Surface grafting of biocompatible phospholipid polymer MPC provides wear resistance of tibial polyethylene insert in artificial knee joints

被引:33
|
作者
Moro, T. [1 ]
Takatori, Y. [1 ]
Kyomoto, M. [1 ]
Ishihara, K. [2 ]
Saiga, K. [1 ]
Nakamura, K. [3 ]
Kawaguchi, H. [3 ]
机构
[1] Univ Tokyo, Dept Sci Joint Reconstruct, Bunkyo Ku, Tokyo 1130033, Japan
[2] Univ Tokyo, Dept Mat Engn, Bunkyo Ku, Tokyo 1130033, Japan
[3] Univ Tokyo, Dept Sensory & Motor Syst Med, Bunkyo Ku, Tokyo 1130033, Japan
关键词
Artificial knee joint; Arthroplasty; Polyethylene; Wear; Loosening; CROSS-LINKED POLYETHYLENE; MOLECULAR-WEIGHT POLYETHYLENE; TOTAL HIP; IN-VITRO; UNITED-STATES; SIMULATOR; UHMWPE; ARTHROPLASTY; REPLACEMENTS; PARTICLES;
D O I
10.1016/j.joca.2010.05.019
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: Aseptic loosening of artificial knee joints induced by wear particles from a tibial polyethylene (PE) insert is a serious problem limiting their longevity. This study investigated the effects of grafting with our original biocompatible phospholipid polymer 2-methacryloyloxyethyl phosphorylcholine (MPC) on the insert surface. Methods: The hydrophilicity of the PE surface was determined by the contact angle of a water droplet, and the friction torque was measured against a cobalt-chromium alloy component. The wear amount was compared among PE inserts with or without cross-linking and MPC grafting during 5 x 10(6) cycles of loading in a knee joint simulator. The surfaces of the insert and the wear particles in the lubricant were subjected to electron and laser microscopic analyses. The mechanical properties of the inserts were evaluated by the small punch test. Results: The MPC grafting increased hydrophilicity and decreased friction torque. In the simulator experiment, the wear of the tibial insert was significantly suppressed in the cross-linked PE (CLPE) insert, and even more dramatically decreased in the MPC-grafted CLPE insert, as compared to that in the non-cross-linked PE insert. Surface analyses confirmed the wear resistance by the cross-linking, and further by the MPC grafting. The particle size distribution was not affected by cross-linking or MPC grafting. The mechanical properties of the insert material remained unchanged during the loading regardless of the cross-linking or grafting. Conclusion: Surface grafting with MPC polymer furnished the PE insert with wear resistance in an artificial knee joint through increased hydrophilicity and decreased friction torque. (C) 2010 Osteoarthritis Research Society International. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:1174 / 1182
页数:9
相关论文
共 6 条
  • [1] Polyethylene wear on the distal tibial insert surface in total knee arthroplasty
    Gabriel, SM
    Dennis, DA
    Honey, MJ
    Scott, RD
    KNEE, 1998, 5 (03): : 221 - 228
  • [2] Surface grafting of artificial joints with a biocompatible polymer for preventing periprosthetic osteolysis
    Toru Moro
    Yoshio Takatori
    Kazuhiko Ishihara
    Tomohiro Konno
    Yorinobu Takigawa
    Tomiharu Matsushita
    Ung-il Chung
    Kozo Nakamura
    Hiroshi Kawaguchi
    Nature Materials, 2004, 3 : 829 - 836
  • [3] Surface grafting of artificial joints with a biocompatible polymer for preventing periprosthetic osteolysis
    Moro, T
    Takatori, Y
    Ishihara, K
    Konno, T
    Takigawa, Y
    Matsushita, T
    Chung, UI
    Nakamura, K
    Kawaguchi, H
    NATURE MATERIALS, 2004, 3 (11) : 829 - 836
  • [4] Wear resistance of the biocompatible phospholipid polymer-grafted highly cross-linked polyethylene liner against larger femoral head
    Moro, Toru
    Takatori, Yoshio
    Kyomoto, Masayuki
    Ishihara, Kazuhiko
    Kawaguchi, Hiroshi
    Hashimoto, Masami
    Tanaka, Takeyuki
    Oshima, Hirofumi
    Tanaka, Sakae
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2015, 33 (07) : 1103 - 1110
  • [5] Computational wear of knee implant polyethylene insert surface under continuous dynamic loading and posterior tibial slope variation based on cadaver experiments with comparative verification
    Alaettin Ozer
    BMC Musculoskeletal Disorders, 23
  • [6] Computational wear of knee implant polyethylene insert surface under continuous dynamic loading and posterior tibial slope variation based on cadaver experiments with comparative verification
    Ozer, Alaettin
    BMC MUSCULOSKELETAL DISORDERS, 2022, 23 (01)