Static MLC transmission simulation using two-dimensional ray tracing

被引:3
|
作者
Adam, David P. [1 ]
Bednarz, Bryan P. [1 ]
Frigo, Sean P. [2 ]
机构
[1] Univ Wisconsin, Dept Med Phys, Sch Med & Publ Hlth, Madison, WI 53792 USA
[2] Univ Wisconsin, Dept Human Oncol, Sch Med & Publ Hlth, 600 Highland Ave, Madison, WI 53792 USA
来源
关键词
HDMLC calibration; MLC modeling; ray tracing; MULTILEAF COLLIMATOR TRANSMISSION; FIELD; IMPLEMENTATION; RADIOTHERAPY; DEPENDENCE;
D O I
10.1002/acm2.13646
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose We investigated the hypothesis that the transmission function of rounded end linearly traveling multileaf collimators (MLCs) is constant with position. This assumption is made by some MLC models used in clinical treatment planning systems (TPSs) and in the Varian MLC calibration convention. If not constant, this would have implications for treatment plan QA results. Methods A two-dimensional ray-tracing tool to generate transmission curves as a function of leaf position was created and validated. The curves for clinically available leaf tip positions (-20 to 20 cm) were analyzed to determine the location of the beam edge (half-attenuation X-ray [XR]) location, the beam edge broadening (BEB, 80%-20% width), as well as the leaf tip zone width. More generalized scenarios were then simulated to elucidate trends as a function of leaf tip radius. Results In the analysis of the Varian high-definition MLC, two regions were identified: a quasi-static inner region centered about central axis (CAX), and an outer one, in which large deviations were observed. A phenomenon was identified where the half-attenuation ray position, relative to that of the tip or tangential ray, increases dramatically at definitive points from CAX. Similar behavior is seen for BEB. An analysis shows that as the leaf radius parameter value is made smaller, the size of the quasi-static region is greater (and vice versa). Conclusion The MLC transmission curve properties determined by this study have implications both for MLC position calibrations and modeling within TPSs. Two-dimensional ray tracing can be utilized to identify where simple behaviors hold, and where they deviate. These results can help clinical physicists engage with vendors to improve MLC models, subsequent fluence calculations, and hence dose calculation accuracy.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] RAY-TRACING IN A TWO-DIMENSIONAL IONOSPHERE
    LABAHN, RW
    RADIO SCIENCE, 1985, 20 (04) : 972 - 976
  • [2] DIGITAL RAY TRACING IN TWO-DIMENSIONAL REFRACTIVE FIELDS
    ANDERSEN, AH
    KAK, AC
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1982, 72 (05): : 1593 - 1606
  • [3] RAY TRACING IN TWO-DIMENSIONAL INHOMOGENEOUS-MEDIA
    XU, Y
    MCDONALD, JA
    AAPG BULLETIN-AMERICAN ASSOCIATION OF PETROLEUM GEOLOGISTS, 1988, 72 (03): : 398 - 398
  • [4] Two-dimensional analytic HF ray tracing in the ionosphere
    Norman, RJ
    2003 PROCEEDINGS OF THE INTERNATIONAL CONFERENCE ON RADAR, 2003, : 375 - 379
  • [5] Effects of earth curvature on two-dimensional ray tracing in underwater acoustics
    Yan, JG
    APPLIED ACOUSTICS, 1999, 57 (02) : 163 - 177
  • [6] Wide two-dimensional field laser ray-tracing aberrometer
    Mazzaferri, Javier
    Navarro, Rafael
    JOURNAL OF VISION, 2012, 12 (02): : 1 - 14
  • [7] Ray tracing method in phase space for two-dimensional optical systems
    Filosa, C.
    Boonkkamp, J. H. M. ten Thije
    Ijzerman, W. L.
    APPLIED OPTICS, 2016, 55 (13) : 3599 - 3606
  • [8] A two-dimensional analytic ray tracing technique accommodating horizontal gradients
    Norman, RJ
    Cannon, PS
    RADIO SCIENCE, 1997, 32 (02) : 387 - 396
  • [9] Simulation and verification of two-dimensional ohmic heating in static system
    Fu, WR
    Hsieh, CC
    JOURNAL OF FOOD SCIENCE, 1999, 64 (06) : 946 - 949
  • [10] Two-dimensional ray-tracing modeling for propagation prediction in microcellular environments
    Rizk, K
    Wagen, JF
    Gardiol, F
    IEEE TRANSACTIONS ON VEHICULAR TECHNOLOGY, 1997, 46 (02) : 508 - 518