摘要
Objective: Since nutritional screening is not routinely and accurately performed by busy care workers, the aim of this study was to evaluate a self-screening electronic measuring station that includes sonic height measurements. Methods: In all, 114 patients, 18 to 85 y of age and attending gastrointestinal outpatient clinics, followed automatically triggered audio-recorded instructions for weight and height measurements. The patients also provided information about unintentional weight loss to establish malnutrition risk using the Malnutrition Universal Screening Tool (MUST). In 56 healthy individuals, the effect of head/foot positions on height was examined using video-recordings. Laboratory studies examined the effects of hair/wigs, the position of a skull and horizontal plates and ambient conditions. Measurements were also made on a mechanical machine for comparison. Results: Of the patients, 21.9% were malnourished, with 99% agreement between sonic and mechanical machine categorization. Patients self-screened in only 35.6 +/- 14.8 s (median 32 s) and 77% rated the screening as very easy (22%, easy), despite encountering some remediable snags. Within-subject precision for height was 0.186 cm in healthy individuals and 0.368 cm in patients. Humidity and barometric pressure had negligible/undetectable effects on height measurements, but temperature corrections were confounded by calibration errors. In the most lateral standing positions, height was underestimated curvilinearly. In healthy individuals, height measurements were 0.353 +/- 0.542 cm lower on the sonic than mechanical device, which was inadequately explained by standing position or body tilt, although hair was found to be "invisible" only to the sonic machine. Conclusion: A method has been developed to rapidly and reliably self-screen for malnutrition using MUST, avoiding calculation and categorization errors, while providing results that can be immediately printed or transmitted electronically into patient notes. Crown Copyright (C) 2019 Published by Elsevier Inc. All rights reserved.
引用
收藏
页数:8
相关论文
共 3 条
  • [1] Rapid and reliable self-screening for nutritional risk in hospital outpatients using an electronic system
    McGurk, Paula
    Jackson, John M.
    Elia, Marinos
    NUTRITION, 2013, 29 (04) : 693 - 696
  • [2] Statistical Evaluation of Localization Performance and its Application in an Internet-Based Self-Screening Test Using a Virtual Environment
    Borss, C.
    Martin, R.
    Lyzenga, J.
    ACTA ACUSTICA UNITED WITH ACUSTICA, 2010, 96 (05) : 936 - 946
  • [3] Implementing a remote self-screening programme for atrial fibrillation using digital health technology among community-dwellers aged 75 years and older: a qualitative evaluation
    Wong, Kam Cheong
    Nguyen, Tu N.
    Trankle, Steven A.
    Usherwood, Tim
    Quintans, Desi
    Marschner, Simone
    Turnbull, Samual
    Indrawansa, Anupama Balasuriya
    White, Rose
    Burns, Mason Jenner
    Gopal, Vishal
    Lindley, Richard, I
    Kumar, Saurabh
    Chow, Clara K.
    BMJ OPEN, 2024, 14 (10):