Projection-reconstruction reduces FOV imaging

被引:16
|
作者
Weiss, S [1 ]
Rasche, V [1 ]
机构
[1] Philips Res Labs, Tech Syst Dept, D-22335 Hamburg, Germany
关键词
projection-reconstruction imaging; local imaging; dynamic imaging;
D O I
10.1016/S0730-725X(98)00212-4
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Reduced field-of-view (rFOV) imaging was introduced recently as a rapid imaging technique that improves temporal resolution while maintaining spatial resolution. It is based on undersampling in k-space and utilizes the fact that the dynamics of an evolving process are often confined to a local area within the full FOV. In the work presented here the reduced FOV approach is applied to projection-reconstruction MRI and compared to the original spin-warp implementation. Results are presented that clearly demonstrate the increased robustness of the projection reconstruction version of rFOV imaging. The technique is successfully applied to an MR-guided biopsy scenario (ex-vivo) and to cine cardiac imaging. Finally an algorithm is proposed that uses the intrinsic advantages of radial k-space sampling to evaluate the projection data to control the adjustment of position and size of the reduced FOV window. (C) 1999 Elsevier Science Inc.