Accelerated imaging of metallic implants using model-based nonlinear reconstruction

被引:5
|
作者
Shi, Xinwei [1 ,2 ]
Levine, Evan [1 ,2 ]
Weber, Hans [1 ]
Hargreaves, Brian A. [1 ,2 ,3 ]
机构
[1] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
[2] Stanford Univ, Dept Elect Engn, Stanford, CA 94305 USA
[3] Stanford Univ, Dept Bioengn, Stanford, CA 94305 USA
基金
美国国家卫生研究院;
关键词
metal implants; model-based reconstruction; multi-spectral imaging; undersampling; TOTAL HIP-ARTHROPLASTY; ARTIFACT CORRECTION; MRI; SUSCEPTIBILITY; SENSE;
D O I
10.1002/mrm.27536
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To accelerate imaging near metallic implants with multi-spectral imaging (MSI) techniques by exploiting a signal model in the spectral dimension. Methods: MSI techniques resolve metal-induced field perturbations by acquiring separate 3D spatial encodings at multiple excitation frequencies, which are referred to as spectral bins. The proposed model-based reconstruction exploits the correlation between spectral bins in image reconstruction by enforcing a signal model to describe the signal profile across bins. This work evaluates the accuracy of the MSI signal model in simulations and in vivo experiments. The proposed model-based reconstruction was evaluated in 6 subjects at an overall undersampling factor of 17.4 and compared with model-free parallel imaging and compressed sensing (PI & CS). The quality of reconstructed images was evaluated using normalized RMS error (nRMSE) and structural similarity index (SSIM) comparisons, with paired Wilcoxon tests in 6 subjects used to determine whether there was a significant difference in the metrics. Results: Both simulations and in vivo experiments show that the proposed signal model can represent the MSI signal profiles in the spectral dimension compactly and accurately. In the in vivo experiments, the model-based reconstruction significantly improved image quality over model-free PI & CS, with P < 0.05 for both nRMSE and SSIM at 17.4x acceleration. Conclusion: This work presents the feasibility of using a model-based reconstruction to accelerate MSI techniques for faster MR imaging near metal.
引用
收藏
页码:2247 / 2263
页数:17
相关论文
共 50 条
  • [1] Accelerated 4D Quantitative Single Point EPR Imaging Using Model-Based Reconstruction
    Jang, Hyungseok
    Matsumoto, Shingo
    Devasahayam, Nallathamby
    Subramanian, Sankaran
    Zhuo, Jiachen
    Krishna, Murali C.
    McMillan, Alan B.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2015, 73 (04) : 1692 - 1701
  • [2] Model-based reconstruction for simultaneous multislice and parallel imaging accelerated multishot diffusion tensor imaging
    Dong, Zijing
    Dai, Erpeng
    Wang, Fuyixue
    Zhang, Zhe
    Ma, Xiaodong
    Yuan, Chun
    Guo, Hua
    [J]. MEDICAL PHYSICS, 2018, 45 (07) : 3196 - 3204
  • [3] Wave-Encoded Model-Based Deep Learning for Highly Accelerated Imaging with Joint Reconstruction
    Cho, Jaejin
    Gagoski, Borjan
    Kim, Tae Hyung
    Tian, Qiyuan
    Frost, Robert
    Chatnuntawech, Itthi
    Bilgic, Berkin
    [J]. BIOENGINEERING-BASEL, 2022, 9 (12):
  • [4] Model-Based Reconstruction for Magnetic Particle Imaging
    Knopp, Tobias
    Sattel, Timo F.
    Biederer, Sven
    Rahmer, Juergen
    Weizenecker, Juergen
    Gleich, Bernhard
    Borgert, Joern
    Buzug, Thorsten M.
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2010, 29 (01) : 12 - 18
  • [5] Highly accelerated parameter mapping using model-based alternating reconstruction coupling fitting
    Li, Shaohang
    Wang, Lili
    Priest, Andrew N.
    Horvat-Menih, Ines
    Mendichovszky, Iosif A.
    Gallagher, Ferdia A.
    Wang, He
    Li, Hao
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (14):
  • [6] A practical guide for model-based reconstruction in optoacoustic imaging
    Dean-Ben, Xose Luis
    Razansky, Daniel
    [J]. FRONTIERS IN PHYSICS, 2022, 10
  • [7] ACCELERATED IMAGE RECONSTRUCTION FOR NONLINEAR DIFFRACTIVE IMAGING
    Ma, Yanting
    Mansour, Hassan
    Liu, Dehong
    Boufounos, Petros T.
    Kamilov, Ulugbek S.
    [J]. 2018 IEEE INTERNATIONAL CONFERENCE ON ACOUSTICS, SPEECH AND SIGNAL PROCESSING (ICASSP), 2018, : 6473 - 6477
  • [8] Equilibrium Model With Anisotropy for Model-Based Reconstruction in Magnetic Particle Imaging
    Maass, Marco
    Kluth, Tobias
    Droigk, Christine
    Albers, Hannes
    Scheffler, Konrad
    Mertins, Alfred
    Knopp, Tobias
    [J]. IEEE Transactions on Computational Imaging, 2024, 10 : 1588 - 1601
  • [9] Model-Based Reconstruction for X-ray Diffraction Imaging
    Sridhar, Venkatesh
    Kisner, Sherman J.
    Skatter, Sondre
    Bouman, Charles A.
    [J]. ANOMALY DETECTION AND IMAGING WITH X-RAYS (ADIX), 2016, 9847
  • [10] Model-based network architecture for image reconstruction in lensless imaging
    Zeng, Tianjiao