The response of a radiophotoluminescent glass dosimeter in megavoltage photon and electron beams

被引:16
|
作者
Araki, Fujio [1 ]
Ohno, Takeshi [1 ]
机构
[1] Kumamoto Univ, Fac Life Sci, Dept Hlth Sci, Chuo Ku, Kumamoto 8620976, Japan
关键词
radiophotoluminescent glass dosimeter; beam quality dependence; Monte Carlo simulation; Burlin's general cavity theory; MONTE-CARLO SIMULATIONS; QUALITY DEPENDENCE; ROD DETECTOR; LIF; TLD;
D O I
10.1118/1.4901639
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: This study investigated the response of a radiophotoluminescent glass dosimeter (RGD) in megavoltage photon and electron beams. Methods: The RGD response was compared with ion chamber measurements for 4-18 MV photons and 6-20 MeV electrons in plastic water phantoms. The response was also calculated via Monte Carlo (MC) simulations with EGSnrc/egs_chamber and Cavity user-codes, respectively. In addition, the response of the RGD cavity was analyzed as a function of field sizes and depths according to Burlin's general cavity theory. The perturbation correction factor, P-Q, in the RGD cavity was also estimated from MC simulations for photon and electron beams. Results: The calculated and measured RGD energy response at reference conditions with a 10 x 10 cm(2) field and 10 cm depth in photons was lower by up to 2.5% with increasing energy. The variation in RGD response in the field size range of 5 x 5 cm(2) to 20 x 20 cm(2) was 3.9% and 0.7%, at 10 cm depth for 4 and 18 MV, respectively. The depth dependence of the RGD response was constant within 1% for energies above 6 MV but it increased by 2.6% and 1.6% for a large (20 x 20 cm(2)) field at 4 and 6 MV, respectively. The dose contributions from photon interactions (1 - d) in the RGD cavity, according to Burlin's cavity theory, decreased with increasing energy and decreasing field size. The variation in (1 - d) between field sizes became larger with increasing depth for the lower energies of 4 and 6 MV. P-Q for the RGD cavity was almost constant between 0.96 and 0.97 at 10 MV energies and above. Meanwhile, P-Q depends strongly on field size and depth for 4 and 6 MV photons. In electron beams, the RGD response at a reference depth, d(ref), varied by less than 1% over the electron energy range but was on average 4% lower than the response for 6 MV photons. Conclusions: The RGD response for photon beams depends on both (1 - d) and perturbation effects in the RGD cavity. Therefore, it is difficult to predict the energy dependence of RGD response by Burlin's theory and it is recommended to directly measure RGD response or use the MC-calculated RGD response, regarding the practical use. The response for electron beams decreased rapidly at a depth beyond d(ref) for lower mean electron energies <3 MeV and in contrast P-Q increased. (C) 2014 American Association of Physicists in Medicine.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] The Beam Quality Dependence of a Radiophotoluminescent Glass Dosimeter in Megavoltage Photon and Electron Beams
    Araki, F.
    Ohno, T.
    Yoshiyama, F.
    Kohno, T.
    Wulff, J.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)
  • [2] Impact of transverse magnetic fields on dose response of a radiophotoluminescent glass dosimeter in megavoltage photon beams
    Ito, Shotaro
    Araki, Fujio
    Ohno, Takeshi
    [J]. MEDICAL PHYSICS, 2020, 47 (04) : 1995 - 2004
  • [3] Evaluation of the dosimetric characteristics of a radiophotoluminescent glass dosimeter for high-energy photon and electron beams in the field of radiotherapy
    Son, Kihong
    Jung, Haijo
    Shin, Sang Hun
    Lee, Hyun-Ho
    Kim, Mi-Sook
    Ji, Young Hoon
    Kim, Kum Bae
    [J]. RADIATION MEASUREMENTS, 2011, 46 (10) : 1117 - 1122
  • [4] Investigation of the dosimetric characteristics of radiophotoluminescent glass dosimeter for high-energy photon beams
    Oonsiri, Puntiwa
    Kingkaew, Sakda
    Vannavijit, Chulee
    Suriyapee, Sivalee
    [J]. JOURNAL OF RADIATION RESEARCH AND APPLIED SCIENCES, 2019, 12 (01) : 65 - 71
  • [5] DOSIMETRIC PROPERTIES OF RADIOPHOTOLUMINESCENT GLASS DOSIMETER IN HIGH-ENERGY ELECTRON BEAMS
    Son, K.
    Jung, H.
    Shin, S. H.
    Lee, H. H.
    Kim, M. S.
    Ji, Y. H.
    Kim, B.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S407 - S407
  • [6] Response of a Radiophotoluminescent Glass Dosimeter for Diagnostic Kilovoltage X-Ray Beams
    Nakatake, C.
    Araki, F.
    Ohno, T.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E552 - E552
  • [7] Energy response of radiophotoluminescent glass dosimeter for diagnostic kilovoltage x-ray beams
    Nakatake, Chihiro
    Araki, Fujio
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 82 : 144 - 149
  • [8] Response of a NanoDot OSLD System in Megavoltage Photon and Electron Beams
    Hoshida, K.
    Araki, F.
    Ohno, T.
    Shimohigashi, Y.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E314 - E314
  • [9] Response of a Radiophotoluminescent Glass Dosimeter Under Transverse Magnetic Fields
    Ohno, T.
    Araki, F.
    Ito, S.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E413 - E413
  • [10] Difference in the relative response of the alanine dosimeter to megavoltage x-ray and electron beams
    Anton, M.
    Kapsch, R-P
    Krauss, A.
    von Voigts-Rhetz, P.
    Zink, K.
    McEwen, M.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (10): : 3259 - 3282