Comparison of the Dosimetric Influence of Applicator Displacement on 2D and 3D Brachytherapy for Cervical Cancer Treatment

被引:5
|
作者
Wu, Ailin [1 ]
Tang, Du [2 ]
Wu, Aidong [1 ]
Liu, Yunqin [1 ]
Qian, Liting [1 ]
Zhu, Lei [3 ]
机构
[1] Univ Sci & Technol China, Affiliated Hosp USTC 1, Div Life Sci & Med, Dept Radiat Oncol, Hefei 230031, Anhui, Peoples R China
[2] Cent South Univ, Xiangya Hosp, Dept Oncol, Changsha, Hunan, Peoples R China
[3] Univ Sci & Technol China, Dept Engn & Appl Phys, Hefei, Anhui, Peoples R China
基金
中国国家自然科学基金;
关键词
applicator displacement; 2D brachytherapy; 3D brachytherapy; dosimetric influence; cervical cancer; DOSE-RATE BRACHYTHERAPY; IMAGE-GUIDED BRACHYTHERAPY; GEC-ESTRO; WORKING GROUP; RECOMMENDATIONS; POSITION; RECONSTRUCTION; UNCERTAINTIES; VARIABILITY; VALIDATION;
D O I
10.1177/15330338211041201
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
To compare the dosimetric influence of applicator displacement on two-dimensional brachytherapy (2D-BT) and three-dimensional brachytherapy (3D-BT) for cervical cancer. Nineteen patients who received computed tomography-guided tandem-and-ovoid (T&O) brachytherapy were retrospectively selected. Both 2D (point-based) and 3D (volume-based) plans with and without virtual applicator displacement in the 3 axes were created for each patient. Dose changes at point A, D-90 of the high-risk clinical target volume (HR-CTV) and intermediate-risk CTV (IR-CTV), and the D-0.1cc, D-1cc, D-2cc, and D-5cc of organs-at-risk (OARs) caused by applicator displacement were evaluated. Both 2D-BT and 3D-BT plans were sensitive to T&O applicator displacement. The D-90 of the CTV and the dose at point A were very sensitive to applicator displacement in the right-left direction (X-axis). An applicator shift of >2 mm in the X-axis resulted in a change of >5% in the dose at point A and D-90 of HR-CTV and IR-CTV. In addition, the doses to the OARs were mostly affected by applicator displacement in the anterior-posterior direction (Z-axis). A displacement of Z-axis was required to avoid a dose change of >10% for OARs. For both 2D-BT and 3D-BT plans, T&O displacement greater than +/- 2 mm in the X-axis or T&O applicator displacement +/- 1.5 mm in the Z-axis resulted in significant dose changes to the tumor and OARs. In comparison with 3D-BT plans, 2D-BT plans delivered a higher dose to the tumor, and the OARs received more undesirable doses when applicator displacement occurred. The influence of applicator displacement on the doses to the tumor and OARs differed between 2D-BT and 3D-BT. Physicians should take individual patient differences into account when selecting a brachytherapy plan to mitigate the influence of applicator displacement.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Dosimetric impacts of applicator displacements and applicator reconstruction-uncertainties on 3D image-guided brachytherapy for cervical cancer
    Schindel, Joshua
    Zhang, Winson
    Bhatia, Sudershan K.
    Sun, Wenqing
    Kim, Yusung
    [J]. JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2013, 5 (04) : 250 - 257
  • [2] Dosimetric Comparison of 2D and 3D Episcleral Plaque Treatment Planning
    Dapaah-Afriyie, B.
    Wilkinson, D.
    Astrahan, M.
    Singh, A.
    Berry, J.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 3080 - 3080
  • [3] Brachytherapy Technique from 2D to 3D in Cervical Cancer - Short Overview
    Manea, Elena
    Condorovici, Daniela
    Ciobanu, T.
    Munteanu, Anca
    [J]. 2019 E-HEALTH AND BIOENGINEERING CONFERENCE (EHB), 2019,
  • [4] A COMPARISON OF TOXICITIES IN CERVICAL CANCER PATIENTS TREATED USING 2D VERSUS 3D BRACHYTHERAPY PLANNING
    Abdullah, T.
    Chitoni, M.
    Reed, N.
    Harrand, R.
    Graham, K.
    Sadozye, A.
    [J]. INTERNATIONAL JOURNAL OF GYNECOLOGICAL CANCER, 2015, 25 (09) : 911 - 911
  • [5] DOSE AND VOLUME PARAMETERS FOR CT BASED TREATMENT PLANNING IN INTRACAVITARY BRACHYTHERAPY FOR CERVICAL CANCER. A COMPARISON BETWEEN 2D AND 3D BRACHYTHERAPY
    Enciu, Lucia Elena
    Constantinescu, Amalia
    Savu, Mircea
    Baranga, Gabriel
    Oprea, Alexandru
    Croitoriu, Madalina
    Busuioc, Laurentiu
    [J]. ROMANIAN REPORTS IN PHYSICS, 2018, 70 (02)
  • [6] Dosimetric Verification of a 3D Printed HDR Skin Brachytherapy Applicator
    Rasmussen, K.
    Baumgarten, A.
    Stanley, D.
    Pelletier, C.
    Corbett, M.
    Jung, J.
    Feng, Y.
    Huang, Z.
    Ju, A.
    Eng, T.
    Kirby, N.
    Gutierrez, A.
    Stathakis, S.
    Papanikolaou, N.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3624 - 3624
  • [7] Applicator reconstruction and applicator shifts in 3D MR-based PDR brachytherapy of cervical cancer
    De Leeuw, Astrid A. C.
    Moerland, Marinus A.
    Nomden, Christel
    Tersteeg, Robert H. A.
    Roesink, Judith M.
    Jurgenliemk-Schulz, Ira M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2009, 93 (02) : 341 - 346
  • [8] Pulse dose-rate brachytherapy and treatment of uterine cervix cancer:: impact of a 3D or a 2D dosimetric support
    Tournat, H.
    Chilles, A.
    Charra-Brunaud, C.
    Peiffert, D.
    Ahmad, F.
    Metayer, Y.
    [J]. CANCER RADIOTHERAPIE, 2007, 11 (04): : 188 - 196
  • [9] Impact of brachytherapy technique (2D versus 3D) on outcome following radiotherapy of cervical cancer
    Derks, Kris
    Steenhuijsen, Jacco L. G.
    van den Berg, Hetty A.
    Houterman, Saskia
    Cnossen, Jeltsje
    van Haaren, Paul
    De Jaeger, Katrien
    [J]. JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2018, 10 (01) : 17 - 25
  • [10] Vaginal dose point reporting in cervical cancer patients treated with combined 2D/3D external beam radiotherapy and 2D/3D brachytherapy
    Westerveld, Henrike
    Poetter, Richard
    Berger, Daniel
    Dankulchai, Pittaya
    Doerr, Wolfgang
    Sora, Mircea-Constantin
    Poetter-Lang, Sarah
    Kirisits, Christian
    [J]. RADIOTHERAPY AND ONCOLOGY, 2013, 107 (01) : 99 - 105