Audit on source strength determination for HDR and PDR 192Ir brachytherapy in Sweden

被引:18
|
作者
Tedgren, Asa Carlsson [1 ]
Grindborg, Jan-Erik [2 ]
机构
[1] Linkoping Univ, Div Med Radiat Phys, Dept Med & Hlth Sci, Fac Hlth Sci, SE-58183 Linkoping, Sweden
[2] Swedish Radiat Protect Author, Stockholm, Sweden
关键词
(192)Ir brachytherapy; dosimetry audit; RAKR;
D O I
10.1016/j.radonc.2007.12.008
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: To investigate the status of source strength determination in terms of reference air kerma rate (RAKR) for HDR and PDR (192)Ir brachytherapy in Sweden. Materials and methods: RAKR was determined in each of the 14 Swedish afterloaders, using calibrated equipment from the Swedish Secondary Standard Dosimetry Laboratory. Results: Values of RAKR from the external audit, the hospitals and vendors agreed within the uncertainty limits guaranteed by the vendors. Conclusions: The accuracy in RAKR determination has increased over the last years as a result of increased availability of interpolation standards for HDR (192)Ir and the increased use of robust well-type ion chambers designed for brachytherapy. It is recommended to establish a ratio between the RAKR value from own measurements at the hospital and that of the vendor since such a ratio embeds constant systematic differences due to e.g. varying traceability and therefore has the potential of being less uncertain than the RAKR alone. Traceability to primary standards for HDR (192)Ir sources will in the future significantly decrease the uncertainty in RAKR of (192)Ir brachytherapy. (C) 2007 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:126 / 130
页数:5
相关论文
共 50 条
  • [1] EVALUATION OF SCATTER CONTRIBUTION AND DISTANCE ERROR BY ITERATIVE METHODS FOR STRENGTH DETERMINATION OF HDR 192IR BRACHYTHERAPY SOURCE
    Kumar, Sudhir
    Srinivasan, Panchapakesan
    Sharma, Sunil D.
    Subbaiah, Kamatam V.
    Mayya, Yelia S.
    [J]. MEDICAL DOSIMETRY, 2010, 35 (03) : 230 - 237
  • [2] Gafchromic film dosimetry of a new HDR 192Ir brachytherapy source
    Ayoobian, Navid
    Asl, Akbar Sarabi
    Poorbaygi, Hosein
    Javanshir, Mohammad Reza
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (02): : 194 - 205
  • [3] Determination of the dose rate around a HDR 192Ir brachytherapy source with the microDiamond and the microSilicon detector
    Rossi, Giulio
    Failing, Thomas
    Gainey, Mark
    Kollefrath, Michael
    Hensley, Frank
    Zink, Klemens
    Baltas, Dimos
    [J]. ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2023, 33 (04): : 463 - 478
  • [4] Determination of transit dose profile for a 192Ir HDR source
    Fonseca, G. P.
    Rubo, R. A.
    Minamisawa, R. A.
    dos Santos, G. R.
    Antunes, P. C. G.
    Yoriyaz, H.
    [J]. MEDICAL PHYSICS, 2013, 40 (05)
  • [5] Calibration of MOSFET detectors for absolute dosimetry with an 192IR HDR brachytherapy source
    Zilio, V
    Joneja, O.
    Popowski, Y.
    Rosenfeld, A.
    Chawla, R.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2006, 182 : 175 - 175
  • [6] Energy dependence of a radiophotoluminescent glass dosimeter for HDR 192Ir brachytherapy source
    Hashimoto, Shimpei
    Nakajima, Yujiro
    Kadoya, Noriyuki
    Abe, Kota
    Karasawa, Katsuyuki
    [J]. MEDICAL PHYSICS, 2019, 46 (02) : 964 - 972
  • [7] Developments of a TLD mailed system for remote dosimetry audit for 192Ir HDR and PDR sources
    Roue, Amelie
    Venselaar, Jack L. M.
    Ferreira, Ivaldo H.
    Bridier, Andre
    Van Dam, Jan
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 83 (01) : 86 - 93
  • [8] Dosimetry of a two new IR-192 source models for HDR and PDR brachytherapy
    Granero, D.
    Perez-Calatayud, J.
    Gimeno, J.
    Gonzalez, V.
    Ballester, F.
    Casal, E.
    Pujades, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S147 - S147
  • [9] Determination of contributions of scatter and distance error to the source strength of Ir-192 HDR brachytherapy source
    BONDEL, Shwetha
    RAVIKUMAR, Manickam
    [J]. POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2016, 22 (03) : 55 - 59
  • [10] On source models for 192Ir HDR brachytherapy dosimetry using model based algorithms
    Pantelis, Evaggelos
    Zourari, Kyveli
    Zoros, Emmanouil
    Lahanas, Vasileios
    Karaiskos, Pantelis
    Papagiannis, Panagiotis
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (11): : 4235 - 4246