New bench test for venous cannula performance assessment

被引:13
|
作者
Abdel-Sayed, S. [1 ]
Favre, J. [2 ]
Horisberger, J. [1 ]
Taub, S. [1 ]
Hayoz, D. [3 ]
von Segesser, L. K. [1 ]
机构
[1] CHU Vaudois, Dept Cardiovasc Surg, CH-1011 Lausanne, Switzerland
[2] Swiss Fed Inst Technol, Lab Movement & Measurement, CH-1015 Lausanne, Switzerland
[3] Hop Cantonal Fribourg, Dept Internal Med, Fribourg, Switzerland
来源
PERFUSION-UK | 2007年 / 22卷 / 06期
关键词
cardiopulmonary bypass; cannulas; venous drainage; flow rate; lumen resistance;
D O I
10.1177/0267659108089759
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Cannula design is of prime importance for venous drainage during cardiopulmonary bypass (CPB). To evaluate cannulas intended for CPB, an in vitro circuit was set up with silicone tubing between the test cannula encased in a movable preload reservoir and another static reservoir. The pressure-drop (Delta P) value (P-drainage - P-preload) was measured using Millar pressure transducers. Flow rate (Q) was measured using an ultrasound flowmeter. Data display and data recording were controlled using a LabView application, custom made particularly for our experiments. Our results demonstrated that Delta P, Q, and cannula resistance (Delta P/Q) values were significantly decreased when the cannula diameter was increased for Smart and Medtronic cannulas. Smartcanula(R) showed 36% and 43% less resistance compared to Medtronic venous and Medtronic femoral cannulas, respectively. The cannula shape (straight- or curved-tips) did not affect the DLP cannula resistance. Out of five cannulas tested, the Smartcanula(R) outperforms the other commercially available cannulas. The mean (Delta P/Q) values were 3.3 +/- 0.08, 4.07 +/- 0.08, 5.58 +/- 0.10, 5.74 +/- 0.15, and 6.45 +/- 0.15 for Smart, Medtronic, Edwards, Sarns, and Gambro cannulas, respectively (two-way ANOVA, p < 0.0001). In conclusion, the present assay allows discrimination between different forms of cannula with high or low lumen resistance.
引用
收藏
页码:411 / 416
页数:6
相关论文
共 50 条
  • [1] SELF EXPANDING VENOUS CANNULA PERFORMANCE ASSESSMENT
    Abdel-Sayed, S.
    Favre, J.
    Jaber, K.
    von Segesser, L. K.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2009, 32 (07): : 432 - 432
  • [2] Venous cannula performance assessment in a realistic caval tree model
    Li, Liang
    Abdel-Sayed, Saad
    Berdajs, Denis
    Ferrari, Enrico
    von Segesser, Ludwig K.
    INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2015, 20 (02) : 194 - 199
  • [3] Assessment of the Performance of Agricultural Tires Using a Mobile Test Bench
    Fanigliulo, Roberto
    Biocca, Marcello
    Grilli, Renato
    Fornaciari, Laura
    Gallo, Pietro
    Benigni, Stefano
    Mattei, Paolo
    Pochi, Daniele
    AGRICULTURE-BASEL, 2023, 13 (01):
  • [4] A NEW VENOUS CANNULA FOR USE DURING CPB
    BAHADIR, I
    DIETHRICH, EB
    ANNALS OF THORACIC SURGERY, 1984, 37 (05): : 439 - 440
  • [5] A new uterine cannula for use in the Rubin test
    Gladstone, E
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1925, 9 : 695 - 696
  • [6] PERCUTANEOUS VENOUS CANNULA
    LAUB, GW
    ANNALS OF THORACIC SURGERY, 1989, 48 (04): : 610 - 610
  • [7] A new experimental test bench for a high performance double electropneumatic actuator system
    Girin, Alexis
    Plestan, Franck
    2009 AMERICAN CONTROL CONFERENCE, VOLS 1-9, 2009, : 3488 - 3493
  • [8] VENOUS CANNULA FOR CPB
    PHILLIPS, SJ
    ANNALS OF THORACIC SURGERY, 1981, 32 (06): : 620 - 620
  • [9] VEIN TEST BENCH - NEW INSTRUMENT
    KARTHAUS, AJ
    VASA-JOURNAL OF VASCULAR DISEASES, 1974, 3 (02): : 200 - 202
  • [10] A new expandable venous cannula for minimal access heart surgery
    Mueller, XM
    Tevaearai, HT
    Jegger, D
    Horisberger, J
    Mucciolo, G
    von Segesser, LK
    ANNALS OF THORACIC SURGERY, 2002, 74 (04): : S1330 - S1333