Vacuum leak problem in low energy of pelletron

被引:0
|
作者
Gargari, S [1 ]
Chopra, S [1 ]
Joshi, R [1 ]
Ojha, S [1 ]
Sota, M [1 ]
Barua, P [1 ]
Patel, V [1 ]
Prasad, J [1 ]
Kumar, R [1 ]
Singh, B [1 ]
Panwar, NS [1 ]
Singh, MP [1 ]
Datta, SK [1 ]
Kanjilal, D [1 ]
机构
[1] Ctr Nucl Sci, New Delhi 110067, India
关键词
D O I
暂无
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
During unit wise conditioning of unit # 8, the vacuum started deteriorating inside the tube after a spark. The RGA leading was taken and it was found out that residual gas inside tube was sulphur hexafluoride. A leak was detected in second tube of unit number eight in between electrode 6 to 8. Leak was sealed with the sealant. Again leak check was done and no leak was found. The tank was closed and conditioning was started again. During the same unit number eight conditioning, leak developed again followed by a spark. So the damaged tube was replaced with a new accelerator tube. During the installation time the alignment of the machine was taken care. Again leak checking was done and the tube was baked properly. The tank was closed again and this particular unit was conditioned for about four days. The maximum voltage it has attained was 1.1 MV.
引用
收藏
页码:78 / 80
页数:3
相关论文
共 50 条
  • [1] LOW COST VACUUM LEAK DETECTOR.
    Sharma, J.K.N.
    Gupta, A.C.
    1600, (36):
  • [2] A LOW-COST VACUUM LEAK DETECTOR
    SHARMA, JKN
    GUPTA, AC
    VACUUM, 1986, 36 (05) : 279 - 280
  • [3] A NEW DESIGN FOR THE LOW-ENERGY OPTICS OF THE LUND PELLETRON ACCELERATOR
    HELLBORG, R
    HAKANSSON, K
    SKOG, G
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 287 (1-2): : 161 - 165
  • [4] Vanderbilt Pelletron - Low Energy Protons and Other Ions for Radiation Effects on Electronics
    McCurdy, Michael W.
    Mendenhall, Marcus H.
    Reed, Robert A.
    Rogers, Bridget R.
    Weller, Robert A.
    Schrimpf, Ronald D.
    2015 IEEE RADIATION EFFECTS DATA WORKSHOP (REDW), 2015, : 146 - 151
  • [5] LOW-RANGE FLOWMETERS FOR USE WITH VACUUM AND LEAK STANDARDS
    MCCULLOH, KE
    TILFORD, CR
    EHRLICH, CD
    LONG, FG
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1987, 5 (03): : 376 - 381
  • [6] On the problem of vacuum energy in brane theories
    Gurwich, Ilya
    Rubin, Shimon
    Davidson, Aharon
    PHYSICS LETTERS B, 2009, 679 (05) : 515 - 518
  • [7] Vacuum leak testing
    Manley, JH
    Haworth, LJ
    Luebke, EA
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1939, 10 (12): : 389 - 390
  • [8] CALCULATIONS OF THE BEAM TRANSPORT THROUGH THE LOW-ENERGY SIDE OF THE LUND PELLETRON ACCELERATOR
    DYMNIKOV, A
    HELLBORG, R
    PALLON, J
    SKOG, G
    YANG, C
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 334 (2-3): : 266 - 270
  • [9] An adjustable vacuum leak
    Smythe, WR
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1936, 7 (11): : 435 - 435
  • [10] VACUUM LEAK DETECTION
    LAFORTE, JT
    SEMICONDUCTOR PRODUCTS AND SOLID STATE TECHNOLOGY, 1965, 8 (12): : 25 - &