New set-up for capillary isoelectric focusing in uncoated capillaries

被引:33
|
作者
Kilár, F [1 ]
Végvári, A [1 ]
Mód, A [1 ]
机构
[1] Univ Pecs, Sch Med, Cent Res Lab, H-7643 Pecs, Hungary
基金
匈牙利科学研究基金会;
关键词
isoelectric focusing; sandwich injection; injection methods; haemoglobin; dyes; nitrophenols;
D O I
10.1016/S0021-9673(98)00377-X
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Substituted aminomethylphenol dyes, low-molecular-mass isoelectric point (pI) markers and hemoglobin samples from normal individuals and diabetic patients were used to test a new set-up of capillary isoelectric focusing (cIEF) in uncoated capillaries. In previous cIEF methods, a mixture of sample components and carrier ampholytes was applied in the capillary and analyzed. In the new set-up a fractionated injection protocol is used to apply a 'sandwich' ampholyte-sample-ampholyte plug in the capillary for analysis. This new set-up allows the separation of amphoteric compounds having pI values outside the pH region of the ampholytes applied in the capillary with high precision. The high resolution power of this technique was proven with the analysis of hemoglobin variants. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:349 / 360
页数:12
相关论文
共 50 条