Accuracy of CoaguChek XS for Point-of-Care Antithrombotic Monitoring in Children with Heart Disease

被引:0
|
作者
Moon, Ju Ryoung [2 ]
Jeong, Soo In [1 ]
Huh, June [1 ]
Lee, Heung Jae [1 ]
Park, Pyo Won [3 ]
Kang, I-Seok [1 ]
机构
[1] Sungkyunkwan Univ, Sch Med, Samsung Med Ctr, Dept Pediat, Seoul 135710, South Korea
[2] Sungkyunkwan Univ, Sch Med, Samsung Med Ctr, Cardiac & Vasc Ctr, Seoul 135710, South Korea
[3] Sungkyunkwan Univ, Sch Med, Samsung Med Ctr, Dept Thorac & Cardiovasc Surg, Seoul 135710, South Korea
来源
关键词
antithrombotic monitoring; point-of-care assay; prothrombin time; ORAL ANTICOAGULANT-THERAPY; SELF-MANAGEMENT; INR;
D O I
暂无
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
The CoaguChek XS international normalized ratio (INR) assay was compared to INR assay by a standard laboratory method in children with heart disease on anticoagulant therapy. The data comprised 120 pairs of INR values for 42 patients (age <16 yr) who attended a cardiology clinic between 1 May 2007 and 30 January 2008. Parallel INR assays by the CoaguChek XS and the standard method were performed within 1 hr by a single qualified technician and the paired results were evaluated by linear regression and Bland-Altman analysis. The mean difference in the INR values was -0.08 +/- 0.04 units (p = 0.63); the difference between the two results was consistently <0.5 INR units. The slope of the regression line was 0.98 (95% CI: 0.96 to 1.01) and the y-intercept was 0.014 (95% CI: -0.01 to 0.04). In the Bland-Altman analysis, the mean difference in INR between the two methods was 0.08 units and values for 99.4% of the patients fell within the limit of agreement (-0.17 to 0.28 units). In summary, INR assays in children by the CoaguChek XS device are as accurate as the standard method, but faster and more convenient.
引用
收藏
页码:247 / 251
页数:5
相关论文
共 50 条
  • [1] Accuracy of the point-of-care coagulometer CoaguChek XS in the hands of patients
    Nagler, M.
    Raddatz-Mueller, P.
    Schmid, P.
    Bachmann, L. M.
    Wuillemin, W. A.
    JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2013, 11 (01) : 197 - 199
  • [2] Accuracy of the CoaguChek XS for point-of-care international normalized ratio (INR) measurement in children requiring warfarin
    Bauman, Mary E.
    Black, Karina L.
    Massicotte, Mary P.
    Bauman, Michelle L.
    Kuhle, Stefan
    Howlett-Clyne, Susan
    Cembrowski, George S.
    Bajzar, Laszlo
    THROMBOSIS AND HAEMOSTASIS, 2008, 99 (06) : 1097 - 1103
  • [3] Point-of-care antithrombotic monitoring in children
    Newall, Fiona
    Bauman, Mary
    THROMBOSIS RESEARCH, 2006, 118 (01) : 113 - 121
  • [4] Accuracy of capillary whole blood international normalized ratio on the CoaguChek S, CoaguChek XS, and i-STAT 1 point-of-care analyzers
    Karon, Brad S.
    McBane, Robert D., II
    Chaudhry, Rajeev
    Beyer, Lisa K.
    Santrach, Paula J.
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2008, 130 (01) : 88 - 92
  • [5] Assessment of the CoaguChek-XS portable prothrombin time point-of-care analyzer for horses
    Berlin, Noa
    Kelmer, Efrat
    Segev, Gilad
    Aroch, Itamar
    Kelmer, Gal
    JOURNAL OF VETERINARY DIAGNOSTIC INVESTIGATION, 2019, 31 (03) : 448 - 452
  • [6] The reliability of point-of-care (POC) prothrombin time testing -: A comparison of CoaguChek XS® INR measurements with hospital laboratory monitoring
    Ryan, Fiona
    O'Shea, Susan
    Ortel, Thomas L.
    Byrne, Stephen
    BLOOD, 2007, 110 (11) : 296A - 296A
  • [7] Point-of-care monitoring of vitamin K-antagonists: validation of CoaguChek XS test strips with International Standard thromboplastin
    van den Besselaar, Anton M. H. P.
    Pequeriaux, Nathalie C. V.
    Ebben, Marij
    van der Feest, Joke
    de Jong, Kerst
    Ganzeboom, Maria B. J.
    van Ooijen, Jetty
    Postema, Ferna
    Witteveen, Evelina
    van der Meer, Felix J. M.
    JOURNAL OF CLINICAL PATHOLOGY, 2012, 65 (11) : 1031 - 1035
  • [8] The reliability of point-of-care prothrombin time testing. A comparison of CoaguChek S® and XS® INR measurements with hospital laboratory monitoring
    Ryan, F.
    O'Shea, S.
    Byrne, S.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2010, 32 (01) : E26 - E33
  • [9] Evaluation of the CoaguChek S, CoaguChek XS and i-STAT for Point of Care INR testing
    Beyer, L. K.
    Wockenfus, A.
    McBane, R.
    Chaudhry, R.
    Karon, B.
    Santrach, P.
    CLINICAL CHEMISTRY, 2008, 54 (06) : A198 - A199
  • [10] Comparison of point-of-care analysis using Coaguchek XS and standard laboratory-measured prothrombin time in dogs
    Newbould, A. R.
    Norman, E. J.
    NEW ZEALAND VETERINARY JOURNAL, 2013, 61 (01) : 18 - 24