VOLUMETRIC MODULATED ARC THERAPY (VMAT) TREATMENT PLANNING FOR SUPERFICIAL TUMORS

被引:9
|
作者
Zacarias, Albert S. [1 ]
Brown, Mellonie F. [1 ]
Mills, Michael D. [1 ]
机构
[1] Univ Louisville Healthcare, Radiat Therapist Sch, James Graham Brown Canc Ctr, Louisville, KY 40202 USA
关键词
IMRT; VMAT; Surface dose; Optimization; Cranium; RADIOTHERAPY; ACCURACY;
D O I
10.1016/j.meddos.2009.06.006
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The physician's planning objective is often a uniform dose distribution throughout the planning target volume (PTV), including superficial PTVs on or near the surface of a patient's body. Varian's Eclipse treatment planning system uses a progressive resolution optimizer (PRO), version 8.2.23, for Rapid Arc dynamic multileaf collimator volumetric modulated arc therapy planning. Because the PRO is a fast optimizer, optimization convergence errors (OCEs) produce dose nonuniformity in the superficial area of the PTV. We present a postsurgical cranial case demonstrating the recursive method our clinic uses to produce Rapid Arc treatment plans. The initial Rapid Arc treatment plan generated using one 360 arc resulted in substantial dose nonuniformity in the superficial section of the PTV. We demonstrate the use of multiple arcs to produce improved dose uniformity in this region. We also compare the results of this superficial dose compensation method to the results of a recursive method of dose correction that we developed in-house to correct optimization convergence errors in static intensity-modulated radiation therapy treatment plans. The results show that up to 4 arcs may be necessary to provide uniform dose to the surface of the PTV with the current version of the PRO. (C) 2010 American Association of Medical Dosimetrists.
引用
收藏
页码:226 / 229
页数:4
相关论文
共 50 条
  • [1] VALIDATION OF A VOLUMETRIC MODULATED ARC THERAPY (VMAT) TREATMENT PLANNING APPROACH
    Perez-Rozos, A.
    Lobato Munoz, M.
    Jerez Sainz, I.
    Medina, J. A.
    Carrasco Rodriguez, J. L.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S196 - S197
  • [2] Mathematical models for optimal Volumetric Modulated Arc Therapy (VMAT) treatment planning
    Dursun, Pinar
    Taskin, Z. Caner
    Altinel, I. Kuban
    [J]. INTERNATIONAL CONFERENCE ON ENTERPRISE INFORMATION SYSTEMS/INTERNATIONAL CONFERENCE ON PROJECT MANAGEMENT/INTERNATIONAL CONFERENCE ON HEALTH AND SOCIAL CARE INFORMATION SYSTEMS AND TECHNOLOGIES, CENTERIS/PROJMAN / HCIST 2016, 2016, 100 : 644 - 651
  • [3] Comprehensive Study of Parameters for Volumetric Modulated Arc Therapy (VMAT) Treatment Planning
    Talbert, C.
    Fontenot, J.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [4] Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)
    Fathy, S.
    Alalawi, T.
    Al-Dhaibani, N.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 99 (02): : E658 - E658
  • [5] Treatment planning for volumetric modulated arc therapy
    Bedford, James L.
    [J]. MEDICAL PHYSICS, 2009, 36 (11) : 5128 - 5138
  • [6] Treatment planning evaluation for bilateral breast cancer using volumetric modulated arc therapy (VMAT)
    Abdullah, Siham Sabah
    Alabedi, Haydar H.
    Taha, Jenan Hussein
    Abdullah, Khalid
    [J]. REVISTA DE SENOLOGIA Y PATOLOGIA MAMARIA, 2023, 36 (01):
  • [7] Stereotactic Radiation Treatment Planning with Volumetric Modulated Arc Therapy (VMAT): Impact of Duodenal Sparing
    Herman, J. M.
    Kang, J.
    Tuli, R.
    Wolfgang, C.
    Pawlik, T. M.
    Tryggestad, E.
    Smith, K.
    DeWeese, T. L.
    Wong, J.
    Ford, E. C.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S790 - S791
  • [8] COMMISSIONING OF VOLUMETRIC MODULATED ARC THERAPY (VMAT)
    Bedford, James L.
    Warrington, Alan P.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 73 (02): : 537 - 545
  • [9] Validation of Spatial Fractionated GRID Treatment Planning Using Volumetric Modulated Arc Therapy (VMAT)
    Collins, C.
    Schwer, M.
    Keefe, M.
    Wazer, D.
    Cardarelli, G.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [10] Dose planning of a total body irradiation with Volumetric Modulated Arc Therapy (VMAT)
    Springer, A.
    Winkler, E.
    Putz, E.
    Altenburger, A.
    Gruber, R.
    Moser, K.
    Track, C.
    Hammer, J.
    Geinitz, H.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S182 - S182