Nebulizers in cystic fibrosis: a source of bacterial contamination in cystic fibrosis patients?

被引:13
|
作者
Costa Brzezinski, Lorena Xavier [1 ]
Riedi, Carlos Antonio [2 ]
Kussek, Paulo [1 ]
de Melo de Souza, Helena Homem [3 ]
Rosario, Nelson [2 ]
机构
[1] Hosp Pequeno Principe, Curitiba, Parana, Brazil
[2] Univ Fed Parana, BR-80060000 Curitiba, Parana, Brazil
[3] Univ Fed Parana, Unidade Apoio Diagnost, Hosp Clin Parana, BR-80060000 Curitiba, Parana, Brazil
关键词
Cystic fibrosis; Nebulizers and vaporizers; Disinfection; HOME TREATMENT; CHILDREN;
D O I
10.1590/S1806-37132011000300010
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Objective: To determine whether nebulizers are a source of microbial contamination in patients with cystic fibrosis, as well as whether the technique and frequency of disinfection of these devices is appropriate. Methods: This was a cross-sectional, uncontrolled observational study. Samples were collected from 28 patients with cystic fibrosis. Samples were collected at the homes of the patients, who were not previously informed of the purpose of the visit. Three samples were collected from each patient: one from the nebulizer chamber, one from the mask/mouthpiece, and one from the patient (orophatyngeal swab/sputum). The samples were properly stored and taken for analyses. The patients, their parents, or their legal guardians completed a questionnaire regarding nebulizer cleaning and disinfecting methods. Results: We collected 84 samples from the 28 patients. Of those 28 patients, 15 (53.5%) were male. The median age of the patients was 11 years (range, 1-27 years). Of the 28 patients, 15 presented with positive oropharyngeal swab/sputum sample cultures. The most common bacterial isolates were Staphylococcus aureus (in 8 patients) and Pseudomonas aeruginosa (in 4 patients). Although the samples obtained from the nebulizers presented with various pathogens in culture, no specific species predominated. In 27 cases (96.7%), there were no associations between the samples obtained from the nebulizers and those obtained from the patients in terms of the results of the cultures. Cleaning and disinfection of nebulizers were inappropriate in 22 cases (78.6%). Conclusions: In this sample of patients, despite the inappropriate disinfection techniques, nebulizers were not found to be a source of microbial contamination.
引用
收藏
页码:341 / 347
页数:7
相关论文
共 50 条
  • [1] Low bacterial contamination of nebulizers in home treatment of cystic fibrosis patients
    Jakobsson, BM
    Onnered, AB
    Hjelte, L
    Nystrom, B
    [J]. JOURNAL OF HOSPITAL INFECTION, 1997, 36 (03) : 201 - 207
  • [2] Nebulizers: Source of bacterial contamination in patients with cystic fibrosis? (vol 37, pg 341, 2011)
    Brzezinski, L. X.
    Riedi, C. A.
    Kussek, P.
    Souza, H. H.
    Rosario, N.
    [J]. JORNAL BRASILEIRO DE PNEUMOLOGIA, 2011, 37 (04) : 571 - 571
  • [3] Nebulizers in cystic fibrosis
    Hambleton, G
    Hung, JC
    Super, M
    [J]. JOURNAL OF THE ROYAL SOCIETY OF MEDICINE, 1996, 89 : 14 - 18
  • [4] Home nebulizers in patients with cystic fibrosis
    Amirav, I
    Newhouse, M
    [J]. JOURNAL OF PEDIATRICS, 1998, 133 (05): : 715 - 715
  • [5] Microbial contamination of nebulizers in the home treatment of cystic fibrosis
    Blau, H.
    Mussaffi, H.
    Zahav, M. Mei
    Prais, D.
    Livne, M.
    Czitron, B. M.
    Cohen, H. A.
    [J]. CHILD CARE HEALTH AND DEVELOPMENT, 2007, 33 (04) : 491 - 495
  • [6] Microbiological contamination of nebulizers used by cystic fibrosis patients: an underestimated problem
    Riquena, Barbara
    Velloso Monte, Luciana de Freitas
    Lopes, Agnaldo Jose
    Ferreira da Silva-Filho, Luiz Vicente Ribeiro
    Damaceno, Neiva
    Aquino, Evanirso da Silva
    Cauduro Marostica, Paulo Jose
    Ribeiro, Jose Dirceu
    [J]. JORNAL BRASILEIRO DE PNEUMOLOGIA, 2019, 45 (03)
  • [7] Bacterial contamination of cystic fibrosis clinics
    Zuckerman, Jonathan B.
    Zuaro, Deborah E.
    Prato, B. Stephen
    Ruoff, Kathryn L.
    Sawicki, Rafal W.
    Quinton, Hebe B.
    Saiman, Lisa
    [J]. JOURNAL OF CYSTIC FIBROSIS, 2009, 8 (03) : 186 - 192
  • [8] Home nebulizers in patients with cystic fibrosis - Reply
    Rosenfeld, M
    [J]. JOURNAL OF PEDIATRICS, 1998, 133 (05): : 716 - 716
  • [9] Microbiologic contamination study of nebulizers after aerosol therapy in patients with cystic fibrosis
    Vassal, S
    Taamma, R
    Marty, N
    Sardet, A
    d'Athis, P
    Brémont, F
    Dalphin, ML
    Plésiat, P
    Rault, G
    Thubert, J
    Dominique, S
    Lemeland, JF
    Derelle, J
    Blech, MF
    Roussey, M
    Perrin, M
    Sautegeau, A
    [J]. AMERICAN JOURNAL OF INFECTION CONTROL, 2000, 28 (05) : 347 - 351
  • [10] Bacterial Contamination of Home Nebulizers in Children with Cystic Fibrosis and Clinical Implication on the Number of Pulmonary Exacerbations
    Farahbakhsh, Nazanin
    Tabatabaii, Seyed Ahmad
    Khanbabaee, Ghamartaj
    Sadr, Saeed
    Aghdam, Maryam Kazemi
    [J]. EUROPEAN RESPIRATORY JOURNAL, 2019, 54