Characterization of the ringing artifacts in rotator-based reconstruction with Monte Carlo-based resolution compensation for PET

被引:5
|
作者
Zhang, Long [1 ]
Staelens, Steven [1 ]
Van Holen, Roel [1 ]
Verhaeghe, Jeroen [2 ]
Vandenberghe, Stefaan
机构
[1] Univ Ghent, MEDISIP, IBBT, B-9000 Ghent, Belgium
[2] McGill Univ, McConnell Brain Imaging Ctr, Montreal Neurol Inst, Montreal, PQ H3A 2B4, Canada
关键词
ringing artifacts; Monte Carlo; resolution modeling; Gaussian rotator; IMAGE-RECONSTRUCTION; ITERATIVE RECONSTRUCTION; MATHEMATICAL REMOVAL; RESPONSE FUNCTION; MATRIX; ALGORITHMS; SIMULATION; SPECT; MODEL;
D O I
10.1118/1.3478275
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Studies have shown that Monte Carlo-based reconstruction could effectively improve the image quality of positron emission tomography. The authors have previously used a Gaussian rotator-based algorithm to efficiently reduce the computational cost for system matrix (SM) calculation and to meet the large memory requirements for SM storage. However, pronounced ringing artifacts were observed in the reconstructed image. In this article, the authors investigated and characterized these artifacts. Methods: The authors proposed an "ideal" rotator and used it as a baseline in the artifacts evaluation. This ideal rotator produces perfectly rotated images. The Gaussian rotator method was evaluated by a full system model and a partial system model without positron range and acolinearity, which could be compensated for by the blurring of the Gaussian rotator for F-18 studies. Noiseless data, Monte Carlo simulation data, as well as acquired experimental data were used to quantitatively characterize the behavior of the artifacts. Results: The study of the noiseless data indicated that the artifacts were mainly attributed to the rotator, which further blurred the simulated system responses. The simulation study suggested that the artifacts become less pronounced and not quantitatively significant in practice. This result is consistent with the experimental data study. Better contrast recovery was achieved with an overcompensated system model. Traditionally, an undercompensated system model was preferred to avoid artifacts. The authors' studies suggest that the Gaussian rotator with the full system model yields the best image quality among the evaluated methods with considerably reduced quantitative error and quantitatively insignificant artifacts in practice. Conclusions: The authors' investigation indicated that a moderately overcompensated system model (about 2 mm FWHM in this study) yielded better contrast recovery and quantitatively insignificant artifacts in practice. (c) 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3478275]
引用
收藏
页码:4648 / 4660
页数:13
相关论文
共 50 条
  • [1] Monte Carlo-based analysis of PET scatter components
    Adam, LE
    Bellemann, ME
    Brix, G
    Lorenz, WJ
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1996, 37 (12) : 2024 - 2029
  • [2] Acceleration of Monte Carlo-based scatter compensation for cardiac SPECT
    Sohlberg, A.
    Watabe, H.
    Iida, H.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (14): : N277 - N285
  • [3] Monte Carlo-based SPECT reconstruction within the SIMIND framework
    Gustafsson, Johan
    Brolin, Gustav
    Ljungberg, Michael
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (24):
  • [4] Efficient fully 3-D iterative SPECT reconstruction with Monte Carlo-based scatter compensation
    Beekman, FJ
    de Jong, HWAM
    van Geloven, S
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (08) : 867 - 877
  • [5] Fast and memory-efficient Monte Carlo-based image reconstruction for whole-body PET
    Zhang, Long
    Staelens, Steven
    Van Holen, Roel
    De Beenhouwer, Jan
    Verhaeghe, Jeroen
    Kawrakow, Iwan
    Vandenberghe, Stefaan
    [J]. MEDICAL PHYSICS, 2010, 37 (07) : 3667 - 3676
  • [6] Quantitative Monte Carlo-Based 90Y SPECT Reconstruction
    Elschot, Mattijs
    Lam, Marnix G. E. H.
    van den Bosch, Maurice A. A. J.
    Viergever, Max A.
    de Jong, Hugo W. A. M.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2013, 54 (09) : 1557 - 1563
  • [7] EVALUATION OF THE SPATIAL RESOLUTION IN MONTE CARLO-BASED SPECT/CT RECONSTRUCTION OF 111IN-OCTREOTIDE IMAGES
    Wikberg, Emma
    van Essen, Martijn
    Ryden, Tobias
    Svensson, Johanna
    Gjertsson, Peter
    Bernhardt, Peter
    [J]. RADIATION PROTECTION DOSIMETRY, 2021, 195 (3-4) : 319 - 326
  • [8] Quantitative Monte Carlo-based holmium-166 SPECT reconstruction
    Elschot, Mattijs
    Smits, Maarten L. J.
    Nijsen, Johannes F. W.
    Lam, Marnix G. E. H.
    Zonnenberg, Bernard A.
    van den Bosch, Maurice A. A. J.
    Viergever, Max A.
    de Jong, Hugo W. A. M.
    [J]. MEDICAL PHYSICS, 2013, 40 (11)
  • [9] 3-d Monte Carlo-based scatter compensation in quantitative I-131SPECT reconstruction
    Dewaraja, YK
    Ljungberg, M
    Fessler, JA
    [J]. 2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3653 - 3657
  • [10] 3-D Monte Carlo-based scatter compensation in quantitative I-131SPECT reconstruction
    Dewaraja, YK
    Ljungberg, M
    Fessler, JA
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (01) : 181 - 188