A Non-Invasive Method for Hydration Status Measurement With a Microwave Sensor Using Skin Phantoms

被引:11
|
作者
Kilpijarvi, Joni [1 ]
Tolvanen, Jarrkko [1 ]
Juuti, Jari [1 ]
Halonen, Niina [1 ]
Hannu, Jari [1 ]
机构
[1] Univ Oulu, Microelect Res Unit, Fac Informat Technol & Elect Engn, FI-90014 Oulu, Finland
关键词
Microwave sensors; biomedical monitoring; dielectric measurement; skin phantoms; health monitoring; DIELECTRIC-PROPERTIES; WATER;
D O I
10.1109/JSEN.2019.2945817
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Fluid balance is important for a healthy human being. In this paper, a method to measure hydration status was developed and tested towards non-invasive measurement from human skin. Measurement of hydration status was performed by a microwave sensor utilizing a complementary split ring resonator (CSRR). The sensor was modeled, manufactured and then characterized by measuring tailor made skin phantoms based on the realistic electrical properties of skin with different degrees of hydration status. Qualitative longer term (>24 h) evaluation of the sensor was also performed by measuring polyester tissue that was drying over the time. Hydration status, represented by dehydrated, normal and hydrated skin phantoms, based on polyurethane with carbon and ceramic additives, was measured successfully by monitoring the changes in resonance frequency around 5.52 GHz. All results were compared to the dielectric reference measurements done by a commercial laboratory instrument.
引用
收藏
页码:1095 / 1104
页数:10
相关论文
共 50 条
  • [1] Non-invasive Measurement of Complex permittivity using a Compact Planar Microwave Sensor
    Moolat, Remsha
    Mani, Manoj
    Pezholil, Mohanan
    [J]. 2022 3RD URSI ATLANTIC AND ASIA PACIFIC RADIO SCIENCE MEETING (AT-AP-RASC), 2022,
  • [2] Quantitative and simultaneous non-invasive measurement of skin hydration and sebum levels
    Ezerskaia, Anna
    Pereira, S. F.
    Urbach, H. Paul
    Verhagen, Rieko
    Varghese, Babu
    [J]. BIOMEDICAL OPTICS EXPRESS, 2016, 7 (06): : 2311 - 2320
  • [3] Non-invasive measurement of corneal hydration
    March, WF
    Bauer, NJC
    [J]. JOURNAL OF REFRACTIVE SURGERY, 2001, 17 (02) : S205 - S210
  • [4] A microwave frequency sensor for non-invasive blood-glucose measurement
    Jean, Buford Randall
    Green, Eric C.
    McClung, Melanie J.
    [J]. 2008 IEEE SENSORS APPLICATIONS SYMPOSIUM, 2008, : 4 - 7
  • [5] Bimodal microwave and ultrasound phantoms for non-invasive clinical imaging
    Villa, Enrique
    Arteaga-Marrero, Natalia
    Gonzalez-Fernandez, Javier
    Ruiz-Alzola, Juan
    [J]. SCIENTIFIC REPORTS, 2020, 10 (01)
  • [6] Bimodal microwave and ultrasound phantoms for non-invasive clinical imaging
    Enrique Villa
    Natalia Arteaga-Marrero
    Javier González-Fernández
    Juan Ruiz-Alzola
    [J]. Scientific Reports, 10
  • [7] A NON-INVASIVE METHOD FOR THE MEASUREMENT OF SKIN BLOOD-FLOW
    LARSSON, H
    HELLSTROM, LG
    LINNARSSON, D
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1980, 18 (02) : N15 - N15
  • [8] A Method to Determine Human Skin Heat Capacity Using a Non-Invasive Calorimetric Sensor
    Rodriguez de Rivera, Pedro Jesus
    Rodriguez de Rivera, Miriam
    Socorro, Fabiola
    Rodriguez de Rivera, Manuel
    Marrero Callico, Gustavo
    [J]. SENSORS, 2020, 20 (12) : 1 - 19
  • [9] Non-invasive hemoglobin measurement using optical method
    Ranjith, R.
    Priya, S.
    Dharshini, A. S. Kaviya
    Jeeva, J. B.
    [J]. HELIYON, 2024, 10 (15)
  • [10] A fast, non-invasive method for measurement of cholesterol in skin samples.
    Horsewood, P.
    Evelegh, M. J.
    Dudek, S. M.
    [J]. CLINICAL CHEMISTRY, 2007, 53 (06) : A40 - A40