Understanding nonlinear vibration behaviours in high-power ultrasonic surgical devices

被引:20
|
作者
Mathieson, Andrew [1 ]
Cardoni, Andrea [2 ]
Cerisola, Niccolo [3 ]
Lucas, Margaret [1 ]
机构
[1] Univ Glasgow, Sch Engn, Glasgow, Lanark, Scotland
[2] Pusonics SL, Arganda Del Rey, Spain
[3] Mectron SpA, Carasco, GE, Italy
基金
英国工程与自然科学研究理事会;
关键词
power ultrasonics; ultrasonic surgery; experimental modal analysis; nonlinear behaviour; BONE SURGERY; TEMPERATURE; INSTRUMENTATION; OSTEOTOMY; CHISEL; STRESS; STRAIN;
D O I
10.1098/rspa.2014.0906
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Ultrasonic surgical devices are increasingly used in oral, craniofacial and maxillofacial surgery to cut mineralized tissue, offering the surgeon high accuracy with minimal risk to nerve and vessel tissue. Power ultrasonic devices operate in resonance, requiring their length to be a half-wavelength or multiple-half-wavelength. For bone surgery, devices based on a half-wavelength have seen considerable success, but longer multiple-half-wavelength endoscopic devices have recently been proposed to widen the range of surgeries. To provide context for these developments, some examples of surgical procedures and the associated designs of ultrasonic cutting tips are presented. However, multiple-half-wavelength components, typical of endoscopic devices, have greater potential to exhibit nonlinear dynamic behaviours that have a highly detrimental effect on device performance. Through experimental characterization of the dynamic behaviour of endoscopic devices, it is demonstrated how geometrical features influence nonlinear dynamic responses. Period doubling, a known route to chaotic behaviour, is shown to be significantly influenced by the cutting tip shape, whereas the cutting tip has only a limited effect on Duffing-like responses, particularly the shape of the hysteresis curve, which is important for device stability. These findings underpin design, aiming to pave the way for a new generation of ultrasonic endoscopic surgical devices.
引用
收藏
页数:19
相关论文
共 50 条
  • [31] A high-power traveling wave ultrasonic motor
    Iula, Antonio
    Pappalardo, Massimo
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2006, 53 (07) : 1344 - 1351
  • [32] High-Power Ultrasonic Transducer for Effective Hemolysis
    Wu, Xiaogang
    Chen, Kang
    Hoshijima, Yasushi
    Hariu, Taro
    Yamazaki, Hiroki
    Miyake, Susumu
    Morita, Takeshi
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2022, 69 (01) : 181 - 186
  • [33] How safe is high-power ultrasonic dissection?
    Emam, TA
    Cuschieri, A
    ANNALS OF SURGERY, 2003, 237 (02) : 186 - 191
  • [34] HIGH-POWER, ULTRASONIC FATIGUE TESTING MACHINE
    TIEN, JK
    PURUSHOTHAMAN, S
    ARONS, RM
    WALLACE, JP
    BUCK, O
    MARCUS, HL
    INMAN, RV
    CRANDALL, GJ
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1975, 46 (07): : 840 - 846
  • [35] ULTRASONIC TRANSDUCER FOR HIGH-POWER APPLICATIONS IN GASES
    GALLEGOJUAREZ, JA
    RODRIGUEZCORRAL, G
    GAETEGARRETON, L
    ULTRASONICS, 1978, 16 (06) : 267 - 271
  • [36] HIGH-POWER ULTRASONIC HORNS FOR LIQUID PROCESSING
    LAST, AJ
    BOUCHER, RMG
    IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1972, SU19 (03): : 396 - &
  • [37] MONITORING OF HIGH-POWER ULTRASONIC FATIGUE SYSTEMS
    BAJONS, P
    INTERNATIONAL JOURNAL OF FATIGUE, 1982, 4 (04) : 238 - 238
  • [38] A study into the vibration behaviour of power ultrasonic devices for bone surgery
    Mathieson, Andrew
    Cardoni, Andrea
    INTERNATIONAL CONGRESS ON ULTRASONICS, PROCEEDINGS, 2010, 3 (01): : 767 - 773
  • [39] HCL GETTERING FOR HIGH-POWER SILICON DEVICES
    CHU, CK
    JOHNSON, JE
    RIDING, AJ
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1974, 121 (03) : C88 - C88
  • [40] Passivation in High-Power Si Devices - An Overview
    Grossner, Ulrike
    Mihaila, Andrei
    Vemulapati, Umamaheswara
    GALLIUM NITRIDE AND SILICON CARBIDE POWER TECHNOLOGIES 2, 2012, 50 (03): : 267 - 277