3D Monte Carlo-based treatment planning evaluation of intracavitary HDR balloon catheter brachytherapy

被引:0
|
作者
Lerma, F [1 ]
Le, Y [1 ]
Kelley, J [1 ]
Arthur, D [1 ]
Williamson, J [1 ]
机构
[1] Virginia Commonwealth Univ, Richmond, VA USA
关键词
D O I
10.1118/1.1997688
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate the dosimetric effects of missing tissue and tissue heterogeneities on intracavitary HDR balloon‐catheter brachytherapy using a novel 3D Monte Carlo‐based treatment planning study. Method and Materials: The PTRAN Monte Carlo (MC) photon transport code is used to generate 3D calculations of the doses arising from an HDR 192Ir source in; Anatomy‐based MC simulations modeling the dosimetric effects of missing tissue beyond the skin surface and tissue heterogeneities (such as air pockets) adjacent to the balloon catheter (Mammosite); and in a large water phantom representing those calculations achievable with a conventional 3D brachytherapy RTPS. Each MC simulation employs a track length estimator scoring method in a 3D grid, with 2mm resolution in a 12cm×11cm×8cm volume, scoring 8×106 primary photon histories. The dose matrices from all MC simulation are registered (in the external beam module of a commercial RTPS) to a single, modeled patient CT, which is identical in location, scale, and shape to the patient volume described in the anatomy‐based MC simulations. Several target volumes (e.g. a 1cm margin on the Mammosite balloon) and a skin‐surface contour, are created. Dose‐volume histograms (DVH's) are computed for each volume, per each plan. Comparisons of the resultant DVH's quantify the dosimetric errors expected from routine clinical planning of Mammosite applications. Results: The missing tissue beyond the skin surface reduces target coverage by 1.5–3.5% and skin doses by 8–20%, when compared to calculations achievable with conventional 3D brachytherapy treatment planning systems. A 10cc airpocket neighboring the Mammosite balloon reduces doses to adjacent tissues <0.5%. Conclusion: A detailed 3D Monte Carlo‐based brachytherapy RTP study of the dose prediction errors in balloon catheter HDR applications shows that the missing tissue beyond the skin surface reduces skin doses by as much as 20% and target coverage by as much as 3.5%. © 2005, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:1953 / 1953
页数:1
相关论文
共 50 条
  • [1] RapidBrachyMCTPS: a Monte Carlo-based treatment planning system for brachytherapy applications
    Famulari, Gabriel
    Renaud, Marc-Andre
    Poole, Christopher M.
    Evans, Michael D. C.
    Seuntjens, Jan
    Enger, Shirin A.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (17):
  • [2] Monte Carlo-based inverse treatment planning
    Jeraj, R
    Keall, P
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (08): : 1885 - 1896
  • [3] Verification of a Monte Carlo-based treatment planning toot
    Chow, J.
    Leung, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S92 - S92
  • [4] RapidBrachyMCTPS: Addition of Contouring Tools to the Monte Carlo-Based Treatment Planning System for Brachytherapy Applications
    Glickman, Harry
    Antaki, Majd
    Enger, Shirin A.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E580 - E580
  • [5] An approach to using conventional brachytherapy software for clinical treatment planning of complex, Monte Carlo-based brachytherapy dose distributions
    Rivard, Mark J.
    Melhus, Christopher S.
    Granero, Domingo
    Perez-Calatayud, Jose
    Ballester, Facundo
    [J]. MEDICAL PHYSICS, 2009, 36 (06) : 1968 - 1975
  • [6] Monte Carlo-Based Optimization for Lung SBRT Treatment Planning
    Zhong, H.
    Li, H.
    -Hurst, C. Glide
    Wen, N.
    Siebers, J.
    Nurushev, T.
    Chetty, I.
    [J]. MEDICAL PHYSICS, 2011, 38 (06) : 3648 - +
  • [7] Pre-Clinical Assessment of HDR Eye Plaque Treatment Using a Monte Carlo-Based Treatment Planning System
    Nunez, D. Aramburu
    Antaki, M.
    Morcos, M.
    Huang, S.
    Enger, S. Abbasinejad
    Barker, C.
    Damato, A.
    Cohen, G.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E830 - E831
  • [8] A Monte Carlo-based treatment planning tool for proton therapy
    Mairani, A.
    Bohlen, T. T.
    Schiavi, A.
    Tessonnier, T.
    Molinelli, S.
    Brons, S.
    Battistoni, G.
    Parodi, K.
    Patera, V.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (08): : 2471 - 2490
  • [9] OPTIMISATION TECHNIQUE FOR MRI BASED 3D TREATMENT PLANNING FOR HDR BRACHYTHERAPY CERVIX TREATMENTS
    Aldridge, S.
    Block, G.
    Naeem, M.
    Johnstone, R.
    Humphreys, S.
    Winship, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S266 - S266
  • [10] Preliminary experience with 3D ultrasound-based treatment planning for HDR prostate brachytherapy
    Falco, T
    Evans, M
    Souhami, L
    Duclos, M
    Aprikian, A
    Parker, W
    Patrocinio, H
    Podgorsak, E
    [J]. MEDICAL PHYSICS, 2003, 30 (07) : 1941 - 1941