Development and characterization of a dynamic lesion phantom for the quantitative evaluation of dynamic contrast-enhanced MRI

被引:11
|
作者
Freed, Melanie [1 ,2 ]
de Zwart, Jacco A. [3 ]
Hariharan, Prasanna [4 ]
Myers, Matthew R. [4 ]
Badano, Aldo [1 ]
机构
[1] US FDA, Div Imaging & Appl Math, Off Sci & Engn Labs, Ctr Devices & Radiol Hlth, Silver Spring, MD 20993 USA
[2] Univ Maryland, Dept Bioengn, College Pk, MD 20742 USA
[3] Natl Inst Neurol Disorders & Stroke, Lab Funct & Mol Imaging, Adv MRI Sect, NIH, Bethesda, MD 20892 USA
[4] US FDA, Div Solid & Fluid Mech, Off Sci & Engn Labs, Ctr Devices & Radiol Hlth, Silver Spring, MD 20993 USA
关键词
dynamic phantom; DCE-MRI; breast imaging; SPOILED GRADIENT-ECHO; MYOCARDIAL-PERFUSION; DIAGNOSTIC-ACCURACY; BREAST MRI; PARAMETERS; SEQUENCES; RESONANCE; DTPA; STANDARDIZATION; QUANTIFICATION;
D O I
10.1118/1.3633911
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To develop a dynamic lesion phantom that is capable of producing physiological kinetic curves representative of those seen in human dynamic contrast-enhanced MRI (DCE-MRI) data. The objective of this phantom is to provide a platform for the quantitative comparison of DCE-MRI protocols to aid in the standardization and optimization of breast DCE-MRI. Methods: The dynamic lesion consists of a hollow, plastic mold with inlet and outlet tubes to allow flow of a contrast agent solution through the lesion over time. Border shape of the lesion can be controlled using the lesion mold production method. The configuration of the inlet and outlet tubes was determined using fluid transfer simulations. The total fluid flow rate was determined using x-ray images of the lesion for four different flow rates (0.25, 0.5, 1.0, and 1.5 ml/s) to evaluate the resultant kinetic curve shape and homogeneity of the contrast agent distribution in the dynamic lesion. High spatial and temporal resolution x-ray measurements were used to estimate the true kinetic curve behavior in the dynamic lesion for benign and malignant example curves. DCE-MRI example data were acquired of the dynamic phantom using a clinical protocol. Results: The optimal inlet and outlet tube configuration for the lesion molds was two inlet molds separated by 30 degrees and a single outlet tube directly between the two inlet tubes. X-ray measurements indicated that 1.0 ml/s was an appropriate total fluid flow rate and provided truth for comparison with MRI data of kinetic curves representative of benign and malignant lesions. DCE-MRI data demonstrated the ability of the phantom to produce realistic kinetic curves. Conclusions: The authors have constructed a dynamic lesion phantom, demonstrated its ability to produce physiological kinetic curves, and provided estimations of its true kinetic curve behavior. This lesion phantom provides a tool for the quantitative evaluation of DCE-MRI protocols, which may lead to improved discrimination of breast cancer lesions. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3633911]
引用
收藏
页码:5601 / 5611
页数:11
相关论文
共 50 条
  • [1] Development and Evaluation of a Phantom for Dynamic Contrast-Enhanced Imaging
    Brauweiler, Robert
    Eisa, Fabian
    Hupfer, Martin
    Nowak, Tristan
    Kolditz, Daniel
    Kalender, Willi A.
    INVESTIGATIVE RADIOLOGY, 2012, 47 (08) : 462 - 467
  • [2] Evaluation of Mesorectal Microcirculation With Quantitative Dynamic Contrast-Enhanced MRI
    Yang, Xinyue
    Liu, Yiyan
    Chen, Yan
    Wen, Ziqiang
    Lin, Bomiao
    Quan, Xianyue
    Yu, Shenping
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2020, 215 (06) : 1370 - 1376
  • [3] Development of a dynamic flow imaging phantom for dynamic contrast-enhanced CT
    Driscoll, B.
    Keller, H.
    Coolens, C.
    MEDICAL PHYSICS, 2011, 38 (08) : 4866 - 4880
  • [4] Quantitative dynamic contrast-enhanced MRI of the lung in asthma
    Zhang, Wei-juan
    Niven, Robert
    Young, Simon
    Liu, Yu-zhen
    Parker, Geoffrey
    Naish, Josephine
    EUROPEAN RESPIRATORY JOURNAL, 2014, 44
  • [5] Dynamic contrast-enhanced MRI for evaluation of cardiac tumors
    Libicher, Martin
    Kauffmann, Guenther W.
    Hosch, Waldemar
    EUROPEAN RADIOLOGY, 2006, 16 (08) : 1858 - 1859
  • [6] Dynamic contrast-enhanced MRI for evaluation of cardiac tumors
    Martin Libicher
    Günther W. Kauffmann
    Waldemar Hosch
    European Radiology, 2006, 16 : 1858 - 1859
  • [7] Dynamic contrast-enhanced MRI for oncology drug development
    Sung, Yu Sub
    Park, Bumwoo
    Choi, Yoonseok
    Lim, Hyeong-Seok
    Woo, Dong-Cheol
    Kim, Kyung Won
    Kim, Jeong Kon
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2016, 44 (02) : 251 - 264
  • [8] Dynamic Contrast-Enhanced MRI in Oncology Drug Development
    Cheng, Hai-Ling Margaret
    CURRENT CLINICAL PHARMACOLOGY, 2007, 2 (02): : 111 - 122
  • [9] Quantitative assessment of pulmonary perfusion with dynamic contrast-enhanced MRI
    Hatabu, H
    Tadamura, E
    Levin, DL
    Chen, Q
    Li, W
    Kim, D
    Prasad, PV
    Edelman, RR
    MAGNETIC RESONANCE IN MEDICINE, 1999, 42 (06) : 1033 - 1038
  • [10] Weakly Supervised Breast Lesion Detection in Dynamic Contrast-Enhanced MRI
    Sun, Rong
    Wei, Chuanling
    Jiang, Zhuoyun
    Huang, Gang
    Xie, Yuanzhong
    Nie, Shengdong
    JOURNAL OF DIGITAL IMAGING, 2023, 36 (04) : 1553 - 1564