Bipolar saline-enhanced electrode for radiofrequency ablation:: Results of experimental study of in vivo porcine liver

被引:58
|
作者
Burdío, F
Güemes, A
Burdío, JM
Navarro, A
Sousa, R
Castiella, T
Cruz, I
Burzaco, O
Lozano, R
机构
[1] Univ Zaragoza, Hosp Clin Lozano Blesa, Dept Surg A, Zaragoza, Spain
[2] Univ Zaragoza, Hosp Clin Lozano Blesa, Dept Pathol, Zaragoza, Spain
[3] Univ Zaragoza, Fac Vet, Dept Elect Engn & Commun, E-50009 Zaragoza, Spain
[4] Univ Zaragoza, Fac Vet, Dept Anim Pathol & Surg, E-50009 Zaragoza, Spain
关键词
animals; experimental study; liver; interventional procedures; liver neoplasms; radiofrequency (RF) ablation;
D O I
10.1148/radiol.2292020978
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To evaluate whether a bipolar saline-enhanced radiofrequency (RF) ablation system embedded in one needle is able to consistently produce homogeneous and predictable areas of coagulation necrosis with or without the Pringle maneuver of vascular inflow occlusion. MATERIALS AND METHODS: RF ablation (480 kHz) of the liver was performed in 24 healthy pigs by means of laparotomy: group A (n = 5), 4-cm distance between electrodes 1 and 2; group B (n = 7); 4-cm distance and the Pringle maneuver; group C (n = 5); 2-cm distance; and group D (n = 7), 2-cm distance with the Pringle maneuver: Twenty percent NaCl solution was infused continuously at a rate of 100 mL/h via each electrode during the procedure. The pigs were followed up, and they were euthanized on the 7th day. Livers were removed for histologic assessment. Time, impedance; current, power output, specific voltage of the contacts, energy output, temperatures in the liver, volume of the lesion, and energy delivered per lesion volume were determined and compared among groups. Predictability of lesion volume was evaluated with the coefficient of variability. Mean values of the variables were compared among the groups by means of one-way analysis of variance or Kruskall-Wallis test. RESULTS: Impedance at the end of the RF ablation procedure was almost twofold lower than the corresponding initial value in all groups. In Pringle groups B and D, regular ellipsoids of coagulation necrosis were created (mean lesion volume, 149.50 cm(3) +/- 34.26 and 69.43 cm(3) +/- 15.48, respectively). In non-Pringle groups A and C, the shape of coagulation necrosis was influenced by the vessels encountered, and mean lesion size was lower than that in the Pringle groups (P < .01). The coefficient of variability of lesion size was lower in the Pringle groups (23% and 22%, respectively) than that in the non-Pringle groups (75% and 30%, respectively). CONCLUSION: The bipolar saline-enhanced RF ablation method produces homogeneous and predictable areas of coagulation necrosis between two electrodes, regardless of the distance between them, preferably with vascular inflow occlusion. (C) RSNA 2003.
引用
收藏
页码:447 / 456
页数:10
相关论文
共 50 条
  • [1] Bipolar saline-enhanced electrode for radiofrequency ablation: Results of experimental study of in vivo porcine liver (vol 229, pg 447, 2003)
    Burdio, F
    Guemes, A
    Burdio, JM
    Navarro, A
    Sousa, R
    Castiella, T
    Cru, I
    Burzaco, O
    Lozano, R
    RADIOLOGY, 2004, 230 (01) : 303 - 303
  • [2] Large hepatic ablation with bipolar saline-enhanced radiofrequency:: An experimental study in in vivo porcine liver with a novel approach
    Burdío, F
    Güemes, A
    Burdío, JM
    Navarro, A
    Sousa, R
    Castiella, T
    Cruz, I
    Burzaco, O
    Guirao, X
    Lozano, R
    JOURNAL OF SURGICAL RESEARCH, 2003, 110 (01) : 193 - 201
  • [3] Enhancement of radiofrequency ablation systems by means of bipolar saline perfused electrode.: In vivo porcine liver experimental study
    Navarro, AC
    Burdío, F
    Güemes, A
    Burdío, JM
    Sousa, R
    Tejero, E
    Castiella, T
    Cruz, I
    Lozano, R
    PROCEEDINGS OF THE XXXIX CONGRESS OF THE EUROPEAN SOCIETY FOR SURGICAL RESEARCH, 2004, : 113 - 120
  • [4] Hepatic lesion ablation with bipolar saline-enhanced radiofrequency in the audible spectrum
    Burdío, F
    Güemes, A
    Burdío, JM
    Castiella, T
    De Gregorio, MA
    Lozano, R
    Livraghi, T
    ACADEMIC RADIOLOGY, 1999, 6 (11) : 680 - 686
  • [5] A comparative experimental study of the in-vitro efficiency of hypertonic saline-enhanced hepatic bipolar and monopolar radiofrequency ablation
    Lee, JM
    Han, JK
    Kim, SH
    Sohn, KL
    Lee, KH
    Ah, SK
    Choi, BI
    KOREAN JOURNAL OF RADIOLOGY, 2003, 4 (03) : 163 - 169
  • [6] Automated saline-enhanced radiofrequency thermal ablation:: Initial results in ex vivo bovine livers
    Schmidt, D
    Trübenbach, J
    Brieger, J
    Koenig, C
    Putzhammer, H
    Duda, SH
    Claussen, CD
    Pereira, PL
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2003, 180 (01) : 163 - 165
  • [7] Premature roll-off in radiofrequency ablation using bipolar saline-enhanced electrodes
    Burdio, F
    Navarro, A
    Sousa, R
    Guemes, A
    Burdio, JM
    Tejero, E
    Lozano, R
    EUROPEAN RADIOLOGY, 2005, 15 (07) : 1495 - 1496
  • [8] Saline-enhanced radiofrequency thermal ablation of the lung: A feasibility study in rabbits
    Lee, JM
    Kim, SW
    Li, CA
    Youk, JH
    Kim, YK
    Jin, ZW
    Chung, MJ
    Lee, MS
    KOREAN JOURNAL OF RADIOLOGY, 2002, 3 (04) : 245 - 253
  • [9] Ex vivo experiment of saline-enhanced hepatic bipolar radiofrequency ablation with a perfused needle electrode: Comparison with conventional monopolar and simultaneous monopolar modes
    Lee, JM
    Kim, SH
    Han, JK
    Sohn, KL
    Choi, BI
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2005, 28 (03) : 338 - 345
  • [10] Saline-enhanced radiofrequency ablation of breast tissue -: An in vitro feasibility study
    Böhm, T
    Hilger, I
    Müller, W
    Reichenbach, JR
    Fleck, M
    Kaiser, WA
    INVESTIGATIVE RADIOLOGY, 2000, 35 (03) : 149 - 157