High resolution corneal and single pulse imaging with line field spectral domain optical coherence tomography

被引:31
|
作者
Lawman, Samuel [1 ,2 ]
Dong, Yue [1 ]
Williams, Bryan M. [2 ]
Romano, Vito [3 ]
Kaye, Stephen [2 ,3 ]
Harding, Simon P. [2 ,3 ]
Willoughby, Colin [2 ,3 ]
Shen, Yao-Chun [1 ]
Zheng, Yalin [2 ,3 ]
机构
[1] Univ Liverpool, Dept Elect Engn & Elect, Liverpool L69 3GJ, Merseyside, England
[2] Univ Liverpool, Dept Eye & Vis Sci, Liverpool L7 8TX, Merseyside, England
[3] Royal Liverpool Univ Hosp, St Pauls Eye Unit, Liverpool L7 8XP, Merseyside, England
来源
OPTICS EXPRESS | 2016年 / 24卷 / 11期
基金
英国工程与自然科学研究理事会; 美国国家卫生研究院;
关键词
ENDOTHELIAL KERATOPLASTY; PHASE; TIME; SYSTEM; INTERFEROMETRY;
D O I
10.1364/OE.24.012395
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We report the development of a Spectral Domain Line Field Optical Coherence Tomography (LF-OCT) system, using a broad bandwidth and spatial coherent Super-Continuum (SC) source. With conventional quasi-Continuous Wave (CW) setup we achieve axial resolutions up to 2.1 mu m in air and 3D volume imaging speeds up to 213 kA-Scan/s. Furthermore, we report the use of a single SC pulse, of 2 ns duration, to temporally gate an OCT B-Scan image of 70 A-Scans. This is the equivalent of 35 GA-Scans/s. We apply the CW setup for high resolution imaging of the fine structures of a human cornea sample ex-vivo. The single pulse setup is applied to imaging of a coated pharmaceutical tablet. The fixed pattern noise due to spectral noise is removed by subtracting the median magnitude A-Scan. We also demonstrate that the Fourier phase can be used to remove aberration caused artefacts. (C)2016 Optical Society of America
引用
收藏
页码:2395 / 2405
页数:11
相关论文
共 50 条
  • [1] in vivo imaging of raptor retina with ultra high resolution spectral domain optical coherence tomography
    Ruggeri, Marco
    Major, James C., Jr.
    McKeown, Craig
    Wehbe, Hassan
    Jiao, Shuliang
    Puliafito, Carmen A.
    OPHTHALMIC TECHNOLOGIES XVIII, 2008, 6844 : 84402 - 84402
  • [2] Repeated, noninvasive, high resolution spectral domain optical coherence tomography imaging of zebrafish embryos
    Kagemann, Larry
    Ishikawa, Hiroshi
    Zou, Jian
    Charukamnoetkanok, Puwat
    Wollstein, Gadi
    Townsend, Kelly A.
    Gabriele, Michelle L.
    Bahary, Nathan
    Wei, Xiangyun
    Fujimoto, James G.
    Schuman, Joel S.
    MOLECULAR VISION, 2008, 14 (253-55): : 2157 - 2170
  • [3] Integrated line-field optical coherence tomography and scheimpflug imaging for corneal imaging
    Li, Xiaoran
    Lawman, Samuel
    Dong, Biqin
    Lin, Hungyen
    Shen, Yaochun
    Zheng, Yalin
    OPTICS AND LASERS IN ENGINEERING, 2024, 182
  • [4] Standard resolution spectral domain optical coherence tomography in clinical ophthalmic imaging
    Szkulmowska, A
    Cyganek, M
    Targowski, P
    Kowalczyk, A
    Kaluzny, JJ
    Wojtkowski, M
    Fujimoto, JG
    OPHTHALMIC TECHNOLOGIES XV, 2005, 5688 : 69 - 76
  • [5] In vivo high resolution human corneal imaging using full-field optical coherence tomography
    Mazlin, Viacheslav
    Xiao, Peng
    Dalimier, Eugenie
    Grieve, Kate
    Irsch, Kristina
    Sahel, Jost-Alain
    Fink, Mathias
    Boccara, A. Claude
    BIOMEDICAL OPTICS EXPRESS, 2018, 9 (02): : 557 - 568
  • [6] High Resolution Imaging System Using Spectral Domain Optical Coherence Tomography Using NIR Source
    Bhatia, Prachi
    Choudhari, Suyog
    Rodrigues, Anjana
    Pati, Mukesh
    Makkar, Roshan
    PROCEEDINGS OF THE 2016 IEEE INTERNATIONAL CONFERENCE ON WIRELESS COMMUNICATIONS, SIGNAL PROCESSING AND NETWORKING (WISPNET), 2016, : 2212 - 2216
  • [7] High resolution retinal imaging with a compact adaptive optics spectral domain optical coherence tomography system
    Hammer, Daniel X.
    Iftimia, Nicusor V.
    Bigelow, Chad E.
    Ustun, Teoman E.
    Bloom, Benjamin
    Ferguson, R. Daniel
    Burns, Stephen A.
    OPHTHALMIC TECHNOLOGIES XVII, 2007, 6426
  • [8] Simultaneous dual-band optical coherence tomography in the spectral domain for high resolution in vivo imaging
    Cimalla, Peter
    Walther, Julia
    Mehner, Mirko
    Cuevas, Maximiliano
    Koch, Edmund
    OPTICS EXPRESS, 2009, 17 (22): : 19486 - 19500
  • [9] Clinical retinal imaging by high speed, ultrahigh resolution spectral/Fourier domain optical coherence tomography
    Wojtkowski, M
    Srinivasan, VJ
    Ko, TH
    Duker, JS
    Schuman, JS
    Kowalczyk, A
    Fujimoto, JG
    Ophthalmic Technologies XV, 2005, 5688 : 84 - 89
  • [10] Is high-resolution spectral domain optical coherence tomography reliable in nystagmus?
    Thomas, Mervyn G.
    Kumar, Anil
    Thompson, John R.
    Proudlock, Frank A.
    Straatman, Kees
    Gottlob, Irene
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2013, 97 (04) : 534 - 536