Up-to-date radon-thoron discriminative detector for a large scale survey

被引:89
|
作者
Tokonami, S
Takahashi, H
Kobayashi, Y
Zhuo, WH
Hulber, E
机构
[1] Natl Inst Radiol Sci, Radon Res Grp, Res Ctr Radiat Safety, Inage Ku, Chiba 2638555, Japan
[2] Radosys Co Ltd, H-1116 Budapest, Hungary
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2005年 / 76卷 / 11期
关键词
D O I
10.1063/1.2132270
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An up-to-date radon-thoron discriminative detector has been developed for conducting a large scale survey. Comparing with our previous detector, some functional problems have been solved. The lowest and highest detection limits of the detector were estimated to be around 5 and 1000 Bq m(-3) for radon, and 15 and 1000 Bq m(-3) for thoron, respectively, with a 6 month exposure and several theoretical assumptions. Small indoor survey were carried out in Japan and in Hungary using the present detector. The presence of thoron was demonstrated in any indoor surveys by the two results. Since any measurements without discrimination of radon isotopes will result in different risk estimates from actual situations, a special attention should be paid to thoron and its concentration should be accordingly measured as well as the radon concentration. (c) 2005 American Institute of Physics.
引用
收藏
页码:1 / 5
页数:5
相关论文
共 50 条
  • [1] A new passive radon-thoron discriminative measurement system
    Sciocchetti, G.
    Sciocchetti, A.
    Giovannoli, P.
    DeFelice, P.
    Cardellini, F.
    Cotellessa, G.
    Pagliari, M.
    [J]. RADIATION PROTECTION DOSIMETRY, 2010, 141 (04) : 462 - 467
  • [2] Measurement of Indoor Thoron Gas Concentrations Using a Radon-Thoron Discriminative Passive Type Monitor: Nationwide Survey in Japan
    Sanada, Tetsuya
    [J]. INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 2021, 18 (03) : 1 - 8
  • [3] Performance test of passive radon-thoron discriminative detectors on environmental parameters
    Sorimachi, Atsuyuki
    Tokonami, Shinji
    Omori, Yasutaka
    Ishikawa, Tetsuo
    [J]. RADIATION MEASUREMENTS, 2012, 47 (06) : 438 - 442
  • [4] EFFECTS OF AIR EXCHANGE PROPERTY OF PASSIVE-TYPE RADON-THORON DISCRIMINATIVE DETECTORS ON PERFORMANCE OF RADON AND THORON MEASUREMENTS
    Omori, Y.
    Janik, M.
    Sorimachi, A.
    Ishikawa, T.
    Tokonami, S.
    [J]. RADIATION PROTECTION DOSIMETRY, 2012, 152 (1-3) : 140 - 145
  • [5] Radon-thoron discriminative measurements in Gansu province, China, and their implication for dose estimates
    Yamada, Y
    Sun, QF
    Tokonami, S
    Akiba, S
    Zhou, WH
    Hou, CS
    Zhang, SZ
    Ishikawa, T
    Furukawa, M
    Fukutsu, K
    Yonehara, H
    [J]. JOURNAL OF TOXICOLOGY AND ENVIRONMENTAL HEALTH-PART A-CURRENT ISSUES, 2006, 69 (7-8): : 723 - 734
  • [6] Radon-thoron discriminative measurements in the high natural radiation areas of southwestern Cameroon
    Saidou
    Tokonami, Shinji
    Janik, Miroslaw
    Samuel, Bineng Guillaume
    Abdourahimi
    Emmanuel, Ndjana Nkoulou Ii Joseph
    [J]. JOURNAL OF ENVIRONMENTAL RADIOACTIVITY, 2015, 150 : 242 - 246
  • [7] Electret ion chamber-based passive radon-thoron discriminative monitors
    Kotrappa, P.
    Steck, D.
    [J]. RADIATION PROTECTION DOSIMETRY, 2010, 141 (04) : 386 - 389
  • [8] RADON-THORON DETECTION PROPERTIES OF A CR-39 PLASTIC TRACK DETECTOR
    KHAN, HA
    AHMAD, I
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1981, 185 (1-3): : 401 - 405
  • [9] RADON-THORON ESTIMATION USING LR-115 PLASTIC TRACK DETECTOR
    SINGH, M
    SINGH, NP
    SINGH, S
    VIRK, HS
    [J]. NUCLEAR TRACKS AND RADIATION MEASUREMENTS, 1984, 8 (1-4): : 415 - 418
  • [10] Invited Article: In situ comparison of passive radon-thoron discriminative monitors at subsurface workplaces in Hungary
    Kavasi, Norbert
    Vigh, Tamas
    Nemeth, Csaba
    Ishikawa, Tetsuo
    Omori, Yasutaka
    Janik, Miroslaw
    Yonehara, Hidenori
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2014, 85 (02):