Ionic Liquids as a Reference Material Candidate for the Quick Performance Check of Energy Dispersive X-ray Spectrometers for the Low Energy Range below 1 keV

被引:5
|
作者
Holzweber, Markus [1 ,2 ]
Unger, Wolfgang E. S. [1 ]
Hodoroaba, Vasile-Dan [1 ]
机构
[1] BAM Fed Inst Mat Res & Testing, Div Surface Anal & Interfacial Chem 6 1, Unter Eichen 44-46, D-12203 Berlin, Germany
[2] Vienna Univ Technol, Inst Chem Technol & Analyt, Getreidemarkt 9-164AC, A-1060 Vienna, Austria
基金
奥地利科学基金会;
关键词
EFFICIENCY;
D O I
10.1021/acs.analchem.6b01444
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Ionic liquids (ILs) are proposed as simple and efficient test materials to evaluate the performance of energy dispersive X-ray spectrometers (EDS) in the low energy range below 1 keV. By only one measurement, C K alpha, N K alpha, O K alpha, and F K alpha X-ray lines can be excited. Additionally, the S K alpha line at 2.3 keV and, particularly, the S L series at 149 eV complete the picture with X-ray lines offered by the selected ILs. The well-known (certifiable) elemental composition of the ILs selected in the present study can be used to check the accuracy of results produced with the available EDS quantification routines in the low energy range, simultaneously, for several low atomic number elements. A comparison with other reference materials in use for testing the performance of EDS in the low energy range is included.
引用
收藏
页码:6967 / 6970
页数:4
相关论文
共 29 条
  • [1] The determination of the efficiency of energy dispersive X-ray spectrometers by a new reference material
    Alvisi, Marco
    Blome, Markus
    Griepentrog, Michael
    Hodoroaba, Vasile-Dan
    Karduck, Peter
    Mostert, Marco
    Nacucchi, Michele
    Procop, Mathias
    Rohde, Martin
    Scholze, Frank
    Statham, Peter
    Terborg, Ralf
    Thiot, Jean-Francois
    [J]. MICROSCOPY AND MICROANALYSIS, 2006, 12 (05) : 406 - 415
  • [2] A Test Material and a quick Procedure for the Performance Check of X-Ray Spectrometers attached to the SEM
    Procop, M.
    Hodoroaba, V. -D.
    [J]. MICROSCOPY AND MICROANALYSIS, 2009, 15 : 1120 - 1121
  • [3] New reference and test materials for the characterization of energy dispersive X-ray spectrometers at scanning electron microscopes
    Rackwitz, Vanessa
    Krumrey, Michael
    Laubis, Christian
    Scholze, Frank
    Hodoroaba, Vasile-Dan
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2015, 407 (11) : 3045 - 3053
  • [4] New reference and test materials for the characterization of energy dispersive X-ray spectrometers at scanning electron microscopes
    Vanessa Rackwitz
    Michael Krumrey
    Christian Laubis
    Frank Scholze
    Vasile-Dan Hodoroaba
    [J]. Analytical and Bioanalytical Chemistry, 2015, 407 : 3045 - 3053
  • [5] IMPROVED PERFORMANCE OF HGI2 SPECTROMETERS IN LOW-ENERGY X-RAY REGION
    SEIBT, W
    SLAPA, M
    HUTH, GC
    [J]. NUCLEAR INSTRUMENTS & METHODS, 1976, 135 (03): : 573 - 576
  • [6] Optical performance of materials for X-ray refractive optics in the energy range 8-100keV
    Serebrennikov, Dmitry
    Clementyev, Evgeny
    Semenov, Alexander
    Snigirev, Anatoly
    [J]. JOURNAL OF SYNCHROTRON RADIATION, 2016, 23 : 1315 - 1322
  • [7] SPECTRUM OF COSMIC X-RAY AND GAMMA-RAY BACKGROUND IN ENERGY RANGE 1 KEV-1 MEV
    KASTURIRANGAN, K
    RAO, UR
    [J]. ASTROPHYSICS AND SPACE SCIENCE, 1972, 15 (01) : 161 - +
  • [8] SOLEX:: a tunable monochromatic X-ray source in the 1-20 keV energy range for metrology
    Bonnelle, C
    Jonnard, P
    André, JM
    Avila, A
    Laporte, D
    Ringuenet, H
    Lépy, MC
    Plagnard, J
    Ferreux, L
    Protas, JC
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 516 (2-3): : 594 - 601
  • [9] Performance of the charge-coupled device for direct X-ray detection in the energy range of 1-9 keV at the synchrotron radiation facility
    Tsunemi, Hiroshi
    Kawai, Shin'ichi
    Hayashida, Kiyoshi
    [J]. Japanese Journal of Applied Physics, Part 1: Regular Papers and Short Notes and Review Papers, 1991, 30 (06): : 1299 - 1302
  • [10] Alanine films for EPR dosimetry of low-energy (1-30 keV) X-ray photons
    D'Oca, M. C.
    Marrale, M.
    Abbene, L.
    Bartolotta, A.
    Collura, G.
    D'Errico, F.
    Principato, F.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2019, 459 : 1 - 6