Multi-Parametric Standardization of Fluorescence Imaging Systems Based on a Composite Phantom

被引:18
|
作者
Gorpas, Dimitris [1 ,2 ]
Koch, Maximilian [1 ,2 ]
Anastasopoulou, Maria [1 ,2 ]
Bozhko, Dmitry [1 ,2 ]
Klemm, Uwe [1 ,2 ]
Nieberler, Markus [3 ]
Ntziachristos, Vasilis [2 ,4 ]
机构
[1] Helmholtz Zentrum Munchen, Inst Biol & Med Imaging, Munich, Germany
[2] Tech Univ Munich, Chair Biol Imaging & TranslaTUM, Munich, Germany
[3] Tech Univ Munich, Univ Hosp Rechts Isar, Dept Oral & Maxillofacial Surg, Munich, Germany
[4] Helmholtz Zentrum Munchen, Inst Biol & Med Imaging, D-85764 Neuherberg, Germany
关键词
Fluorescence molecular imaging; composite phantom; standardization; benchmarking; data referencing;
D O I
10.1109/TBME.2019.2910733
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: Fluorescence molecular imaging (FMI) has emerged as a promising tool for surgical guidance in oncology, with one of the few remaining challenges being the ability to offer quality control and data referencing. This paper investigates the use of a novel composite phantom to correct and benchmark FMI systems. Methods: This paper extends on previous work by describing a phantom design that can provide a more complete assessment of FMI systems through quantification of dynamic range and determination of spatial illumination patterns for both reflectance and fluorescence imaging. Various performance metrics are combined into a robust and descriptive "system benchmarking score," enabling not only the comprehensive comparison of different systems, but also for the first time, correction of the acquired data. Results: We show that systems developed for targeted fluorescence imaging can achieve benchmarking scores of up to 70%, while clinically available systems optimized for indocyanine green are limited to 50%, mostly due to greater leakage of ambient and excitation illumination and lower resolution. The image uniformity can also be approximated and employed for image flat-fielding, an important milestone toward data referencing. In addition, we demonstrate composite phantom use in assessing the performance of a surgical microscope and of a raster-scan imaging system. Conclusion: Our results suggest that the new phantom has the potential to support high-fidelity FMI through benchmarking and image correction. Significance: Standardization of the FMI is a necessary process for establishing good imaging practices in clinical environments and for enabling high-fidelity imaging across patients and multi-center imaging studies.
引用
收藏
页码:185 / 192
页数:8
相关论文
共 50 条
  • [1] MULTI-PARAMETRIC FLUORESCENCE CYTOPHOTOMETRY
    FUKUDA, M
    CELL STRUCTURE AND FUNCTION, 1983, 8 (04) : 406 - 406
  • [2] High-throughput, multi-parametric, and correlative fluorescence lifetime imaging
    Poudel, Chetan
    Mela, Ioanna
    Kaminski, Clemens F.
    METHODS AND APPLICATIONS IN FLUORESCENCE, 2020, 8 (02):
  • [3] Multi-parametric Retinal Microvascular Functional Perfusion Imaging Based on Dynamic Fundus Fluorescence Angiography
    Wang D.
    Wang Q.
    Su Q.
    Wang S.
    Jian Z.
    Li J.
    Ye F.
    Hou Y.
    Wan M.
    IEEE Transactions on Biomedical Engineering, 2024, 71 (11) : 1 - 10
  • [4] Multiparameter solid phantom for fluorescence imaging standardization
    Anastasopoulou, Maria
    Gorpas, Dimitris
    Koch, Maximilian
    Garcia-Allende, Pilar Beatriz
    Klemm, Uwe
    Karlas, Angelos
    Ntziachristos, Vasilis
    CLINICAL AND PRECLINICAL OPTICAL DIAGNOSTICS, 2017, 10411
  • [5] Multi-parametric functional imaging for tumour classification
    Van Elmpt, W.
    Even, A.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S64 - S64
  • [6] Standardization phantom for intra-operative fluorescence molecular imaging
    Gorpas, Dimitris
    Anastasopoulou, Maria
    Koch, Maximilian
    Klemm, Uwe
    Nieberler, Markus
    Ntziachristos, Vasilis
    MEDICAL LASER APPLICATIONS AND LASER-TISSUE INTERACTIONS IX, 2019, 11079
  • [7] Multi-parametric imaging of cell heterogeneity in apoptosis analysis
    Vorobjev, Ivan A.
    Barteneva, Natasha S.
    METHODS, 2017, 112 : 105 - 123
  • [8] MULTI-parametric MR imaging with fLEXible design (MULTIPLEX)
    Ye, Yongquan
    Lyu, Jingyuan
    Hu, Yichen
    Zhang, Zhongqi
    Xu, Jian
    Zhang, Weiguo
    MAGNETIC RESONANCE IN MEDICINE, 2022, 87 (02) : 658 - 673
  • [10] Standardization of Fluorescence Molecular Imaging Systems
    Gorpas, Dimitris
    Koch, Maximilian
    Anastasopoulou, Maria
    Klemm, Uwe
    Ntziachristos, Vasilis
    NOVEL BIOPHOTONICS TECHNIQUES AND APPLICATIONS IV, 2017, 10413