Calibration evaluation and radiometric testing of field radiometers with the SeaWiFS Quality Monitor (SQM)

被引:0
|
作者
Hooker, SB [1 ]
Aiken, J
机构
[1] NASA, Goddard Space Flight Ctr, Lab Hydrospher Proc, Greenbelt, MD 20771 USA
[2] Plymouth Marine Lab, Plymouth, Devon, England
关键词
D O I
10.1175/1520-0426(1998)015<0995:CEARTO>2.0.CO;2
中图分类号
P75 [海洋工程];
学科分类号
0814 ; 081505 ; 0824 ; 082401 ;
摘要
One of the goals of calibration and validation programs supporting ocean color satellites is to produce water-leaving radiances with an uncertainty of 5% in clear-water regions. This objective requires field instruments with a calibration and measurement capability that is on the order of 1%. The Sea-viewing Wide Field-of-view Sensor (SeaWiFS) Project, in collaboration with the National Institute of Standards and Technology, has developed a portable illumination source with three temperature-stabilized internal monitors designed to provide a stable light field for checking the optical stability of radiometers used to measure the in situ optical properties of seawater. This device is called the SeaWiFS Quality Monitor (SQM). A recent field evaluation during an extensive research cruise indicates the SQM has the following capabilities: (a) the SQM can be used to track the stability of field radiometers at less than the 1% level in terms of the radiometric response of the instruments-on average 0.30% (+/-0.15%) for radiance sensors and 0.58% (+/-0.20%) for irradiance sensors: (b) the SQM light field is sufficiently stable to allow for a sensitive measure and, thus, modeling of changes in the radiometric detectors; (cf based on the radiometers used during the field evaluation, daily SQM measurements are needed to resolve the temporal changes in the response of the sensors; and (d) SQM performance, in terms of the generated light field and the SQM internal monitors, is very stable and decayed only by approximately 0.60% during the course of the 36-day deployment with most of the decay attributed to a change in the operating voltage of one of the lamps.
引用
收藏
页码:995 / 1007
页数:13
相关论文
共 21 条
  • [1] Radiometric and engineering performance of the SeaWiFS Quality Monitor (SQM): A portable light source for field radiometers
    Johnson, BC
    Shaw, PS
    Hooker, SB
    Lynch, D
    [J]. JOURNAL OF ATMOSPHERIC AND OCEANIC TECHNOLOGY, 1998, 15 (04) : 1008 - 1022
  • [2] The SeaWiFS Quality Monitor - A portable field calibration light source
    Shaw, PS
    Johnson, BC
    Hooker, SB
    Lynch, D
    [J]. OCEAN OPTICS XIII, 1997, 2963 : 772 - 776
  • [3] Evaluation of field calibration methods and performance of AQMesh, a low-cost air quality monitor
    Wahlborg, Dan
    Bjorling, Mikael
    Mattsson, Magnus
    [J]. ENVIRONMENTAL MONITORING AND ASSESSMENT, 2021, 193 (05)
  • [4] Evaluation of field calibration methods and performance of AQMesh, a low-cost air quality monitor
    Dan Wahlborg
    Mikael Björling
    Magnus Mattsson
    [J]. Environmental Monitoring and Assessment, 2021, 193
  • [5] VNIR, MWIR, and LWIR source assemblies for optical quality testing and spectro-radiometric calibration of earth observation satellites
    Compain, Eric
    Maquet, Philippe
    Leblay, Pierrick
    Gavaud, Eric
    Marque, Julien
    Glastre, Wilfried
    Cortese, Maxime
    Sugranes, Pierre
    Gaillac, Stephanie
    Potheau, Herve
    [J]. OPTICAL SYSTEMS DESIGN 2015: OPTICAL DESIGN AND ENGINEERING VI, 2015, 9626
  • [6] Evaluation and Field Calibration of a Low-cost Ozone Monitor at a Regulatory Urban Monitoring Station
    Masiol, Mauro
    Squizzato, Stefania
    Chalupa, David
    Rich, David Q.
    Hopke, Philip K.
    [J]. AEROSOL AND AIR QUALITY RESEARCH, 2018, 18 (08) : 2029 - 2037
  • [7] Evaluation of a novel integrated quality control system for a patient self testing INR monitor.
    Goodman, DE
    Blunt, J
    Hewett, G
    [J]. CLINICAL CHEMISTRY, 2004, 50 (06) : A184 - A184
  • [8] Evaluation and field calibration of the Miniram PDM-3 aerosol monitor for measuring respirable and total coal dust
    Middendorf, PJ
    Lehocky, AH
    Williams, PL
    [J]. AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL, 1999, 60 (04): : 502 - 511
  • [9] Power Quality: Harmonic Current Distortion, Standards, Field Testing and Evaluation
    Makasheva, Svetlana
    Pinchukov, Pavel
    [J]. 2020 INTERNATIONAL URAL CONFERENCE ON ELECTRICAL POWER ENGINEERING (URALCON), 2020, : 44 - 49
  • [10] Field Testing, Refinement, and Psychometric Evaluation of a New Measure of Quality of Care for Assisted Living
    Rantz, Marilyn J.
    Aud, Myra A.
    Zwygart-Stauffacher, Mary
    Mehr, David R.
    Petroski, Gregory F.
    Owen, Steven V.
    Madsen, Richard W.
    Flesner, Marcia
    Conn, Vicki
    Maas, Meridean
    [J]. JOURNAL OF NURSING MEASUREMENT, 2008, 16 (01) : 16 - 30