Determination of E-coli by a Graphene Oxide-Modified Quartz Crystal Microbalance

被引:9
|
作者
Nehra, Anuj [1 ]
Pandey, Khyati [1 ]
Singh, Krishna Pal [1 ]
Ahalawat, Saurabh [2 ]
Joshi, Rajendra Prasad [1 ]
机构
[1] GB Pant Univ Agr & Technol, Biophys Unit, Nanobiosensor Res Lab, CBSH, Pantnagar, Uttarakhand, India
[2] CSIR Cent Bldg Res Inst, Mat Evaluat Res Lab, Roorkee, Uttarakhand, India
关键词
Escherichia coli; graphene oxide nanofilm; quartz crystal microbalance; scanning electron microscopy; ELECTROCHEMICAL BIOSENSORS; CARBON; SENSOR; FOOD; FILM; NANOMATERIALS; SECRETION; CHEMISTRY; ARRAYS; CELLS;
D O I
10.1080/00032719.2016.1253708
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A biosensor for the determination of Escherichia coli using graphene oxide on the crystal (gold) surface was fabricated by the drop cast method. The E. coli sensing characteristics of the biosensor, such as a change in frequency, were examined by exposing the graphene oxide-coated crystal to various functionalization steps at room temperature. Graphene oxide was functionalized by 1-ethyl-3-(3-dimethylaminopropyl) carbodiimide hydrochloride-N-hydroxysuccinimide to covalently conjugate -galactosidase antibodies to recognize microorganisms that produce this material. Frequency changes in the quartz crystal microbalance are dependent on the absorbed/desorbed masses of the analytes on the functional surface of the crystal. In addition, various characterization techniques were optimized for the morphological elemental analysis of the nanocoating that included field emission scanning electron microscopy, scanning electron microscopy, and electron diffraction spectroscopy. This surface was used in a quartz crystal microbalance nanoplatform for the rapid, sensitive, and label-free detection of E. coli. Under optimal conditions, the frequency of quartz crystal microbalance biosensor was directly proportional to the concentration of antigen with a dynamic range from 0.5mgmL(-1) to 5ngmL(-1) and a minimum detection limit of 5ngmL(-1), and a sensitivity of 0.037Hzgml(-1)cm(-1). These results show that the graphene oxide-coated crystal had excellent performance for E. coli. This research reports a simple, inexpensive, and effective highly stable biosensor using graphene oxide as the sensing medium.
引用
收藏
页码:1897 / 1911
页数:15
相关论文
共 50 条
  • [1] Graphene oxide-modified electrodes for sensitive determination of diethylstilbestrol
    Yu, Chunmei
    Ji, Wanyu
    Wang, Yidan
    Bao, Ning
    Gu, Haiying
    NANOTECHNOLOGY, 2013, 24 (11)
  • [2] Deposition and Release of Graphene Oxide Nanomaterials Using a Quartz Crystal Microbalance
    Chowdhury, Indranil
    Duch, Matthew C.
    Mansukhani, Nikhita D.
    Hersam, Mark C.
    Bouchard, Dermont
    ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2014, 48 (02) : 961 - 969
  • [3] Graphene oxide as quartz crystal microbalance sensing layers for detection of formaldehyde
    Yang, Mingqing
    He, Junhui
    SENSORS AND ACTUATORS B-CHEMICAL, 2016, 228 : 486 - 490
  • [4] Graphene oxide thin film coated quartz crystal microbalance for humidity detection
    Yao, Yao
    Chen, Xiangdong
    Guo, Huihui
    Wu, Zuquan
    APPLIED SURFACE SCIENCE, 2011, 257 (17) : 7778 - 7782
  • [5] Determination of microorganisms with a quartz crystal microbalance sensor
    Bao, LL
    Deng, L
    Nie, LH
    Yao, SZ
    Wei, WZ
    ANALYTICA CHIMICA ACTA, 1996, 319 (1-2) : 97 - 101
  • [6] Quartz crystal microbalance ELISA for determination of IgM
    Department of Chemistry and Biology Engineering, Dongguan University of Technology, Dongguan, Guangdong 523106, China
    不详
    Chin. J. Sens. Actuators, 2006, 2 (305-308+331):
  • [7] A humidity sensor based on quartz crystal microbalance using graphene oxide as a sensitive layer
    Jin, Hao
    Tao, Xiang
    Feng, Bin
    Yu, Liyang
    Wang, Demiao
    Dong, Shurong
    Luo, Jikui
    VACUUM, 2017, 140 : 101 - 105
  • [8] Graphene oxide/polystyrene composite nanofibers on quartz crystal microbalance electrode for the ammonia detection
    Jia, Yongtang
    Chen, Lizhu
    Yu, Hui
    Zhang, Yumei
    Dong, Fengchun
    RSC ADVANCES, 2015, 5 (51): : 40620 - 40627
  • [9] A Highly Sensitive Immunosorbent Assay Based on Biotinylated Graphene Oxide and the Quartz Crystal Microbalance
    Deng, Xudong
    Chen, Mengsu
    Fu, Qiang
    Smeets, Niels M. B.
    Xu, Fei
    Zhang, Zhuyuan
    Filipe, Carlos D. M.
    Hoare, Todd
    ACS APPLIED MATERIALS & INTERFACES, 2016, 8 (03) : 1893 - 1902
  • [10] A Mica-Modified Quartz Resonator for a Quartz Crystal Microbalance Study
    Bo Xu
    Hongda Wang
    Ying Wang
    Guoyi Zhu
    Zhuang Li
    Erkang Wang
    Analytical Sciences, 2000, 16 : 1061 - 1063