Comparative biomechanical analysis of rigid vs. flexible fixation devices for the lumbar spine: A geometrically patient-specific poroelastic finite element study

被引:13
|
作者
Khalaf, Kinda [1 ,2 ]
Nikkhoo, Mohammad [3 ]
机构
[1] Khalifa Univ Sci & Technol, Dept Biomed Engn, POB 127788, Abu Dhabi, U Arab Emirates
[2] Hlth Engn Innovat Ctr, POB 127788, Abu Dhabi, U Arab Emirates
[3] Islamic Azad Univ, Dept Biomed Engn, Sci & Res Branch, Tehran, Iran
关键词
Poroelastic modeling; Finite element analysis; Patient-specific; Lumbar fixation device; Dynamic rod; Interspinous process devices; INTERBODY FUSION; MOTION SEGMENT; INTERVERTEBRAL DISCS; INTRADISCAL PRESSURE; LOAD-BEARING; ADJACENT; MODEL; STABILIZATION; DEGENERATION; VALIDATION;
D O I
10.1016/j.cmpb.2021.106481
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Background and Objective: Lumbar spinal stenosis (LSS), or the narrowing of the spinal canal, continues to be the leading preoperative diagnosis for adults older than 65 years who undergo spine surgery. Although the treatment of LSS depends on its severity, the optimal surgical technique towards decreasing the risk of adjacent segment disease (ASD) remains elusive. This study aimed to comparatively analyze spinal biomechanics with rigid and flexible fixation devices (i.e., rigid and dynamic posterolateral fusion (PLF) and interspinous process (ISP) devices) during daily activities. Methods: Using a validated parametric poroelastic finite element modeling approach, 8 subject-specific pre-operative models were developed, and their validity was evaluated. Parametric FE models of the lumbar spines were then regenerated based on post-operation images for (A) rigid PLF (B) dynamic PLF (C) rigid ISP device (Coflex) and (D) flexible ISP device (DIAM) at L4-L5 level. Biomechanical responses for instrumented and adjacent intervertebral discs (IVDs) were analyzed and compared subject to static and cyclic loading. Results: The preoperative models were well comparable with previous works in literature. The postoperative results for the PLF and Coflex rigid systems, demonstrated greater ROM; higher values of stress and strain in the AF region; and increased disc height and fluid loss at the adjacent levels, as compared with the pre-op models and the post-op results of the flexible systems (i.e., dynamic PLF and DIAM). The calculated forces on the facet joint were of smaller magnitude for the ISP devices as compared to the PLF, particularly during extension. Conclusions: This study demonstrates that the dynamic PLF construct and DIAM implants could be effective to maintain the natural poroelastic characteristics of adjacent IVDs, which could be beneficial for enhancing long-term clinical outcomes. FEM provides clinicians with an invaluable patient-specific quantitative tool for informed surgical planning and discerning follow-up management. (c) 2021 Published by Elsevier B.V.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Comparative biomechanical analyses of lower cervical spine post anterior fusion versus intervertebral disc arthroplasty: A geometrically patient-specific poroelastic finite element investigation
    Khalaf, Kinda
    Nikkhoo, Mohammad
    [J]. JOURNAL OF ORTHOPAEDIC TRANSLATION, 2022, 36 : 33 - 43
  • [2] The Biomechanical Response of the Lower Cervical Spine Post Laminectomy: Geometrically-Parametric Patient-Specific Finite Element Analyses
    Nikkhoo, Mohammad
    Cheng, Chih-Hsiu
    Wang, Jaw-Lin
    Niu, Chi-Chien
    Parnianpour, Mohamad
    Khalaf, Kinda
    [J]. JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2021, 41 (01) : 59 - 70
  • [3] The Biomechanical Response of the Lower Cervical Spine Post Laminectomy: Geometrically-Parametric Patient-Specific Finite Element Analyses
    Mohammad Nikkhoo
    Chih-Hsiu Cheng
    Jaw-Lin Wang
    Chi-Chien Niu
    Mohamad Parnianpour
    Kinda Khalaf
    [J]. Journal of Medical and Biological Engineering, 2021, 41 : 59 - 70
  • [4] Comparison of the effects of bilateral posterior dynamic and rigid fixation devices on the loads in the lumbar spine: a finite element analysis
    Rohlmann, Antonius
    Burra, Nagananda K.
    Zander, Thomas
    Bergmann, Georg
    [J]. EUROPEAN SPINE JOURNAL, 2007, 16 (08) : 1223 - 1231
  • [5] Comparison of the effects of bilateral posterior dynamic and rigid fixation devices on the loads in the lumbar spine: a finite element analysis
    Antonius Rohlmann
    Nagananda K. Burra
    Thomas Zander
    Georg Bergmann
    [J]. European Spine Journal, 2007, 16 : 1223 - 1231
  • [6] Biomechanical comparative analysis of conventional pedicle screws and cortical bone trajectory fixation in the lumbar spine: An in vitro and finite element study
    Beijing key laboratory for design and evaluation technology of advanced implantable & interventional medical devices, Beijing Advanced Innovation Center for Biomedical Engineering, School of Biological Science and Medical Engineering, Beihang University, Beijing, China
    不详
    不详
    不详
    [J]. Front. Bioeng. Biotechnol.,
  • [7] Biomechanical comparative analysis of conventional pedicle screws and cortical bone trajectory fixation in the lumbar spine: An in vitro and finite element study
    Pei, Baoqing
    Xu, Yangyang
    Zhao, Yafei
    Wu, Xueqing
    Lu, Da
    Wang, Haiyan
    Wu, Shuqin
    [J]. FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2023, 11
  • [8] Biomechanical Investigation Between Rigid and Semirigid Posterolateral Fixation During Daily Activities: Geometrically Parametric Poroelastic Finite Element Analyses
    Nikkhoo, Mohammad
    Lu, Meng-Ling
    Chen, Wen-Chien
    Fu, Chen-Ju
    Niu, Chi-Chien
    Lin, Yang-Hua
    Cheng, Chih-Hsiu
    [J]. FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2021, 9
  • [9] Biomechanical Study on the Efficacy of the Periacetabular Osteotomy using Patient-specific Finite Element Analysis
    Lee, Kyung-Jin
    Park, Sung-Jae
    Lee, Sung-Jae
    Naito, Masatoshi
    Kwon, Soon-Yong
    [J]. INTERNATIONAL JOURNAL OF PRECISION ENGINEERING AND MANUFACTURING, 2015, 16 (04) : 823 - 829
  • [10] Biomechanical study on the efficacy of the periacetabular osteotomy using Patient-specific finite element analysis
    Kyung-Jin Lee
    Sung-Jae Park
    Sung-Jae Lee
    Masatoshi Naito
    Soon-Yong Kwon
    [J]. International Journal of Precision Engineering and Manufacturing, 2015, 16 : 823 - 829