Detectability of regional lung ventilation with flat-panel detector-based dynamic radiography

被引:14
|
作者
Tanaka, Rie [1 ]
Sanada, Shigeru [1 ]
Okazaki, Nobuo [2 ]
Kobayashi, Takeshi [3 ]
Suzuki, Masayuki [1 ]
Matsui, Takeshi [4 ]
Matsui, Osamu [4 ]
机构
[1] Kanazawa Univ, Grad Sch Med Sci, Kanazawa, Ishikawa 9200942, Japan
[2] Ohtsuka Ryogoku Clin, Dept Internal Med, Tokyo 130026, Japan
[3] Kanazawa Univ, Ishikawa Prefectural Cent Hosp, Dept Radiol, Kanazawa, Ishikawa 9208530, Japan
[4] Kanazawa Univ Hosp, Dept Radiol, Kanazawa, Ishikawa 9208642, Japan
关键词
digital imaging; functional imaging; computer analysis; chest radiographs; flat-panel detector; FPD; ventilation;
D O I
10.1007/s10278-007-9017-8
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study was performed to investigate the ability of breathing chest radiography using flat-panel detector (FPD) to quantify relative local ventilation. Dynamic chest radiographs during respiration were obtained using a modified FPD system. Imaging was performed in three different positions, ie, standing and right and left decubitus positions, to change the distribution of local ventilation. We measured the average pixel value in the local lung area. Subsequently, the interframe differences, as well as difference values between maximum inspiratory and expiratory phases, were calculated. The results were visualized as images in the form of a color display to show more or less x-ray translucency. Temporal changes and spatial distribution of the results were then compared to lung physiology. In the results, the average pixel value in each lung was associated with respiratory phase. In all positions, respiratory changes of pixel value in the lower area were greater than those in the upper area (P<0.01), which was the same tendency as the regional differences in ventilation determined by respiratory physiology. In addition, in the decubitus position, it was observed that areas with large respiratory changes in pixel value moved up in the vertical direction during expiration, which was considered to be airway closure. In conclusion, breathing chest radiography using FPD was shown to be capable of quantifying relative ventilation in local lung area and detecting regional differences in ventilation and timing of airway closure. This method is expected to be useful as a new diagnostic imaging modality for evaluating relative local ventilation.
引用
收藏
页码:109 / 120
页数:12
相关论文
共 50 条
  • [1] Detectability of Regional Lung Ventilation with Flat-panel Detector-based Dynamic Radiography
    Rie Tanaka
    Shigeru Sanada
    Nobuo Okazaki
    Takeshi Kobayashi
    Masayuki Suzuki
    Takeshi Matsui
    Osamu Matsui
    [J]. Journal of Digital Imaging, 2008, 21
  • [2] Dynamic chest radiography with a flat-panel detector (FPD): Ventilation-perfusion study
    Tanaka, R.
    Sanada, S.
    Fujimura, M.
    Yasui, M.
    Tsuji, S.
    Hayashi, N.
    Okamoto, H.
    Nanbu, Y.
    Matsui, O.
    [J]. MEDICAL IMAGING 2011: BIOMEDICAL APPLICATIONS IN MOLECULAR, STRUCTURAL, AND FUNCTIONAL IMAGING, 2011, 7965
  • [3] Development and characterization of a flat-panel detector-based microtomography system
    Lee, SC
    Kim, HK
    Chun, IK
    Cho, MH
    Cho, MH
    Lee, SY
    [J]. ADVANCES IN NONDESTRUCTIVE EVALUATION, PT 1-3, 2004, 270-273 : 245 - 251
  • [4] Comparison of dynamic flat-panel detector-based chest radiography with nuclear medicine ventilation-perfusion imaging for the evaluation of pulmonary function: A clinical validation study
    Tanaka, Rie
    Matsumoto, Isao
    Tamura, Masaya
    Takata, Munehisa
    Kasahara, Kazuo
    Ohkura, Noriyuki
    Inoue, Dai
    Matsuura, Yukihiro
    [J]. MEDICAL PHYSICS, 2020, 47 (10) : 4800 - 4809
  • [5] Development of a selenium-based flat-panel detector for radiography and fluoroscopy
    Yamada, S
    Tsukamoto, A
    Tanaka, M
    Asahina, H
    Nishiki, M
    Suzuki, K
    Uematsu, Y
    [J]. CAR '98 - COMPUTER ASSISTED RADIOLOGY AND SURGERY, 1998, 1165 : 542 - 547
  • [6] Imaging and Dosimetric Study on Direct Flat-Panel Detector-Based Digital Mammography System
    Sharma, Reena
    Sharma, S. D.
    Sarkar, P. S.
    Datta, D.
    [J]. JOURNAL OF MEDICAL PHYSICS, 2018, 43 (04) : 255 - 263
  • [7] Dynamic chest radiography: flat-panel detector (FPD) based functional X-ray imaging
    Tanaka R.
    [J]. Radiological Physics and Technology, 2016, 9 (2) : 139 - 153
  • [8] Breathing chest radiography using a dynamic flat-panel detector combined with computer analysis
    Tanaka, R
    Sanada, S
    Suzuki, M
    Kobayashi, T
    Matsui, T
    Inoue, H
    Yoshihisa, N
    [J]. MEDICAL PHYSICS, 2004, 31 (08) : 2254 - 2262
  • [9] Performance evaluation of a flat-panel detector-based microtomography system for small-animal imaging
    Kim, HK
    Lee, SC
    Chun, IK
    Cho, MH
    Cho, MH
    Lee, SY
    Joo, KS
    Cho, GS
    [J]. 2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 2108 - 2113
  • [10] A flat-panel detector digital radiography and a storage phosphor computed radiography: Screening for pneumoconioses
    Takashima, Yoshihiro
    Suganuma, Narufumi
    Sakurazawa, Hirofumi
    Itoh, Harumi
    Hirano, Harukazu
    Shida, Hisao
    Kusaka, Yukinori