Assessment of two 3-D fluoroscopic systems for articular fracture reduction: a cadaver study

被引:12
|
作者
Weil, Yoram A. [1 ]
Liebergall, Meir [1 ]
Mosheiff, Rami [1 ]
Singer, Syndie B. [2 ]
Joskowicz, Leo [3 ]
Khoury, Amal [1 ]
机构
[1] Hadassah Hebrew Univ Hosp, Dept Orthopaed, IL-91120 Jerusalem, Israel
[2] Univ Toronto, Dept Orthopaed, Toronto, ON M5S 1A1, Canada
[3] Hebrew Univ Jerusalem, Sch Engn & Comp Sci, Jerusalem, Israel
关键词
Articular fracture; 3D fluoroscopy; Intraoperative imaging; Tibial plateau fracture; IMAGE INTENSIFIER; CT; CONSEQUENCES; ISO-C-3D; CARE;
D O I
10.1007/s11548-011-0548-6
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective The most commonly used imaging device for assessment of fracture reduction is the two-dimensional X-ray fluoroscope. Two recently introduced 3D fluoroscopic devices, the Siremobil ISO-C3D (Siemens) and the C-InSight (Mazor Surgical Technologies), enable the surgeon to obtain spatial information for the assessment of articular reduction and hardware placement. The purpose of this study was to assess the reliability and accuracy of these two 3D fluoroscopic systems in measuring articular reduction in a cadaveric tibial plateau fracture. Methods Six cadaveric knee specimens were osteotomized at the lateral tibial plateau and fixed with a maximal articular step-off of 0, 1, 2.5, 5 and 7.5 mm. Each specimen was scanned 10 times with two 3D fluoroscopes, the Siremobil ISO-C3D and the C-InSight. The resulting images were reformatted and interpreted for articular displacements at four different locations at the plateau level and were compared with high-resolution CT scans by an independent observer. Results For the non-displaced fracture, no displacement (mean < 0.1 mm) was observed in either modality. The mean scanning time for the ISO-C3D was 2 min, while each C-InSight scan took 20 s. The readings at four different points along the malreduced fractures were similar for most measurements with either of the two modalities. The C-InSight readings were less accurate than those of the ISO-C3D, relative to the CT scan, but most errors were within clinically acceptable limits (< 2 mm) and used less radiation. Conclusions Intraoperative 3D fluoroscopes can detect clinically significant intra-articular step-off with acceptable measurement errors, using newer devices that enable the use of a conventional C-arm and reduced radiation.
引用
收藏
页码:685 / 692
页数:8
相关论文
共 50 条
  • [1] Assessment of two 3-D fluoroscopic systems for articular fracture reduction: a cadaver study
    Yoram A. Weil
    Meir Liebergall
    Rami Mosheiff
    Syndie B. Singer
    Leo Joskowicz
    Amal Khoury
    International Journal of Computer Assisted Radiology and Surgery, 2011, 6 : 685 - 692
  • [2] Detection of articular perforations of the proximal humerus fracture using a mobile 3D image intensifier – a cadaver study
    Jan Theopold
    Kevin Weihs
    Christine Feja
    Bastian Marquaß
    Christoph Josten
    Pierre Hepp
    BMC Medical Imaging, 17
  • [3] Detection of articular perforations of the proximal humerus fracture using a mobile 3D image intensifier - a cadaver study
    Theopold, Jan
    Weihs, Kevin
    Feja, Christine
    Marquass, Bastian
    Josten, Christoph
    Hepp, Pierre
    BMC MEDICAL IMAGING, 2017, 17
  • [4] Fluoroscopic 3-D reconstruction of coronary stents
    Close, RA
    Morioka, CA
    Whiting, JS
    PHYSICS OF MEDICAL IMAGING - MEDICAL IMAGING 1997, 1997, 3032 : 445 - 450
  • [5] Assessment of the stability of fracture fixation systems: Mechanical device to investigate the 3-D stiffness in vitro
    Kassi, JP
    Hoffmann, JE
    Heller, M
    Raschke, M
    Duda, GN
    BIOMEDIZINISCHE TECHNIK, 2001, 46 (09): : 247 - 252
  • [6] Materials and systems for two photon 3-D ROM devices
    Dvornikov, AS
    Cokgor, I
    Wang, M
    McCormick, FB
    Esener, SC
    Rentzepis, PM
    IEEE TRANSACTIONS ON COMPONENTS PACKAGING AND MANUFACTURING TECHNOLOGY PART A, 1997, 20 (02): : 203 - 212
  • [7] The zygomatic-sphenoid fracture line in malar reduction - A cadaver study
    Burns, JA
    Park, SS
    ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 1997, 123 (12) : 1308 - 1311
  • [8] 3-D FLUOROSCOPIC INVESTIGATION OF SOLVENT EFFECTS IN BINARY-MIXTURES
    SMITH, GC
    BORGES, SM
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1990, 199 : 77 - ANYL
  • [9] Two-dimensional fluoroscopic navigation in posterior cruciate ligament reconstruction: a preclinical cadaver study
    Ralf E. Rosenberger
    Reto J. Bale
    Cornelia Kneisl
    Dietmar Krappinger
    Martin Knoflach
    Rene Attal
    Archives of Orthopaedic and Trauma Surgery, 2010, 130 : 971 - 976
  • [10] Two-dimensional fluoroscopic navigation in posterior cruciate ligament reconstruction: a preclinical cadaver study
    Rosenberger, Ralf E.
    Bale, Reto J.
    Kneisl, Cornelia
    Krappinger, Dietmar
    Knoflach, Martin
    Attal, Rene
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2010, 130 (08) : 971 - 976