共 50 条
Evaluation of eccentric aortic regurgitation by color Doppler jet and color Doppler-imaged vena contracta measurements: An animal study of quantified aortic regurgitation
被引:13
|作者:
Ishii, M
Jones, M
Shiota, T
Heinrich, R
Yamada, I
Sinclair, B
Yoganathan, AP
Sahn, DJ
机构:
[1] NHLBI, LAMS, NIH, BETHESDA, MD 20892 USA
[2] OREGON HLTH SCI UNIV, PORTLAND, OR 97201 USA
[3] GEORGIA INST TECHNOL, ATLANTA, GA 30332 USA
关键词:
D O I:
10.1016/S0002-8703(96)90314-2
中图分类号:
R5 [内科学];
学科分类号:
1002 ;
100201 ;
摘要:
To evaluate the utility of measurements of the color Doppler jet area, jet length, and width of the color Doppler-imaged vena contracta (the smallest flow diameter in any part of the flow acceleration field) as methods for quantifying aortic regurgitation (AR), eight sheep with surgically induced AR were studied. AR was quantified as peak and mean regurgitant flow rates, regurgitant stroke volumes, and regurgitant fractions as determined with pulmonary acid aortic electromagnetic flow probes and flowmeters balanced against each other. Simple linear regression analysis between the maximal color jet areas, jet length, and flowmeter data showed only moderately good correlation (jet area: 0.42 less than or equal to r less than or equal to 0.57, SEE = 2.85 cm(2); jet length: 0.42 less than or equal to r less than or equal to 0.59, SEE = 1.23 cm). In contrast, the width of color Doppler-imaged vena contracta was a better indicator of the severity of AR on the basis of the electromagnetic flowmeter methods (0.73 less than or equal to r less than or equal to 0.90, SEE = 0.15 cm). Therefore the color Doppler jet length and jet area methods have limited use for determining AR, whereas the width of the color Doppler-imaged vena contracta can be used for quantifying the severity of AR.
引用
收藏
页码:796 / 804
页数:9
相关论文