Combined use of a true-bipolar sensing implantable cardioverter defibrillator in a patient having a prior implantable spinal cord stimulator for intractable pain

被引:25
|
作者
Monahan, K
Casavant, D
Rasmussen, C
Hallet, N
机构
[1] Beth Israel Deaconess Med Ctr, Boston, MA 02215 USA
[2] Medtron Inc, Minneapolis, MN USA
来源
关键词
implantable cardioverter defibrillator; spinal cord stimulator;
D O I
10.1111/j.1540-8159.1998.tb00045.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Reported is a case involving a patient with a previously implanted spinal cord stimulator (SCS) who presented for an implantable cardioverter defibrillator (ICD), The SCS device was located in the left lower abdominal quadrant with a stimulation electrode array placed on the dorsal aspect of the spinal cord at the T-11 thoracic level. interaction testing demonstrated that the bipolar sensing transvenous ICD system (Medtronic 7221Cx PCD) did not detect the stimulator's output at burst rates ranging from 20-130 pulses/s, even with the ICD set to its maximum sensitivity of 0.15 m V and the stimulator programmed to the highest patient tolerated output combinations of 5 V, 0.45 ms in the bipolar configuration and 3 V, 0.45 ms in the unipolar (i.e., case-electrode) configuration.
引用
收藏
页码:2669 / 2672
页数:4
相关论文
共 5 条
  • [1] Device-Device Interference of a Bipolar Spinal Cord StimulatorWith a Bipolar Implantable Cardioverter- Defibrillator Device- Device Interference of a Bipolar Spinal Cord StimulatorWith a Bipolar Implantable Cardioverter- Defibrillator
    Qu, Wenchun
    Muir, Jeffery J.
    Osborne, Michael D.
    NEUROMODULATION, 2014, 17 (02): : 202 - 204
  • [2] Spinal cord stimulation in a patient with an implantable cardioverter defibrillator for managing chronic chest pain: a case report and literature review
    Orlando, Porras Bueno Cristian
    William, Rangel Jaimes German
    Anamaria, Camargo Cardenas
    Ximena, Cediel Carrillo
    Milena, Berdugo Pereira Eliana
    PAIN MANAGEMENT, 2024,
  • [3] Potential device interaction of a dual chamber implantable cardioverter defibrillator in a patient with continuous spinal cord stimulation
    Schimpf, R
    Wolpert, C
    Herwig, S
    Schneider, C
    Esmailzadeh, B
    Lüderitz, B
    EUROPACE, 2003, 5 (04): : 397 - 402
  • [4] Myopotential oversensing notified by Lead Integrity Alert in a patient with implantable cardioverter defibrillator with a dedicated bipolar epicardial sensing lead
    Yazaki, Kyoichiro
    Ajiro, Yoichi
    Hasegawa, Shun
    Watanabe, Masahiro
    Tsukamoto, Kei
    Mizobuchi, Keiko
    Saito, Takashi
    Mori, Fumiaki
    Iwade, Kazunori
    CLINICAL CASE REPORTS, 2016, 4 (12): : 1091 - 1095
  • [5] Simultaneous use of implantable cardioverter-defibrillator in a patient with a preexisting deep brain stimulator: A proposed protocol of implantation to avoid dangerous interactions
    Chmielewska-Michalak, Lidia
    Migaj, Jacek
    Katarzynska-Szymanska, Agnieszka
    Cieplucha, Aleksandra
    Grymuza, Maciej
    Ochotny, Romuald
    Lesiak, Maciej
    Mitkowski, Przemyslaw
    KARDIOLOGIA POLSKA, 2022, 80 (09) : 949 - 951