Deformable left-ventricle mesh model for motion-compensated filtering in cardiac gated SPECT

被引:11
|
作者
Marin, Thibault [1 ]
Brankov, Jovan G. [1 ]
机构
[1] IIT, Dept Elect & Comp Engn, Med Imaging Res Ctr, Chicago, IL 60616 USA
基金
美国国家卫生研究院;
关键词
cardiac-gated SPECT; motion compensation; motion estimation; mesh modeling; EMISSION TOMOGRAPHY; RECONSTRUCTION; QUANTIFICATION; IMAGES;
D O I
10.1118/1.3483098
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: In this article, the authors present a motion-compensated spatiotemporal processing algorithm to reduce noise in cardiac gated SPECT. Cardiac gated SPECT data are particularly noisy because the acquired photon data are divided among a number of time frames (gates). Classical spatial reconstruction and processing techniques offer noise reduction but they are usually applied on each frame separately and fail to utilize temporal correlation between frames. Methods: In this work, the authors present a motion-compensated spatiotemporal postreconstruction filter offering noise reduction while minimizing motion-blur artifacts. The proposed method can be used regardless of the type of image-reconstruction method (analytical or iterative). The between-frame volumetric myocardium motion is estimated using a deformable mesh model based on the model of the myocardial surfaces. The estimated motion is then used to perform spatiotemporal filtering along the motion trajectories. Both the motion-estimation and spatiotemporal filtering methods seek to maintain the wall brightening seen during cardiac contraction. Wall brightening is caused by the partial volume effect, which is usually viewed as an artifact; however, wall brightening is a useful signature in clinical practice because it allows the clinician to visualize wall thickening. Therefore, the authors seek in their method to preserve the brightening effect. Results: The authors find that the proposed method offers better noise reduction than several existing methods as quantitatively evaluated by signal-to-noise ratio, bias-variance plots, and ejection fraction analysis as well as on tested clinical data. Conclusions: The proposed method mitigates for noise in cardiac gated SPECT images using a postreconstruction motion-compensated filtering approach. Visual as well as quantitative evaluation show considerable improvement in image quality. (C) 2010 American Association of Physicists in Medicine. [DOT: 10.1118/1.3483098]
引用
收藏
页码:5471 / 5481
页数:11
相关论文
共 43 条
  • [1] MOTION-COMPENSATED RECONSTRUCTION OF GATED CARDIAC SPECT IMAGES USING A DEFORMABLE MESH MODEL
    Marin, Thibault
    Wernick, Miles N.
    Yang, Yongyi
    Brankov, Jovan G.
    [J]. 2010 7TH IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: FROM NANO TO MACRO, 2010, : 520 - 523
  • [2] Motion-compensated temporal summation of cardiac gated SPECT images using a deformable mesh model
    Mann, Thibault
    Wernick, Miles N.
    Yang, Yongyi
    Brankov, Jovan G.
    [J]. 2009 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-20, 2009, : 3657 - 3660
  • [3] 4D smoothing of gated SPECT images using a left-ventricle shape model and a deformable mesh
    Brankov, JG
    Yang, YY
    Feng, B
    King, MA
    Wernick, MN
    [J]. 2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 2845 - 2848
  • [4] Motion-compensated dynamic image reconstruction for gated cardiac spect
    Jin, Mingwu
    Yang, Yongyi
    Wernick, Miles N.
    King, Michael A.
    [J]. 2006 3RD IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: MACRO TO NANO, VOLS 1-3, 2006, : 267 - +
  • [5] Motion compensated spatio-temporal filtering of cardiac gated SPECT images
    Marin, Thibault
    Wernick, Miles N.
    Yang, Yongyi
    Brankov, Jovan G.
    [J]. 2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 3009 - 3012
  • [6] Motion-compensated 4D processing of gated SPECT perfusion studies
    Brankov, JG
    Yang, YY
    Narayanan, MV
    Wernick, MN
    [J]. 2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 1380 - 1384
  • [7] EVALUATION OF REGURGITANT FRACTION OF THE LEFT-VENTRICLE BY GATED CARDIAC BLOOD-POOL SCANNING USING SPECT
    OHTAKE, T
    NISHIKAWA, J
    MACHIDA, K
    MOMOSE, T
    MASUO, M
    SERIZAWA, T
    YOSHIZUMI, M
    YAMAOKI, K
    TOYAMA, H
    MURATA, H
    IIO, M
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1987, 28 (01): : 19 - 24
  • [8] Fully 4D motion-compensated reconstruction of cardiac SPECT images
    Gravier, Erwan
    Yang, Yongyi
    King, Michael A.
    Jin, Mingwu
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (18): : 4603 - 4619
  • [9] Motion-compensated fully 4D reconstruction of gated cardiac sequences
    Gravier, E
    Yang, YY
    [J]. COMPUTATIONAL IMAGING III, 2005, 5674 : 305 - 315
  • [10] MOTION OF POSTERIOR WALL OF CARDIAC LEFT-VENTRICLE IN EFFUSIVE PERICARDITIS
    NAKAMURA, K
    MIYAHARA, K
    KITAHARA, T
    IRIE, Y
    MISAKA, K
    OTSUJI, Y
    SATO, H
    [J]. JAPANESE CIRCULATION JOURNAL-ENGLISH EDITION, 1975, 39 (09): : 1021 - 1021