The agreement between oscillometric and intra-arterial technique for blood pressure monitoring in the lower extremities for infants and toddlers undergoing aortic coarctation repair

被引:2
|
作者
Peng, Zhe-Zhe [1 ,2 ]
Zhang, Ma-Zhong [1 ,2 ]
Sun, Ying [1 ,2 ]
Bai, Jie [2 ]
Gu, Hong-Bin [2 ]
Liu, Pei-Pei [1 ,2 ]
Li, Min [1 ,2 ]
Cai, Mei-Hua [1 ]
机构
[1] Shanghai Jiao Tong Univ, Pediat Clin Pharmacol Lab, Shanghai Childrens Med Ctr, Sch Med, Shanghai, Peoples R China
[2] Shanghai Jiao Tong Univ, Dept Anesthesiol, Shanghai Childrens Med Ctr, Sch Med, Shanghai, Peoples R China
关键词
oscillometry; infant; coarctation of aorta; blood pressure; catheterization; ACCURACY; PRETERM; ARM;
D O I
10.1111/pan.12989
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
ObjectiveAnesthetic management for patients undergoing surgical repair of aortic coarctation (CoA) should include constant blood pressure monitoring of the right upper extremity and a lower extremity. The delayed or absent pulse in the lower limbs often leads to unsuccessful arterial cannulation in infants and the oscillometric technique used for blood pressure measurement. The aim of this study was to evaluate the agreement between the oscillometric method and intra-arterial technique for blood pressure monitoring in the lower limbs of infants undergoing CoA. MethodsA total of 45 infants diagnosed with isolated CoA were initially enrolled in this study and five were excluded because of cannulation failure. Thus, 40 patients had their blood pressure measured simultaneously by both oscillometric technique on the thigh and femoral artery catheterization. After induction and intubation, five pairs of blood pressure readings from each patient were collected in an interval of 3 min. Statistical analysis was accomplished by revised Bland-Altman analysis. ResultsThere was a strong correlation between oscillometric and invasive blood pressure measurements [systolic blood pressure (SBP) r = 0.771, diastolic blood pressure (DBP) r = 0.704 and mean artery pressure (MAP) r = 0.850]. The mean difference and 95% limits of agreement (95% LOA) between oscillometric and femoral artery blood pressure readings was 3.830 mmHg (-19.297, 26.957) for SBP, -8.725 mmHg (-26.236, 8.786) for DBP, and -3.235 mmHg (-18.842, 12.372) for MAP. There were only one pair of MAP (1/40) and two pairs of SBP readings (2/40) out of range (95% LOA), and all of paired DBP readings were within 95% LOA. ConclusionThere was a good agreement between oscillometric and invasive blood pressure measurements of lower extremities in infants with isolated CoA statistically. However, the oscillometry-measured SBP showed a tendency to overestimate the intra-arterial blood pressure reference, while oscillometry-measured DBP underestimated its reference. MAP measurement provided the most accurate and reliable results in this study.
引用
收藏
页码:1091 / 1096
页数:6
相关论文
共 20 条
  • [1] Comment on: Z-Z Peng "The agreement between oscillometric and intra-arterial technique for blood pressure monitoring in the lower extremities of infants undergoing aortic coarctation repair'
    Withington, Davinia E.
    PEDIATRIC ANESTHESIA, 2017, 27 (05) : 552 - 552
  • [2] Reply to Withington, Davinia, regarding their comment "Comment on: Z-Z Peng "The agreement between ocillometric and intra-arterial technique for blood pressure monitoring in the lower extremities of infants undergoing aortic coarctation repair"
    Sun, Ying
    Wu, Jun-Zheng
    Peng, Zhe-Zhe
    Zhang, Ma-Zhong
    Liu, Pei-Pei
    PEDIATRIC ANESTHESIA, 2017, 27 (05) : 553 - 554
  • [3] Agreement between noninvasive oscillometric and invasive intra-arterial blood pressure in children with ruptured brain arteriovenous malformations
    Chang, Nathan
    Poon, Diana
    Casazza, May
    Medrano, Amanda
    Basnett, Kaitlyn
    Koilparampil, Lesley
    Rasmussen, Lindsey
    CLINICAL NEUROLOGY AND NEUROSURGERY, 2024, 243
  • [4] Blood Pressure Measurement Compared Between Oscillometric Blood Pressure Monitoring and Gold Standard Intra-Arterial Blood Pressure Monitoring in Adult Shock Resuscitation Patients
    Tongyoo, Surat
    Visuthisakchai, Suharit
    Permpikul, Chairat
    JOURNAL OF CRITICAL & INTENSIVE CARE, 2022, 13 (03): : 102 - 109
  • [5] Blood pressure monitoring during arrhythmia: agreement between automated brachial cuff and intra-arterial measurements
    Lakhal, K.
    Ehrmann, S.
    Martin, M.
    Faiz, S.
    Reminiac, F.
    Cinotti, R.
    Capdevila, X.
    Asehnoune, K.
    Blanloeil, Y.
    Rozec, B.
    Boulain, T.
    BRITISH JOURNAL OF ANAESTHESIA, 2015, 115 (04) : 540 - 549
  • [6] Oscillometric and Intra-arterial Blood Pressure in Preterm and Term Infants: Extent of Discrepancy and Factors Associated with Inaccuracy
    Shimokaze, Tomoyuki
    Akaba, Kazuhiro
    Saito, Emi
    AMERICAN JOURNAL OF PERINATOLOGY, 2015, 32 (03) : 277 - 282
  • [7] Continuous intra-arterial blood gas monitoring in patients undergoing suprarenal aortic surgery
    Gordon, G
    Flaherty, P
    Strafford, M
    Belkin, M
    ODonnell, TF
    Bojar, R
    VASCULAR SURGERY, 1996, 30 (01): : 29 - 36
  • [8] Comparison of blood pressure variability by means of non-invasive oscillometric and intra-arterial blood pressure monitoring in acute stroke patients
    Manios, E
    Tsivgoulis, G
    Spengos, K
    Spiliopoulou, J
    Dolianitis, K
    Tseke, P
    Xynos, K
    Dimitriou, A
    Vemmos, K
    Zakopoulos, N
    Mavrikakis, M
    JOURNAL OF HYPERTENSION, 2005, 23 : S125 - S125
  • [9] Accuracy of oscillometric noninvasive blood pressure compared with intra-arterial blood pressure in infants and small children during neurosurgical procedures An observational study
    Meidert, Agnes S.
    Tholl, Martin
    Huettl, Tanija K.
    Bernasconi, Patricia
    Peraud, Aurelia
    Briegel, Josef
    EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2019, 36 (06) : 400 - 405
  • [10] Comparison between oscillometric and intra-arterial blood pressure measurements in ill preterm and full-term neonates
    Lalan, Shwetal
    Blowey, Douglas
    JOURNAL OF THE AMERICAN SOCIETY OF HYPERTENSION, 2014, 8 (01) : 36 - 44