Dual-Energy Spectral CT: Various Clinical Vascular Applications

被引:94
|
作者
Machida, Haruhiko [1 ]
Tanaka, Isao [1 ]
Fukui, Rika [1 ]
Shen, Yun [1 ]
Ishikawa, Takuya [1 ]
Tate, Etsuko [1 ]
Ueno, Eiko [1 ]
机构
[1] Tokyo Womens Med Univ, Med Ctr East, Dept Radiol, Arakawa Ku, 2-1-10 Nishiogu, Tokyo 1168567, Japan
基金
日本学术振兴会;
关键词
REDUCED IODINE LOAD; COMPUTED-TOMOGRAPHY; IMAGE QUALITY; ANGIOGRAPHY; CONTRAST; VENOGRAPHY; ARTIFACTS; PHANTOM; SCANNER; REPAIR;
D O I
10.1148/rg.2016150185
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Single-source dual-energy (DE) computed tomography (CT) with fast switching of tube voltage allows projection-based image reconstruction, substantial reduction of beam-hardening effects, reconstruction of accurate monochromatic images and material decomposition images (MDIs), and detailing of material composition by using x-ray spectral information. In vascular applications, DE CT is expected to overcome limitations of standard single-energy CT angiography, including patient exposure to nephrotoxic contrast medium and carcinogenic radiation, insufficient contrast vascular enhancement, interference from metallic and beam-hardening artifacts and severe vessel calcification, and limited tissue characterization and perfusion assessment. Acquisition of low-energy monochromatic images and iodine/water MDIs can reasonably reduce contrast agent dose and improve vessel enhancement. Acquisition of virtual noncontrast images, such as water/iodine MDIs, can reduce overall radiation exposure by replacing true noncontrast CT in each examination. Acquisition of monochromatic images by using metal artifact reduction software or acquisition of iodine/water MDIs can reduce metal artifacts with preserved or increased vessel contrast, and subtraction of monochromatic images between two energy levels can subtract coils composed of dense metallic materials. Acquisition of iodine/calcium (ie, hydroxyapatite) MDIs permits subtraction of vessel calcification and improves vessel lumen delineation. Sensitive detection of lipid-rich plaque can be achieved by using fat/water MDIs, the spectral Hounsfield unit curve (energy level vs CT attenuation), and a histogram of effective atomic numbers included in an image. Various MDIs are useful for accurate differentiation among materials with high attenuation values, including contrast medium, calcification, and fresh hematoma. Iodine/water MDIs are used to assess organ perfusion, such as in the lungs and myocardium. Understanding these DE CT techniques enhances the value of CT for vascular applications. (C) RSNA, 2016.
引用
收藏
页码:1215 / 1232
页数:18
相关论文
共 50 条
  • [1] Dual-Energy CT: Vascular Applications
    Vlahos, Ioannis
    Chung, Raymond
    Nair, Arjun
    Morgan, Robert
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 199 (05) : S87 - S97
  • [2] Dual-Energy CT: Clinical Applications in Various Pulmonary Diseases
    Kang, Mi-Jin
    Park, Chang Min
    Lee, Chang-Hyun
    Goo, Jin Mo
    Lee, Hyun Ju
    RADIOGRAPHICS, 2010, 30 (03) : 685 - 698
  • [3] Clinical Applications of Dual-Energy CT
    Hamid, Saira
    Nasir, Muhammad Umer
    So, Aaron
    Andrews, Gordon
    Nicolaou, Savvas
    Qamar, Sadia Raheez
    KOREAN JOURNAL OF RADIOLOGY, 2021, 22 (06) : 970 - 982
  • [4] Dual-Energy (Spectral) CT: Applications in Abdominal Imaging
    Blake, Michael A.
    RADIOGRAPHICS, 2011, 31 (04) : 1047 - 1050
  • [5] Dual-energy CT: technical considerations and clinical applications
    Fernandez-Perez, G. C.
    Fraga Pineiro, C.
    Onate Miranda, M.
    Diez Blanco, M.
    Mato Chain, J.
    Collazos Martinez, M. A.
    RADIOLOGIA, 2022, 64 (05): : 445 - 455
  • [6] Cardiac Dual-Energy CT Applications and Clinical Impact
    Albrecht M.H.
    De Cecco C.N.
    Nance J.W., Jr.
    Varga-Szemes A.
    De Santis D.
    Eid M.
    Tesche C.
    Apfaltrer G.
    von Knebel Doeberitz P.L.
    Jacobs B.
    Vogl T.J.
    Joseph Schoepf U.
    Current Radiology Reports, 5 (9)
  • [7] Dual-energy CT: theoretical principles and clinical applications
    Agostini, Andrea
    Borgheresi, Alessandra
    Mari, Alberto
    Floridi, Chiara
    Bruno, Federico
    Carotti, Marina
    Schicchi, Nicolo
    Barile, Antonio
    Maggi, Stefania
    Giovagnoni, Andrea
    RADIOLOGIA MEDICA, 2019, 124 (12): : 1281 - 1295
  • [8] Dual-energy CT: theoretical principles and clinical applications
    Andrea Agostini
    Alessandra Borgheresi
    Alberto Mari
    Chiara Floridi
    Federico Bruno
    Marina Carotti
    Nicolò Schicchi
    Antonio Barile
    Stefania Maggi
    Andrea Giovagnoni
    La radiologia medica, 2019, 124 : 1281 - 1295
  • [9] Dual-Energy CT: Dual-KV Techniques and Clinical Applications
    McCollough, C.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [10] Clinical applications of dual-energy CT in head and neck imaging
    Ginat, Daniel Thomas
    Mayich, Michael
    Daftari-Besheli, Laleh
    Gupta, Rajiv
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2016, 273 (03) : 547 - 553