The study on the spectral irradiance responsiveness calibration for UV-VUV solar spectrometer

被引:2
|
作者
Li, Hanshuang [1 ]
Li, Bo [1 ]
Wang, Shurong [1 ]
Li, Zhigang [1 ]
Li, Futian [1 ]
机构
[1] Chinese Acad Sci, Changchun Inst Opt Fine Mech & Phys, 3888 Dong Nanhu Rd, Changchun 130033, Jilin, Peoples R China
来源
OPTIK | 2020年 / 208卷
基金
美国国家科学基金会;
关键词
Spectral irradiance responsiveness calibration; UV-VUV solar spectrometer; Solar spectral irradiance; RADIATION FACILITY; ULTRAVIOLET; BLACKBODY; MONITOR;
D O I
10.1016/j.ijleo.2020.164578
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The monitoring of the solar spectral irradiance has important scientific significance and application value. This paper describes the principle, structure and performance of the UV-VUV solar spectrometer. The calibration principle and the calibration method of the spectral irradiance responsiveness are emphatically discussed, based on synchrotron radiation. The spectral irradiance responsiveness calibration of UV-VUV solar spectrometer in the 165 nm-320 nm spectral range is realized under vacuum environment with PTB transfer standard deuterium lamp as standard source. The results of the calibration are achieved, factors affecting the accuracy of the calibration are analyzed. The synthesized uncertainty of the calibration is 5.1 %.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] UV-VUV Spectral irradiance responsivity calibration for space remote sensing spectroradiometer
    Huang, Yu
    Wang, Shurong
    Lin, Guanyu
    Wang, Longqi
    ADVANCED MATERIALS AND DEVICES FOR SENSING AND IMAGING III, 2008, 6829
  • [2] Quantum Efficiency Calibration of UV-VUV Photomultiplier Tube
    Li Hanshuang
    Li Bo
    Wang Shurong
    CHINESE JOURNAL OF LASERS-ZHONGGUO JIGUANG, 2018, 45 (08):
  • [3] Spectral irradiance calibration of space remote sensing spectral radiometer in range of UV and VUV
    Huang, Yu
    Wang, Shu-Rong
    Zhang, Zhen-Duo
    Lin, Guan-Yu
    Li, Fu-Tian
    Guangzi Xuebao/Acta Photonica Sinica, 2007, 36 (SUPPL.): : 187 - 191
  • [4] UV-VUV radiation calibration facility based on synchrotron radiation
    Department of Optical Engineering, School of Information Science and Technology, Beijing Institute of Technology, Beijing 100081, China
    不详
    不详
    Beijing Ligong Daxue Xuebao, 2007, 9 (807-810):
  • [5] The depolarization characteristic of quartz depolarizer in UV-VUV spectral region
    Liu, Y
    Fu, SJ
    Li, FT
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2004, 24 (03) : 372 - 375
  • [6] Polarization characteristic of LiF polarizer at UV-VUV spectral range
    Liu, Y
    Li, FT
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2002, 22 (04) : 552 - 555
  • [7] Polarization characteristic of LiF polarizer at UV-VUV spectral range
    Liu, Ying
    Li, Fu-Tian
    Guang Pu Xue Yu Guang Pu Fen Xi/Spectroscopy and Spectral Analysis, 2002, 22 (04):
  • [8] Characteristics of various diffuser materials at UV-VUV spectral range
    Liu, Y.
    Li, F.T.
    Guangxue Xuebao/Acta Optica Sinica, 2001, 21 (03): : 371 - 375
  • [9] HIGH SENSITIVE DETECTION OF MOLECULES WITH A TUNABLE UV-VUV LASER MASS-SPECTROMETER
    ZACHARIAS, H
    ROTTKE, H
    WELGE, KH
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1980, 70 (11) : 1386 - 1386
  • [10] SOLAR UV SPECTRAL IRRADIANCE VARIATIONS
    DONNELLY, RF