Evaluation of Tibial Fixation Devices for Quadrupled Hamstring ACL Reconstruction

被引:6
|
作者
Ammann, Elias [1 ]
Hecker, Andreas [1 ]
Bachmann, Elias [1 ]
Snedeker, Jess G. [1 ]
Fucentese, Sandro F. [1 ]
机构
[1] Balgrist Univ Hosp, Zurich, Switzerland
关键词
anterior cruciate ligament; tibia; fixation; biomechanics; CRUCIATE LIGAMENT RECONSTRUCTION; GRAFT FIXATION; BIOMECHANICAL EVALUATION; SUSPENSORY FIXATION; INTERFERENCE SCREW; TENDON GRAFTS; LOOP; TUNNEL; TENSION; KNEE;
D O I
10.1177/23259671221096107
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Shortcomings to tibial-side fixation have been reported as causes of failure after anterior cruciate ligament reconstruction. Adjustable-loop suspensory devices have become popular; however, no comparison with hybrid fixation (ie, interference screw and cortical button) exists to our knowledge. Purpose: The purpose of this study was to compare the biomechanical properties of adjustable loop devices (ALDs) in full-tunnel and closed-socket configurations in relation to hybrid fixation. We hypothesized that primary stability of fixation by a tibial ALD will not be inferior to hybrid fixation. Study Design: Controlled laboratory study. Methods: Tibial fixation of a quadrupled tendon graft was biomechanically investigated in a porcine tibia-bovine tendon model using 5 techniques (n = 6 specimens each). The tested constructs included hybrid fixation with a cortical fixation button and interference screw (group 1), single cortical fixation with the full-tunnel technique using an open-suture strand button (group 2) or an ALD (group 3), or closed-socket fixation using 2 different types of ALDs (groups 4 and 5). Each specimen was evaluated using a materials testing machine (1000 cycles from 50-250 N and pull to failure). Force at failure, cyclic displacement, stiffness, and ability to pretension the graft during insertion were compared among the groups. Results: No differences in ultimate load to failure were found between the ALD constructs (groups 3, 4, and 5) and hybrid fixation (group 1). Cyclic displacement was significantly higher in group 2 vs all other groups (P < .001); however, no difference was observed in groups 3, 4, and 5 as compared with group 1. The remaining tension on the construct after fixation was significantly higher in groups 3 and 4 vs groups 1, 2, and 5 (P < .02 for all comparisons), irrespective of whether a full-tunnel or closed-socket approach was used. Conclusion: Tibial anterior cruciate ligament graft fixation with knotless ALDs achieved comparable results with hybrid fixation in the full-tunnel and closed-socket techniques. The retention of graft tension appears to be biomechanically more relevant than tunnel type.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Tibial press-fit fixation of the hamstring tendons for ACL-reconstruction
    M. Jagodzinski
    K. Scheunemann
    K. Knobloch
    K. Albrecht
    C. Krettek
    C. Hurschler
    J. Zeichen
    [J]. Knee Surgery, Sports Traumatology, Arthroscopy, 2006, 14 : 1281 - 1287
  • [2] Tibial press-fit fixation of the hamstring tendons for ACL-reconstruction
    Jagodzinski, M.
    Scheunemann, K.
    Knobloch, K.
    Albrecht, K.
    Krettek, C.
    Hurschler, C.
    Zeichen, J.
    [J]. KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2006, 14 (12) : 1281 - 1287
  • [3] The fixation of hamstring tendons in ACL reconstruction
    Aglietti, P
    Giron, F
    [J]. 6TH WORLD CONGRESS OF ENDOSCOPIC SURGERY, PTS 1 AND 2, 1998, : 695 - 700
  • [4] Biomechanical evaluation of a bioabsorbable expansion bolt for hamstring graft fixation in ACL reconstruction: an experimental study in calf tibial bone
    Piltz, S
    Dieckmann, R
    Meyer, L
    Strunk, P
    Plitz, W
    Lob, G
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2005, 125 (09) : 577 - 584
  • [5] Biomechanical evaluation of a bioabsorbable expansion bolt for hamstring graft fixation in ACL reconstruction: an experimental study in calf tibial bone
    S. Piltz
    R. Dieckmann
    L. Meyer
    P. Strunk
    W. Plitz
    G. Lob
    [J]. Archives of Orthopaedic and Trauma Surgery, 2005, 125 : 577 - 584
  • [6] Clinical stability and outcome of supplementing tibial fixation with a staple for ACL reconstruction using hamstring tendons
    Lim, Chin Tat
    Tan, Ken-Jin
    Chuan, Ang Kian
    [J]. CURRENT ORTHOPAEDIC PRACTICE, 2009, 20 (06): : 660 - 664
  • [7] Tibial tunnel enlargement is affected by the tunnel diameter-screw ratio in tibial hybrid fixation for hamstring ACL reconstruction
    Flury, Andreas
    Wild, Linda
    Waltenspul, Manuel
    Zindel, Christoph
    Vlachopoulos, Lazaros
    Imhoff, Florian B.
    Fucentese, Sandro F.
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2023, 143 (04) : 1923 - 1930
  • [8] Tibial tunnel enlargement is affected by the tunnel diameter-screw ratio in tibial hybrid fixation for hamstring ACL reconstruction
    Wild, Linda
    Flury, Andreas
    Waltenspul, Manuel
    Zindel, Christoph
    Vlachopoulos, Lazaros
    Imhoff, Florian
    Fucentese, Sandro F.
    [J]. SWISS MEDICAL WEEKLY, 2022, 152 : 38S - 38S
  • [9] Tibial tunnel enlargement is affected by the tunnel diameter-screw ratio in tibial hybrid fixation for hamstring ACL reconstruction
    Andreas Flury
    Linda Wild
    Manuel Waltenspül
    Christoph Zindel
    Lazaros Vlachopoulos
    Florian B. Imhoff
    Sandro F. Fucentese
    [J]. Archives of Orthopaedic and Trauma Surgery, 2023, 143 : 1923 - 1930
  • [10] Evaluation of Isokinetic Knee Strengths after ACL Reconstruction with Quadrupled Semitendinosus Suspensory Femoral and Tibial Fixation versus Four-Strand Semitendinosus and Gracilis Suspensory Femoral and Tibial Screw Fixation
    Cerci, Mehmet Halis
    Yilmaz, Ali Kerim
    Kehribar, Lokman
    Surucu, Serkan
    Aydin, Mahmud
    Mahirogullari, Mahir
    [J]. JOURNAL OF CLINICAL MEDICINE, 2023, 12 (12)