Skin dose measurements on patients for diagnostic and interventional neuroradiology: A multicentre study

被引:33
|
作者
Struelens, L
Vanhavere, F
Bosmans, H
Van Loon, R
Mol, H
机构
[1] CEN SCK, Belgian Nuck Res Ctr, B-2400 Mol, Belgium
[2] UZ GHB Leuven, Dept Radiol, B-3000 Louvain, Belgium
[3] VUB, B-1050 Brussels, Belgium
[4] EHSAL, European Inst Higher Educ, Dept Med Imaging, B-1000 Brussels, Belgium
关键词
D O I
10.1093/rpd/nch537
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
In Belgium, the number of interventional neuroradiology procedures for the treatment of vascular lesions in the brain and neck region is increasing. These procedures, however, are associated with high doses of X rays. The present study reports on the skin dose measurements on patients undergoing these procedures. Grids of thermoluminescence dosemeters were attached on the head of the patient. The exposure parameters of the X-ray systems and the clinical procedures were fully documented. We searched for correlations between these parameters and the skin doses. While for the diagnostic procedure, the dose threshold of 2 Gy for deterministic effects was not reached, this situation was much different for the embolisations. For angiography of the carotid arteries, no skin doses were measured > 320 mGy. For the cerebral embolisations, maximum skin doses up to 5.4 Gy were measured. Doses to the patients varied largely between different hospitals and within the same hospital for similar procedures. On the one hand, the complexity of the pathology for interventional procedures was responsible for the large variability in dose. On the other hand, large differences in clinical protocol and technical parameters of the X-ray systems, explaining the dose variations, were also observed. A correlation was found between the maximum skin dose measured on a patient and the total dose-area product (DAP) value for cerebral embolisations. This correlation makes it possible to estimate the maximum skin dose from these DAP values and to determine a trigger level. In conclusion, management of patient doses in interventional radiology requires training, specialisation and well-documented procedure guidelines.
引用
收藏
页码:143 / 146
页数:4
相关论文
共 50 条
  • [1] Diagnostic neuroradiology for the interventional neuroradiologist
    Pereira, Vitor Mendes
    Vargas, Maria Isabel
    Marcos, Ana
    Bijlenga, Philippe
    Narata, Ana Paula
    Haller, Sven
    Lovblad, Karl-Olof
    [J]. WORLD JOURNAL OF RADIOLOGY, 2013, 5 (11): : 386 - 397
  • [2] A study of the relationship between peak skin dose and cumulative air kerma in interventional neuroradiology and cardiology
    Neil, S.
    Padgham, C.
    Martin, C. J.
    [J]. JOURNAL OF RADIOLOGICAL PROTECTION, 2010, 30 (04) : 659 - 672
  • [3] Effective Dose and Thyroid Dose in Interventional Neuroradiology
    Magill, D.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)
  • [4] Absorbed dose and deterministic effects to patients from interventional neuroradiology
    Mooney, RB
    McKinstry, CS
    Kamel, HAM
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2000, 73 (871): : 745 - 751
  • [5] EUROPEAN SOCIETY OF NEURORADIOLOGY Diagnostic and Interventional
    不详
    [J]. NEURORADIOLOGY, 2024, 66 (SUPPL 1) : 1 - 179
  • [6] A method for a real time estimation of entrance skin dose distribution in interventional neuroradiology
    Rampado, O
    Ropolo, R
    [J]. MEDICAL PHYSICS, 2004, 31 (08) : 2356 - 2361
  • [7] Effective and organ doses in patient undergoing interventional neuroradiology procedures: A multicentre study
    D'Alessio, Andrea
    Strocchi, Sabina
    Dalmasso, Federico
    Cannillo, Barbara
    Matheoud, Roberta
    Ponzetti, Alex
    Aimonetto, Stefania
    Cernigliaro, Massimiliano
    Azzalin, Giulia
    Giorgianni, Andrea
    Natrella, Massimiliano
    Carriero, Alessandro
    Guzzardi, Giuseppe
    Brambilla, Marco
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2024, 122
  • [8] Monte Carlo investigation of backscatter factors for skin dose determination in interventional neuroradiology procedures
    Omar, Artur
    Benmakhlouf, Hamza
    Marteinsdottir, Maria
    Bujila, Robert
    Nowik, Patrik
    Andreo, Pedro
    [J]. MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING, 2014, 9033
  • [9] ASPIRIN AND SYSTEMIC HEPARINIZATION IN DIAGNOSTIC AND INTERVENTIONAL NEURORADIOLOGY
    DEBRUN, GM
    VINUELA, FV
    FOX, AJ
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 1982, 3 (03) : 337 - 340
  • [10] ASPIRIN AND SYSTEMIC HEPARINIZATION IN DIAGNOSTIC AND INTERVENTIONAL NEURORADIOLOGY
    DEBRUN, GM
    VINUELA, FV
    FOX, AJ
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1982, 139 (01) : 139 - 142