Efficacy of the RADPAD Protection Drape in Reducing Operators' Radiation Exposure in the Catheterization Laboratory: A Sham-Controlled Randomized Trial

被引:48
|
作者
Vlastra, Wieneke [1 ]
Delewi, Ronak [1 ]
Sjauw, Krischan D. [1 ]
Beijk, Marcel A. [1 ]
Claessen, Bimmer E. [1 ]
Streekstra, Geert J. [2 ]
Bekker, Robbert J. [1 ]
van Hattum, Juliette C. [1 ]
Wykrzykowska, Joanna J. [1 ]
Vis, Marije M. [1 ]
Koch, Karel T. [1 ]
de Winter, Robbert J. [1 ]
Piek, Jan J. [1 ]
Henriques, Jose P. S. [1 ]
机构
[1] Univ Amsterdam, Acad Med Ctr, Heart Ctr, Amsterdam, Netherlands
[2] Univ Amsterdam, Acad Med Ctr, Dept Biomed Engn & Phys & Radiol & Nucl Med, Amsterdam, Netherlands
关键词
fluoroscopy; percutaneous coronary intervention; radiation; radiation protection; risk; PERCUTANEOUS CORONARY INTERVENTIONS; DOSE REDUCTION; RADIOLOGY; SHIELD; CARDIOLOGISTS; ANGIOGRAPHY; PATIENT; TUMORS; STAFF; RISK;
D O I
10.1161/CIRCINTERVENTIONS.117.006058
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Interventional cardiologists are increasingly exposed to radiation-induced diseases like cataract and the stochastic risk of left-sided brain tumors. The RADPAD is a sterile, disposable, lead-free shield placed on the patient with the aim to minimize operator-received scatter radiation. The objective of the trial was to examine the RADPAD's efficacy in a real-world situation. Methods and Results In the current, double-blind, sham-controlled, all-comer trial, patients undergoing diagnostic catheterization or percutaneous coronary interventions were randomized in a 1:1:1 ratio to a radiation absorbing shield (RADPAD), standard treatment (NOPAD), or a sham shield (SHAMPAD). The sham shield allowed testing for shield-induced radiation behavior. The primary outcome was the difference in relative exposure of the primary operator between the RADPAD and NOPAD arms and was defined as the ratio between operator's exposure (E in mu Sv) and patient exposure (dose area product in mGycm(2)), measured per procedure. A total of 766 consecutive coronary procedures were randomized to the use of RADPAD (N=255), NOPAD (N=255), or SHAMPAD (N=256). The use of RADPAD was associated with a 20% reduction in relative operator exposure compared with that of NOPAD (P=0.01) and a 44% relative exposure reduction compared with the use of a SHAMPAD (P<0.001). Use of the SHAMPAD was associated with a 43% higher relative radiation exposure than procedures with NOPAD (P=0.009). Conclusions In clinical daily practice, the standard use of the RADPAD radiation shield reduced operator radiation exposure compared with procedures with NOPAD or SHAMPAD. This study supports the routine use of RADPAD in the catheterization laboratory. Clinical Trial Registration URL: https://www.clinicaltrials.gov. Unique identifier: NCT03139968.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] ASSESSING THE EFFICACY OF RADPAD PROTECTION DRAPE IN REDUCING RADIATION EXPOSURE TO OPERATORS IN THE CATHETERIZATION LABORATORY: A SYSTEMATIC REVIEW AND METAANALYSIS
    Khanal, Resha
    Bahar, Abdul Rasheed
    Goru, Rohit
    Basit, Salman Abdul
    Umer, Ahmed Muaaz
    Haider, Mobeen Z.
    Shafiq, Aimen
    Ishaq, Syed
    Yadav, Ritu
    Almas, Talal
    Hamza, Mohammad
    Sattar, Yasar
    Alraies, M. Chadi
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2024, 83 (13) : 1132 - 1132
  • [2] Assessing the Efficacy of RADPAD Protection Drape in Reducing Radiation Exposure to Operators in the Cardiac Catheterization Laboratory: A Systematic Review and MetaAnalysis
    Bahar, Abdul Rasheed
    Khanal, Resha
    Hamza, Mohammad
    Goru, Rohit K.
    Shafiq, Aimen
    Haider, Mobeen Z.
    Basit, Salman Abdul
    Bahar, Yasemin
    Umer, Ahmed Muaaz
    Sattar, Yasar
    Alraies, M. Chadi
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (04)
  • [3] Efficacy of RADPAD protection drape in reducing radiation exposure in the catheterization laboratory-First Indian study
    Shah, Pujan
    Khanna, Roopali
    Kapoor, Aditya
    Goel, Pravin K.
    INDIAN HEART JOURNAL, 2018, 70 : S265 - S268
  • [4] Efficacy of RADPAD® protection drape in reducing radiation exposure to the primary operator during Transcatheter Aortic Valve Implantation (TAVI)
    Sharma, Divyesh
    Ramsewak, Adesh
    Manoharan, Ganesh
    Spence, Mark S.
    MINERVA CARDIOANGIOLOGICA, 2016, 64 (01): : 41 - 46
  • [5] A Double-Blind, Randomized, Sham-Controlled Trial of the Effect of a Radiation-Attenuating Drape on Radiation Exposure to Endoscopy Staff During ERCP
    Muniraj, Thiruvengadam
    Aslanian, Harry R.
    Laine, Loren
    Farrell, James
    Ciarleglio, Maria M.
    Deng, Yanhong
    Ho, Henry
    Jamidar, Priya A.
    AMERICAN JOURNAL OF GASTROENTEROLOGY, 2015, 110 (05): : 690 - 696
  • [6] Efficacy of acupuncture for psychogenic erectile dysfunction: a randomized, sham-controlled trial
    Wang, Hao
    Lei, Xulu
    Ma, Dongyue
    Zhao, Ziwei
    Wang, Anmin
    Du, Guanchao
    Zhang, Jiwei
    Wang, Fu
    Guo, Jun
    BASIC AND CLINICAL ANDROLOGY, 2023, 33 (01)
  • [7] Efficacy of acupuncture in migraine attack prophylaxis: a randomized sham-controlled trial
    Alecrim-Andrade, J
    Maciel, JA
    Cladellas, XCI
    Correa, HR
    Machado, HC
    Vasconcelos, GMS
    CEPHALALGIA, 2005, 25 (10) : 942 - 942
  • [8] Efficacy of acupuncture for psychogenic erectile dysfunction: a randomized, sham-controlled trial
    Hao Wang
    Xulu Lei
    Dongyue Ma
    Ziwei Zhao
    Anmin Wang
    Guanchao Du
    Jiwei Zhang
    Fu Wang
    Jun Guo
    Basic and Clinical Andrology, 33
  • [9] Efficacy and Safety of Acupuncture for Cyclic Mastalgia: Study Protocol for a Randomized, Sham-Controlled Trial
    Gao, Shuai
    Sun, Yuanjie
    Shi, Hangyu
    Fang, Jiufei
    Liu, Zhishun
    INTERNATIONAL JOURNAL OF WOMENS HEALTH, 2023, 15 : 845 - 855
  • [10] Efficacy of Kinesio taping on Mutational Falsetto: A Double Blind, Randomized, Sham-controlled Trial
    Atar, Sevgi
    Atar, Yavuz
    Sari, Seyin
    Karaketir, Semih
    Uygan, Ugur
    Karaketir, Seyma Gorcio
    Saltu, Ziya
    Kuru, Omer
    JOURNAL OF VOICE, 2023, 37 (06) : 968.e1 - 968.e12