Radiation safety survey on a flattening filter-free medical accelerator

被引:17
|
作者
Vassiliev, O. N. [1 ]
Titt, U. [1 ]
Kry, S. F. [1 ]
Mohan, R. [1 ]
Gillin, M. T. [1 ]
机构
[1] Univ Texas, MD Anderson Canc Ctr, Houston, TX 77030 USA
关键词
D O I
10.1093/rpd/ncm179
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Dose rates at several locations outside a treatment room were measured for 6 and 18 MV photon beams from a Varian Clinac 21EX accelerator operated with and without a flattening filter. Also, dose rates in the treatment room due to activation were measured at 18 MV. An analysis of the measured data is presented. The results suggest that substantial reduction in doses outside the treatment room and lower activation can be achieved with a flattening-filter free accelerator.
引用
收藏
页码:187 / 190
页数:4
相关论文
共 50 条
  • [1] Treatment vault shielding for a flattening filter-free medical linear accelerator
    Kry, Stephen F.
    Howell, Rebecca M.
    Polf, Jerimy
    Mohan, Radhe
    Vassiliev, Oleg N.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (05): : 1265 - 1273
  • [2] Pion effects in flattening filter-free radiation beams
    Corns, Robert A.
    Huang, Vicky W.
    Thomas, Steven D.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (06): : 376 - 385
  • [3] Survey on utilization of flattening filter-free photon beams in Japan
    Kodama, Takumi
    Yasui, Keisuke
    Nishioka, Shie
    Miyaura, Kazunori
    Takakura, Toru
    Katayose, Tetsurou
    Nakamura, Mitsuhiro
    [J]. JOURNAL OF RADIATION RESEARCH, 2021, 62 (04) : 726 - 734
  • [4] A Monte Carlo study of a flattening filter-free linear accelerator verified with measurements
    Dalaryd, Marten
    Kragl, Gabriele
    Ceberg, Crister
    Georg, Dietmar
    McClean, Brendan
    af Wetterstedt, Sacha
    Wieslander, Elinore
    Knoos, Tommy
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (23): : 7333 - 7344
  • [5] On flattening filter-free portal dosimetry
    Pardo, Eduardo
    Castro Novais, Juan
    Molina Lopez, Maria Yolanda
    Ruiz Maqueda, Sheila
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (04): : 132 - 145
  • [6] The comparison of properties for radiotherapy with flattening filter-free and flattening filter beam
    Gu, J. H.
    Won, H. S.
    Hong, J. W.
    Chang, N. J.
    Park, J. H.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S978 - S978
  • [7] Monte Carlo study on a flattening filter-free 18-MV photon beam of a medical linear accelerator
    Mesbahi, Asghar
    Nejad, Farshad Seyed
    [J]. RADIATION MEDICINE, 2008, 26 (06): : 331 - 336
  • [8] RADIATION SAFETY SURVEY OF A FILTER FREE ELEKTA LINEAR ACCELERATOR
    Phillips, J.
    George, L.
    Ramtohul, M.
    Cashmore, J.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S559 - S559
  • [9] Monte Carlo study on a flattening filter-free 18-MV photon beam of a medical linear accelerator
    Asghar Mesbahi
    Farshad Seyed Nejad
    [J]. Radiation Medicine, 2008, 26 : 331 - 336
  • [10] Monte Carlo study of photon fields from a flattening filter-free clinical accelerator
    Vassiliev, ON
    Titt, U
    Kry, SF
    Pönisch, F
    Gillin, MT
    Mohan, R
    [J]. MEDICAL PHYSICS, 2006, 33 (04) : 820 - 827