Effect of maximum inhalation flow and inhaled volume on formoterol drug deposition in-vitro from an Easyhaler® dry powder inhaler

被引:20
|
作者
Abadelah, Mohamad [1 ]
Hazim, Firisteh [1 ]
Chrystyn, Henry [3 ]
Bagherisadeghi, Golshan [1 ]
Rahmoune, Hassan [2 ]
Larhrib, Hassan [1 ]
机构
[1] Univ Huddersfield, Dept Pharm & Pharmaceut Sci, Huddersfield HD1 3DH, W Yorkshire, England
[2] Univ Cambridge, Dept Chem Engn & Biotechnol, Cambridge CB2 3RA, England
[3] Inhalat Consultancy Ltd, Leeds LS19 7SP, W Yorkshire, England
基金
英国工程与自然科学研究理事会;
关键词
Andersen Cascade impactor; Easyhaler; Formoterol; Inhaled volume; Maximum inhalation flow; LUNG DEPOSITION; CASCADE IMPACTOR; PARTICLE-SIZE; DEVICE DESIGN; PERFORMANCE; ASTHMA; COPD; AEROSOLIZATION; BUDESONIDE; SALBUTAMOL;
D O I
10.1016/j.ejps.2017.03.035
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Most patients using dry powder inhalers (DPIs) do not achieve the inhalation parameters recommended for pharmacopoeial in-vitro dose emission testing. The dose emission characteristics of formoterol from an Easyhaler((R)) have been measured using the Andersen Cascade Impactor (ACI) with a maximum inhalation flow (MIF) of 28.3, 60 and 90 L/min and inhaled volumes (Vin) of 240, 750, 1500 and 2000 mL. The total emitted dose (TED) was significantly higher at 90 L/min (p < 0.05), but the difference in the TED between low (28.3 L/min) and high (90 L/min) flow rate was significantly reduced by increasing the Vin. The fine particle dose (FPD) was higher (p < 0.05) at 90 L/min for all Vin values compared to 28.3 and 60 L/min. Similarly the mass median aerodynamic diameter (MMAD) was smaller at 90 L/min across all Vin values. Dose emission characteristics were lower at 240 mL for both 1VIIFs. The results for 240 mL could be due to an insufficient Vin pulled through the ACI or incomplete emptying of the dose metering cup. This study shows that the FPD, %FPF and MMAD were not significantly affected by the vin >= 750 mL and that an inhaled volume as low as 750 mL could be used with the ACI.
引用
收藏
页码:180 / 187
页数:8
相关论文
共 50 条
  • [1] Remarks on the article "Effect of maximum inhalation flow and inhaled volume on formoterol drug deposition in-vitro from an Easyhaler dry-powder inhaler" by Abadelah, et al., volume 104, page 180-87 (2017)
    Roberts, Daryl L.
    Mitchell, Jolyon P.
    EUROPEAN JOURNAL OF PHARMACEUTICAL SCIENCES, 2017, 109 : 144 - 145
  • [2] Effect of Inhalation Profiles on the in vitro Deposition Performance of Dry Powder Inhaler Models
    Wu, Fei
    Chen, Lan
    Chen, Donghao
    Chinese Journal of Pharmaceuticals, 2023, 54 (10) : 1467 - 1474
  • [3] EFFECT OF INHALATION FLOW, INHALATION VOLUME AND INHALATION DEVICE CHANGES ON DRY POWDER INHALER TOTAL EMITTED DOSE
    Boshra, Mariane S.
    Almeldien, Ahmed G.
    Eldin, Randa Salah
    Elberry, Ahmed A.
    Abdelwahab, Nada Saed
    Rabea, Hoda
    Abdelrahim, Mohamed
    JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2017, 30 (03) : 4 - 4
  • [4] Effect of Human Error, Inhalation Flow, and Inhalation Volume on Dose Delivery from Ellipta® Dry-Powder Inhaler
    Saeed, Haitham
    Salem, Heba F.
    Rabea, Hoda
    Abdelrahim, Mohamed E. A.
    JOURNAL OF PHARMACEUTICAL INNOVATION, 2019, 14 (03) : 239 - 244
  • [5] Effect of Human Error, Inhalation Flow, and Inhalation Volume on Dose Delivery from Ellipta® Dry-Powder Inhaler
    Haitham Saeed
    Heba F. Salem
    Hoda Rabea
    Mohamed E. A. Abdelrahim
    Journal of Pharmaceutical Innovation, 2019, 14 : 239 - 244
  • [6] Correlation Of Pharmacokinetic Parameters Of Inhaled Drug Delivery With Acoustics Of Inhalation From A Diskus Dry Powder Inhaler
    Seheult, J. N.
    O'Connell, P.
    Sulaiman, I.
    D'Arcy, S.
    Holmes, M.
    Reilly, R.
    Boran, G.
    Costello, R. W.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2014, 189
  • [7] INVITRO AND INVIVO DEPOSITION OF DRUG PARTICLES INHALED FROM PRESSURIZED AEROSOL AND DRY POWDER INHALER
    VIDGREN, M
    KARKKAINEN, A
    KARJALAINEN, P
    NUUTINEN, J
    PARONEN, P
    DRUG DEVELOPMENT AND INDUSTRIAL PHARMACY, 1988, 14 (15-17) : 2649 - 2665
  • [8] EFFECT OF DIFFERENT MODES OF INHALATION ON DRUG-DELIVERY FROM A DRY POWDER INHALER
    NEWMAN, SP
    HOLLINGWORTH, A
    CLARK, AR
    INTERNATIONAL JOURNAL OF PHARMACEUTICS, 1994, 102 (1-3) : 127 - 132
  • [9] A COMPARISON OF THE LUNG DEPOSITION OF SALBUTAMOL INHALED FROM A NEW DRY POWDER INHALER, AT 2 INHALED FLOW-RATES
    PITCAIRN, G
    LUNGHETTI, G
    VENTURA, P
    NEWMAN, S
    INTERNATIONAL JOURNAL OF PHARMACEUTICS, 1994, 102 (1-3) : 11 - 18
  • [10] Flow-dependent effect of formoterol dry-powder inhaled from the Aerolizer(R)
    Nielsen, KG
    Skov, M
    Klug, B
    Ifversen, M
    Bisgaard, H
    EUROPEAN RESPIRATORY JOURNAL, 1997, 10 (09) : 2105 - 2109