Cerenkov counting as a complement to liquid scintillation counting

被引:4
|
作者
Scarpitta, SC
Fisenne, IM
机构
[1] Environ. Measurements Laboratory, U.S. Department of Energy, New York
关键词
Charged particles - Radiation - Radioisotopes - Scintillation - Spectrometers;
D O I
10.1016/0969-8043(96)00061-9
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
A commercially available spectrometer was calibrated for liquid scintillation (LS) and Cerenkov counting efficiency (CCE) using National Institute of Standards and Technology traceable solutions. The CCE increased linearly over a 3 order of magnitude range in K-40 beta activity, and by 42% per MeV as beta-energies increased from 0.300 to 3.54 MeV, achieving a maximum value of 80% for Ru-106/Rh-106. The CCE can be enhanced by 10-15% when a wavelength shifter is used. A comparison of the data showed that the CCE was typically 20-50% less than the LS counting efficiency for beta-particles with maximum energies >1 MeV. Applications that utilize sequential CCE and LS counting to quantitate activity concentrations are discussed for samples containing two beta-emitting nuclides of differing energies. Published by Elsevier Science Ltd
引用
收藏
页码:795 / 800
页数:6
相关论文
共 50 条