Continuous-wave near-infrared spectroscopy is not related to brain tissue oxygen tension

被引:11
|
作者
Kerz, Thomas [1 ]
Beyer, Christian [1 ]
Huthmann, Alexandra [3 ]
Kalasauskas, Darius [1 ]
Amr, Amr Nimer [1 ]
Boor, Stephan [2 ]
Welschehold, Stefan [1 ]
机构
[1] Johannes Gutenberg Univ Mainz, Univ Med Ctr, Dept Neurosurg, Langenbeckstr 1, D-55131 Mainz, Germany
[2] Johannes Gutenberg Univ Mainz, Inst Neuroradiol, Univ Med Ctr, Mainz, Germany
[3] Ludmillenstift, Meppen, Germany
关键词
Near-infrared spectroscopy; Brain oxygen tension; Brain trauma; Subarachnoid haemorrhage; Brain death; CEREBRAL-BLOOD-FLOW; VENOUS OXIMETRY; SATURATION; VALIDATION; MONITOR;
D O I
10.1007/s10877-015-9755-y
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Near-infrared spectroscopy (NIRS) has gained acceptance for cerebral monitoring, especially during cardiac surgery, though there are few data showing its validity. We therefore aimed to correlate invasive brain tissue oxygen measurements (PtiO2) with the corresponding NIRS-values (regional oxygen saturation, rSO2). We also studied whether NIRS was able to detect ischemic events, defined as a PtiO2-value of < 15 mmHg. Eleven patients were studied with invasive brain tissue oxygen monitoring and continuous-wave NIRS. PtiO2-correlation with corresponding NIRS-values was calculated. We found no correlation between PtiO2- and NIRS-readings. Measurement of rSO2 was no better than flipping a coin in the detection of cerebral ischemia when a commonly agreed ischemic PtiO2 cut-off value of < 15 mmHg was chosen. Continuous-wave-NIRS was unable to reliably detect ischemic cerebral episodes, defined as a PtiO2 value < 15 mmHg. Displayed NIRS-values did not correlate with invasively measured PtiO2-values. CW-NIRS should not be used for the detection of cerebral ischemia.
引用
收藏
页码:641 / 647
页数:7
相关论文
共 50 条
  • [1] Continuous-wave near-infrared spectroscopy is not related to brain tissue oxygen tension
    Thomas Kerz
    Christian Beyer
    Alexandra Huthmann
    Darius Kalasauskas
    Amr Nimer Amr
    Stephan Boor
    Stefan Welschehold
    Journal of Clinical Monitoring and Computing, 2016, 30 : 641 - 647
  • [2] Brain tissue oxygen evaluation by wireless near-infrared spectroscopy
    Wang, Che-Chuan
    Kuo, Jinn-Rung
    Chen, Yu-Chih
    Chio, Chung-Ching
    Wang, Jhi-Joung
    Lin, Bor-Shyh
    JOURNAL OF SURGICAL RESEARCH, 2016, 200 (02) : 669 - 675
  • [3] Algorithm Establishment with Model Experiment for Continuous-wave Functional Near-infrared Spectroscopy
    Li, Ting
    Zhang, Zhili
    Zheng, Yi
    APPLIED MECHANICS AND MECHANICAL ENGINEERING II, PTS 1 AND 2, 2012, 138-139 : 553 - +
  • [4] In-vivo quantitative measurement of tissue oxygen saturation of human webbing using a transmission type continuous-wave near-infrared spectroscopy
    Aizimu, Tuerxun
    Adachi, Makoto
    Nakano, Kazuya
    Ohnishi, Takashi
    Nakaguchi, Toshiya
    Takahashi, Nozomi
    Nakada, Taka-aki
    Oda, Shigeto
    Haneishi, Hideaki
    OPTICAL DIAGNOSTICS AND SENSING XVIII: TOWARD POINT-OF-CARE DIAGNOSTICS, 2018, 10501
  • [5] A broadband multi-distance approach to measure tissue oxygen saturation with continuous wave near-infrared spectroscopy
    Kovacsova, Zuzana
    Bale, Gemma
    Veesa, Joshua Deepak
    Dehghani, Hamid
    Tachtsidis, Ilias
    DIFFUSE OPTICAL SPECTROSCOPY AND IMAGING VII, 2019, 11074
  • [6] Continuous-wave near-infrared spectroscopy using path length-independent hypoxia normalization
    Kennan, RP
    Behar, KL
    JOURNAL OF BIOMEDICAL OPTICS, 2002, 7 (02) : 228 - 235
  • [7] Correlation of brain tissue oxygen tension with cerebral near-infrared spectroscopy and mixed venous oxygen saturation during extracorporeal membrane oxygenation
    Tyree, Kreangkai
    Tyree, Melissa
    DiGeronimo, Robert
    PERFUSION-UK, 2009, 24 (05): : 325 - 331
  • [8] Continuous-wave near-infrared topography system and its clinical application
    Xu, Canxing
    Wang, Daifa
    Zhang, Yonghong
    Bai, Jing
    2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2005, : 196 - 198
  • [9] Ocular hazards of tunable continuous-wave near-infrared laser sources
    Lund, DJ
    Edsall, PR
    Fuller, DR
    Hoxie, SW
    LASER-INFLICTED EYE INJURIES: EPIDEMIOLOGY, PREVENTION, AND TREATMENT, PROCEEDINGS OF, 1996, 2674 : 53 - 61
  • [10] Ocular hazards of tunable continuous-wave near-infrared laser sources
    Lund, DJ
    Edsall, PR
    Fuller, DR
    Hoxie, SW
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1996, 37 (03) : 3580 - 3580