In Vitro Models of the Fetal Lung Comparison of Lung Volume Measurements With 3-Dimensional Sonography and Magnetic Resonance Imaging

被引:0
|
作者
Kehl, Sven [1 ]
Zirulnik, Anna [1 ]
Debus, Angelika [2 ]
Suetterlin, Marc [1 ]
Siemer, Joern [1 ]
Neff, Wolfgang [2 ]
机构
[1] Mannheim Univ Hosp, Dept Obstet & Gynecol, Mannheim, Germany
[2] Mannheim Univ Hosp, Dept Radiol, Mannheim, Germany
关键词
congenital diaphragmatic hernia; fetal lung volumetry; magnetic resonance imaging; 3-dimensional sonography; CONGENITAL DIAPHRAGMATIC-HERNIA; EXTRACORPOREAL MEMBRANE-OXYGENATION; LOGISTIC-REGRESSION ANALYSIS; TO-HEAD RATIO; ULTRASOUND; ULTRASONOGRAPHY; PREDICTION; FETUSES; RELIABILITY; SURVIVAL;
D O I
暂无
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Objectives-Three-dimensional (3D) sonography is an established volumetric method in gynecology and obstetrics. The aim of this study was to investigate the variability of 3D sonographic measurements and their accuracy in comparison with magnetic resonance imaging (MM) for assessing fetal lung volume using in vitro lung models. Methods-Twenty-three in vitro lung models with randomly defined volumes ranging from 1 to 60 mL were made from gelatin with plastic sheaths, manually molded into the shape of fetal lungs. The models were measured using 3D sonography and MRI. The 3D sonographic volumes were calculated using the rotational technique with angles of 6 degrees and 30 degrees. Multiplanar T2-weighted sequences were used for the MM measurements. The percentage error and absolute percentage error were calculated for each method, and intraobserver and interobserver variability in 3D sonographic measurements was assessed with intraclass correlation coefficients (ICCs). Agreement between calculated and real volumes using the limits of agreement method was also evaluated. Results-The ICCs for the rotation angles indicated very good intraobserver and interobserver variability (6 degrees, 0.995 and 0.996; 30 degrees, 0.997 and 0.985). No systematic errors were observed in the mean percentage errors for 3D sonographic measurements or MM volumetry. The lowest median absolute percentage error (1.76) was obtained with MM volumetry, significantly lower than the values for sonography (6 degrees, 5.00; P < .001; 30 degrees, 5.49; P < .001). There were no significant differences in absolute percentage errors, between the rotation angles (P = .82) and no significant differences in limits of agreement between 3D sonography and MRI (6 degrees, P = .76; 30 degrees, P = .39). Conclusions-Three-dimensional sonographic volumetry was almost as accurate as MRI in this in vitro model and can be regarded as a good alternative method. Further research is needed to confirm these findings in vivo and to assess the prognostic value in fetuses with lung hypoplasia (eg, congenital diaphragmatic hernias).
引用
收藏
页码:1085 / 1091
页数:7
相关论文
共 50 条
  • [1] Assessment of Lung Volume by 3-Dimensional Sonography and Magnetic Resonance Imaging in Fetuses With Congenital Diaphragmatic Hernias
    Kehl, Sven
    Kalk, Anna Luisa
    Eckert, Sven
    Schaible, Thomas
    Suetterlin, Marc
    Neff, Wolfgang
    Siemer, Joern
    JOURNAL OF ULTRASOUND IN MEDICINE, 2011, 30 (11) : 1539 - 1545
  • [2] Reproducibility of fetal lung volume measurements with 3-dimensional ultrasonography
    Sabogal, JC
    Becker, E
    Bega, G
    Komwilaisak, R
    Berghella, V
    Weiner, S
    Tolosa, J
    JOURNAL OF ULTRASOUND IN MEDICINE, 2004, 23 (03) : 347 - 352
  • [3] Fetal lung volume estimated by 3-dimensional ultrasonography and magnetic resonance imaging in cases with isolated congenital diaphragmatic hernia
    Ruano, R
    Joubin, L
    Sonigo, P
    Benachi, A
    Aubry, MC
    Thalabard, JC
    Brunelle, F
    Dumez, Y
    Dommergues, M
    JOURNAL OF ULTRASOUND IN MEDICINE, 2004, 23 (03) : 353 - 358
  • [4] Influence of Different Rotation Angles in Assessment of Lung Volumes by 3-Dimensional Sonography in Comparison to Magnetic Resonance Imaging in Healthy Fetuses
    Kehl, Sven
    Eckert, Sven
    Suetterlin, Marc
    Neff, K. Wolfgang
    Siemer, Joern
    JOURNAL OF ULTRASOUND IN MEDICINE, 2011, 30 (06) : 819 - 825
  • [5] Fetal lung volume in fetuses with urinary tract malformations: Comparison by 2D-, 3D-sonography and magnetic resonance imaging
    Araujo Junior, Edward
    De Oliveira, Patricia Soares
    Machado Nardozza, Luciano Marcondes
    Simioni, Christiane
    Rolo, Liliam Cristine
    Goldman, Susan Menasce
    Szejnfeld, Jacob
    Moron, Antonio Fernandes
    JOURNAL OF MATERNAL-FETAL & NEONATAL MEDICINE, 2010, 23 (01): : 60 - 68
  • [6] Fetal lung volume: three-dimensional ultrasonography compared with magnetic resonance imaging
    Gerards, F. A.
    Twisk, J. W. R.
    Bakker, M.
    Barkhof, F.
    Van Vugt, J. M. G.
    ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2007, 29 (05) : 533 - 536
  • [7] Prenatal diagnosis of a fetal epidural hematoma using 2-and 3-dimensional sonography and magnetic resonance imaging
    Muench, Michael V.
    Zheng, Min
    Bilica, Paula M.
    Canterino, Joseph C.
    JOURNAL OF ULTRASOUND IN MEDICINE, 2008, 27 (09) : 1369 - 1373
  • [8] Normal ranges for fetal lung volume measured by 3-dimensional ultrasound
    Gerards, F
    Twisk, J
    Vugt, JMG
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 2003, 189 (06) : S221 - S221
  • [9] Prenatal Diagnosis of Seckel Syndrome on 3-Dimensional Sonography and Magnetic Resonance Imaging
    Napolitano, Raffaele
    Maruotti, Giuseppe M.
    Quarantelli, Mario
    Martinelli, Pasquale
    Paladini, Dario
    JOURNAL OF ULTRASOUND IN MEDICINE, 2009, 28 (03) : 369 - 374
  • [10] Fetal lung volume measurement by magnetic resonance imaging in congenital diaphragmatic hernia
    Mahieu-Caputo, D
    Sonigo, P
    Dommergues, M
    Fournet, JC
    Thalabard, JC
    Abarca, C
    Benachi, A
    Brunelle, F
    Dumez, Y
    BRITISH JOURNAL OF OBSTETRICS AND GYNAECOLOGY, 2001, 108 (08): : 863 - 868