Interdevice comparison of retinal sensitivity assessments in a healthy population: the CenterVue MAIA and the Nidek MP-3 microperimeters

被引:29
|
作者
Balasubramanian, Siva [1 ,2 ]
Uji, Akihito [1 ,2 ]
Lei, Jianqin [1 ,2 ,3 ]
Velaga, Swetha [1 ,2 ]
Nittala, Muneeswar [1 ,2 ]
Sadda, SriniVas [1 ,2 ]
机构
[1] Doheny Eye Inst, Doheny Image Reading Ctr, 1355 San Pablo St, Los Angeles, CA 90033 USA
[2] Univ Calif Los Angeles, David Geffen Sch Med, Dept Ophthalmol, Los Angeles, CA 90095 USA
[3] Xi An Jiao Tong Univ, Affiliated Hosp 1, Xian, Shaanxi, Peoples R China
关键词
PERIMETRY MICROPERIMETRY; MACULAR DEGENERATION; AUTOMATED PERIMETRY; FUNDUS PERIMETRY;
D O I
10.1136/bjophthalmol-2017-310258
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background To compare and correlate the retinal sensitivity measurements obtained with Nidek Microperimetry-3 (MP-3) and the CenterVue Macular Integrity Assessment (MAIA) microperimeters among healthy subjects. Methods In this prospective comparative study, 31 eyes of 23 subjects underwent complete ophthalmological examination including retinal sensitivity assessments using two microperimeters, the MP-3 (Nidek Technologies) and the MAIA (CenterVue). The mean retinal sensitivity (dB) and its corresponding luminance (asb) and contrast (log units) were analysed between the two instruments. The interdevice reproducibility and level of agreement between the sensitivity values of the devices were assessed. Results The mean retinal sensitivity (dB) measured by the MP-3 (25.02 +/- 1.06 dB, range: 20.90-26.70) was significantly (p < 0.0001) lower compared with the MAIA (30.68 +/- 0.74 dB, range: 28-31.84). The luminosity levels were significantly (p < 0.0001) higher with the MP3 (7.75 +/- 1.31 asb, range: 6.44-9.06) compared with the MAIA (0.92 +/- 0.14 asb, range: 0.78-1.06). The contrast sensitivity was significantly higher for the MP-3 (0.94 +/- 0.33 log units, range: 0.61-1.27) compared with the MAIA (0.23 +/- 0.03 log units, range: 0.20-0.26). Despite these absolute differences, the intraclass coefficient was 0.85 (95% CI 0.70 to 0.92) between the two devices after applying a standard correction factor to each data point (MAIA sensitivity=MP-3 sensitivity+5.65) with a mean difference between MAIA and MP-3 of 0.01. Conclusion Retinal sensitivity measures higher, but luminance and contrast sensitivity measure lower for MAIA-generated values compared with the MP-3. The relationships, however, appeared fairly consistent, and application of a standard correction factor allowed the data to be inter-related, at least for normal eyes.
引用
收藏
页码:109 / 113
页数:5
相关论文
共 8 条
  • [1] Inter-device comparison of retinal sensitivity assessments in a healthy population: the CenterVue MAIA and the Nidek MP-3 microperimeters
    Karamat, Ayesha
    Balasubramanian, Siva
    Hariri, Amir H.
    Tepelus, Tudor
    Uji, Akihito
    Sadda, Srinivas R.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2017, 58 (08)
  • [2] A COMPARISON OF FEASIBILITY: THE NIDEK MP-1S AND THE CENTERVUE MAIA MICROPERIMETERS
    Wong, Evan
    Isaacs, Tim
    Constable, Ian
    McAllister, Ian
    Chen, Fred
    [J]. CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2012, 40 : 116 - 117
  • [3] Inter-device comparison of retinal sensitivity measurements: the CenterVue MAIA and the Nidek MP-1
    Wong, Evan N.
    Mackey, David A.
    Morgan, William H.
    Chen, Fred K.
    [J]. CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2016, 44 (01): : 15 - 23
  • [4] AN INTER-DEVICE, INTRA-SESSION COMPARISON OF 2 AUTOMATED MICROPERIMETERS: THE NIDEK MP-1S AND THE CENTERVUE MAIA
    Wong, Evan
    Isaacs, Timothy
    McAllister, Ian
    Constable, Ian
    Chen, Fred
    [J]. CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2012, 40 : 116 - 116
  • [5] A comparison of microperimetry parameters derived from the CenterVue MAIA and Nidek MP1S instruments in subjects with type 2 idiopathic macular telangiectasia (MacTel)
    Chen, Fred
    Constable, Ian
    Wong, Evan
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (15)
  • [6] MP-3 measurement of retinal sensitivity in macular hole area and its predictive value on visual prognosis
    Wang, Zengyi
    Qi, Yue
    Liang, Xida
    Yu, Yanping
    Chen, Jinqiu
    Wang, Jing
    Liu, Xinxin
    Liu, Wu
    [J]. INTERNATIONAL OPHTHALMOLOGY, 2019, 39 (09) : 1987 - 1994
  • [7] Comparison of Macular Integrity Assessment (MAIA™), MP-3, and the Humphrey Field Analyzer in the Evaluation of the Relationship between the Structure and Function of the Macula
    Hirooka, Kazuyuki
    Misaki, Kana
    Nitta, Eri
    Ukegawa, Kaori
    Sato, Shino
    Tsujikawa, Akitaka
    [J]. PLOS ONE, 2016, 11 (03):
  • [8] MP-3 measurement of retinal sensitivity in macular hole area and its predictive value on visual prognosis
    Zengyi Wang
    Yue Qi
    Xida Liang
    Yanping Yu
    Jinqiu Chen
    Jing Wang
    Xinxin Liu
    Wu Liu
    [J]. International Ophthalmology, 2019, 39 : 1987 - 1994