Study on bubble detectors used as personal neutron dosimeters

被引:7
|
作者
Zhang Guiying [1 ]
Ni Bangfa [1 ]
Li Li [1 ]
Lv Peng [1 ]
Tian Weizhi [1 ]
Wang Zhiqiang [1 ]
Zhang Chunbao [1 ]
Luo Hailong [1 ]
Jiang Shunli [1 ]
Wang Pingsheng [1 ]
Huang Donghui [1 ]
Liu Cunxiong [1 ]
Xiao Caijin [1 ]
机构
[1] China Inst Atom Energy, Beijing 102413, Peoples R China
关键词
Neutron bubble detectors; Relevant equipments; Calibrations; On-site test; SUPERHEATED DROP DETECTOR; X-RAY; ENERGY; SPECTROMETRY; RADIOTHERAPY; LIQUID;
D O I
10.1016/j.apradiso.2011.05.008
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Neutron bubble detector is so far the only personal neutron dosimeter satisfying the energy response criteria of the International Committee of Radiation Protection 60 (ICRP 60). This paper presents our studies on neutron bubble detectors including the manufacture, the relevant equipments, the basic calibrations and on-site tests for monitoring personal neutron dose. The results of calibrations show that the highest sensitivity so far manufactured by the authors reaches about 4 bubbles/mu Sv, the correlation coefficient of dose response is 0.99, and the in-batch consistency and reproducibility are up to the ISO standards. The results of on-site test show that the in-batch consistency and between-batch consistency are within 15% relative standard uncertainty. The results are directly readable. The detectors are portable, especially suitable for on-site neutron dose monitoring in n-gamma mixed radiation fields. (C) 2011 Published by Elsevier Ltd.
引用
收藏
页码:1453 / 1458
页数:6
相关论文
共 50 条
  • [1] Preliminary Study on Application of Neutron Bubble Detectors as Personal Neutron Dosimeters in Nuclear Well Logging
    Zhang, Guiying
    Deng, Jun
    Cao, Lei
    Zhang, Haiqing
    Ni, Bangfa
    Tian, Weizhi
    Wang, Pingsheng
    Huang, Donghui
    Liu, Cunxiong
    Xiao, Caijin
    Zhao, Changjun
    [J]. HEALTH PHYSICS, 2011, 100 (02): : S30 - S34
  • [2] COMPARISON OF NEUTRON MEASUREMENTS AT LINACS USING BUBBLE DOSIMETERS TO OTHER NEUTRON DETECTORS
    NELSON, ME
    GORDON, R
    [J]. RADIATION PROTECTION DOSIMETRY, 1993, 47 (1-4) : 547 - 550
  • [3] Use of bubble detectors in personal neutron dosimetry
    Spurny, F
    Vlcek, B
    [J]. JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1996, 209 (02): : 275 - 278
  • [4] In situ measurement of neutron and gamma radiation exposures during intercontinental flights using electronic personal dosimeters and bubble detectors
    Mukherjee, B.
    Makowsk, D.
    Mares, V.
    Rybka, D.
    Simrock, S.
    [J]. MIXDES 2007: PROCEEDINGS OF THE 14TH INTERNATIONAL CONFERENCE ON MIXED DESIGN OF INTEGRATED CIRCUITS AND SYSTEMS:, 2007, : 110 - +
  • [5] ROLE OF NEUTRON BUBBLE DOSIMETERS
    RIEL, G
    [J]. RADIATION PROTECTION DOSIMETRY, 1993, 47 (1-4) : 543 - 545
  • [6] PERSONAL ALARM NEUTRON DOSIMETERS
    GIBSON, JAB
    [J]. RADIATION PROTECTION DOSIMETRY, 1985, 10 (1-4) : 197 - 205
  • [7] Study of bubble damage detectors for neutron detection
    Tu, CQ
    Guo, SL
    Wang, YL
    Hao, XH
    Chen, CM
    Su, JL
    [J]. RADIATION MEASUREMENTS, 1997, 28 (1-6) : 159 - 162
  • [8] Bubble neutron detector can be used in personal dose measurement
    Zhang, Hong-Jun
    Tian, Qing-Qing
    [J]. Hedianzixue Yu Tance Jishu/Nuclear Electronics and Detection Technology, 2004, 24 (06):
  • [9] ANGULAR DEPENDENT RESPONSE OF NUCLEAR TRACK DETECTORS USED AS FAST-NEUTRON DOSIMETERS
    PITT, E
    WERNER, B
    [J]. KERNTECHNIK, 1988, 52 (01) : 9 - 13
  • [10] ALBEDO DOSIMETERS FOR PERSONAL NEUTRON DOSIMETRY
    BUSUOLI, G
    CAVALLINI, A
    CIVOLANI, O
    LEMBO, L
    [J]. ENERGIA NUCLEARE, 1976, 23 (10): : 518 - 521