Evaluation of fingertip pulse oximeters for monitoring haemoglobin oxygen saturation in arterial blood and pulse rate in isoflurane-anaesthetised horses breathing greater than 90 percent oxygen

被引:0
|
作者
Sage, A. M. [1 ,2 ]
Ambrisko, T. D. [1 ]
Martins, F. D. C. [1 ,3 ]
Dantino, S. C. [1 ,4 ]
Keating, S. C. J. [1 ]
Strahl-Heldreth, D. E. [1 ]
Wilkins, P. A. [1 ]
机构
[1] Univ Illinois, Coll Vet Med, Dept Vet Clin Med, Urbana, IL 61801 USA
[2] Univ Wisconsin Madison, Dept Surg Sci, Sch Vet Med, Madison, WI USA
[3] Univ Missouri, Coll Vet Med, Dept Vet Med & Surg, Columbia, MO 65211 USA
[4] Univ Tennessee, Coll Vet Med, Dept Large Anim Clin Sci, Knoxville, TN USA
关键词
horse; anaesthesia; pulse oximetry; oxygen saturation; accuracy; CONFIDENTIAL INQUIRY; PARTIAL-PRESSURE; ACCURACY; HYPOXEMIA; CATS; DOGS; PERFORMANCE; MOTION; DEATH; RISK;
D O I
10.1111/eve.13592
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Background Pulse oximetry is a noninvasive method for continuous monitoring of oxygen saturation (SpO(2)) and pulse rate (PR) used in equine anaesthesia. Despite species differences in the oxyhaemoglobin dissociation curve, pulse oximeters validated for humans are commonly used in veterinary medicine. These devices are expensive and at an increased risk for damage when used for equine anaesthesia. Human fingertip pulse oximeters (FPO) may be useful as an inexpensive alternative to standard devices. Objectives Evaluate accuracy and repeatability of FPOs applied to tongues of anaesthetised horses breathing >90% oxygen. Study design Prospective clinical study. Methods Duplicate SpO(2) and PR measurements from eight pulse oximeters (PO; seven FPOs and reference PO) placed in random order on tongues of horses under inhalational anaesthesia were recorded. A standard portable PO (RPO) was used as reference SpO(2) and invasive blood pressure trace or manual pulse palpation as reference PR. Nonparametric Bland-Altman analysis was performed. Acceptable coefficient of repeatability (CR) value was <= 6% SpO(2). Agreements were accepted if limits of agreement were <= +/- 3% difference in SpO(2) and <= +/- 10% relative difference in PR. Only PR values between 25 and 50 beats per minute and their corresponding SpO(2) measurements were used for agreement analysis. Results SpO(2) repeatability and agreement were unacceptable for all devices. The agreement of PR measurements was acceptable for one FPO and the RPO. Main limitations Utilisation of devices designed for the human oxyhaemoglobin dissociation curve, high number of absent or erroneous readings, use of healthy horses, small sample size, unavailability of co-oximetry, lack of a desaturation protocol and difficulty with proper placement of FPOs on horse tongues are all considered limitations in this study. Conclusions Accuracy of FPOs used in anaesthetised horses was poor. Their use in equine anaesthesia for SpO(2) monitoring is not recommended, but some models may be considered for continuous PR monitoring in conjunction with intermittent manual confirmation.
引用
收藏
页码:E540 / E546
页数:7
相关论文
共 13 条
  • [1] Repeatability and accuracy of fingertip pulse oximeters for measurement of hemoglobin oxygen saturation in arterial blood and pulse rate in anesthetized dogs breathing 100% oxygen
    Ambrisko, Tamas D.
    Dantino, Stephanie C.
    Keating, Stephanie C. J.
    Strahl-Heldreth, Danielle E.
    Sage, Adrianna M.
    Martins, Felipe Da Costa
    Harper, Tisha A. M.
    Wilkins, Pamela A.
    AMERICAN JOURNAL OF VETERINARY RESEARCH, 2021, 82 (04) : 268 - 273
  • [2] Evaluation of pulse co-oximetry to determine haemoglobin saturation with oxygen and haemoglobin concentration in anaesthetized horses: a retrospective study
    Zoff, A.
    Dugdale, A. H. A.
    Scarabelli, S.
    Rioja, E.
    VETERINARY ANAESTHESIA AND ANALGESIA, 2019, 46 (04) : 452 - 457
  • [3] Clinical evaluation of a smart wristband for monitoring oxygen saturation, pulse rate, and respiratory rate
    Monnink, Stefan H. J.
    van Vliet, Mariska
    Kuiper, Mathijs J.
    Constandse, Jan C.
    Hoftijzer, Dieke
    Muller, Marjolein
    Ronner, Eelko
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2024,
  • [4] Effects of a novel hemoglobin-based oxygen carrier on percent oxygen saturation as determined with arterial blood gas analysis and pulse oximetry
    Hughes, GS
    Francom, SF
    Antal, EJ
    Adams, WJ
    Locker, PK
    Yancey, EP
    Jacobs, EE
    ANNALS OF EMERGENCY MEDICINE, 1996, 27 (02) : 164 - 169
  • [5] Monitoring oxygen saturation and heart rate during neonatal transition. comparison between two different pulse oximeters and electrocardiography
    Rasha Khoury
    Gil Klinger
    Yakir Shir
    Michael Osovsky
    Ruben Bromiker
    Journal of Perinatology, 2021, 41 : 885 - 890
  • [6] Monitoring oxygen saturation and heart rate during neonatal transition. comparison between two different pulse oximeters and electrocardiography
    Khoury, Rasha
    Klinger, Gil
    Shir, Yakir
    Osovsky, Michael
    Bromiker, Ruben
    JOURNAL OF PERINATOLOGY, 2021, 41 (04) : 885 - 890
  • [7] Technical note: Evaluation of a wireless pulse oximeter for measuring arterial oxygen saturation and pulse rate in newborn Holstein Friesian calves
    Kanz, P.
    Krieger, S.
    Drillich, M.
    Iwersen, M.
    JOURNAL OF DAIRY SCIENCE, 2018, 101 (07) : 6437 - 6442
  • [8] A new fibre optic pulse oximeter probe for monitoring splanchnic organ arterial blood oxygen saturation
    Hickey, M.
    Samuels, N.
    Randive, N.
    Langford, R.
    Kyriacou, P. A.
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2012, 108 (03) : 883 - 888
  • [9] OXYGEN-SATURATION MONITORING IN EXPERIMENTAL SURGERY - A COMPARISON OF PULSE OXIMETRY AND ARTERIAL BLOOD-GAS MEASUREMENT
    VENDER, JR
    HAND, CM
    SEDOR, D
    TABOR, SL
    BLACK, P
    LABORATORY ANIMAL SCIENCE, 1995, 45 (02): : 211 - 215
  • [10] ARTERIAL OXYGEN-SATURATION, BLOOD-PRESSURE, AND PULSE-RATE DURING UPPER GASTROINTESTINAL ENDOSCOPY - INFLUENCE OF SEDATION AND AGE
    KINOSHITA, Y
    ISHIDO, S
    NISHIYAMA, K
    KITAJIMA, N
    ITOH, T
    INATOME, T
    INOH, T
    FUKUZAKI, H
    CHIBA, T
    JOURNAL OF CLINICAL GASTROENTEROLOGY, 1991, 13 (06) : 656 - 660