Anatomical and Technical Predictors of Three-Dimensional Mitral Valve Area Reduction After Transcatheter Edge-To-Edge Repair

被引:13
|
作者
Kassar, Mohammad [1 ]
Praz, Fabien [1 ]
Hunziker, Lukas [1 ]
Pilgrim, Thomas [1 ]
Windecker, Stephan [1 ]
Seiler, Christian [1 ]
Brugger, Nicolas [1 ]
机构
[1] Univ Bern, Bern Univ Hosp, Dept Cardiol, Inselspital, CH-3010 Bern, Switzerland
关键词
Mitral valve area; Transcatheter edge-to-edge repair; MitraClip; Latrogenic mitral stenosis; Three-dimensional echocardiography; EUROPEAN ASSOCIATION; ECHOCARDIOGRAPHY; MANAGEMENT; SURGERY; SOCIETY; 2D;
D O I
10.1016/j.echo.2021.08.021
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Among current transcatheter therapies for the treatment of mitral regurgitation, the MitraClip (MC; Abbott Vascular, Abbott Park, IL) system is the most commonly used. MitraClip implantation is usually contraindicated in patients with a mitral valve area (MVA) < 4.0 cm(2). However, little is known about the real impact of MC implantation on MVA. Our goal was to investigate the factors influencing MVA reduction and derive the minimal MVA required to prevent the development of a clinically significant mitral stenosis (MVA < 1.5 cm(2)) in different clinical scenarios. Methods: Using three-dimensional data sets, the annulus and leaflet anatomy and MVA before clip implanta-tion (MVA(BC)) were assessed. After each MC implant (NTR or XTR), the relative MVA reduction and the absolute residual MVA were measured and their predictors evaluated. Results: The present analysis included 116 patients. An MC XTR was the first device implanted in 50% of the subjects, and 53% were treated with a single implant. The MVA reduction following one XTR was 57% +/- 7% versus 52% +/- 8% after one NTR (P = .001). A lower MVA reduction was observed when the MC was placed commissural/central versus paracentral (50% +/- 8% vs 57% +/- 7%, P < .0001). After a second device, the additional MVA reduction was higher when creating a triple-compared with a double-orifice morphology (34% +/- 11% vs 25% +/- 9%, P = .001). The MVA after one MC correlated with MVA(BC) as well as with the clip type and position (r = 0.91, P < .0001). The MVA(BC), orifice morphology, and first device position predicted MVA after two implants (r = 0.82, P < .0001). Based on the mathematical relationship between these param-eters, the minimal MVA(BC) needed in eight different clinical scenarios was summarized in a decision algorithm: the values ranged from 3.5 to 4.7 cm(2) for one and 4.5 to 6.3 cm(2) for two MC strategies. Conclusions: The minimal native MVA preventing clinically relevant MS after transcatheter edge-to-edge repair is predicted by the number and location of clip(s), orifice morphology, and device type. Based on these parameters, an algorithm has been derived to optimize patient selection and preprocedural planning.
引用
收藏
页码:96 / 104
页数:9
相关论文
共 50 条
  • [1] Three-dimensional echocardiography in transcatheter edge-to-edge tricuspid valve repair
    Orban, M.
    Wolff, S.
    Stolz, L.
    Braun, D.
    Stark, K.
    Mehr, M.
    Stocker, T.
    Orban, M.
    Hagl, C.
    Massberg, S.
    Nabauer, M.
    Hausleiter, J.
    EUROPEAN HEART JOURNAL, 2021, 42 : 268 - 268
  • [2] Mitral valve transcatheter edge-to-edge repair
    Hausleiter, Joerg
    Stocker, Thomas J.
    Adamo, Marianna
    Karam, Nicole
    Swaans, Martin J.
    Praz, Fabien
    EUROINTERVENTION, 2023, 18 (12) : 957 - +
  • [3] Predictors of Outcomes Following Transcatheter Edge-to-Edge Mitral Valve Repair
    Ben-Shoshan, Jeremy
    Overtchook, Pavel
    Buithieu, Jean
    Mousavi, Negar
    Martucci, Giuseppe
    Spaziano, Marco
    de Varennes, Benoit
    Lachapelle, Kevin
    Brophy, James
    Modine, Thomas
    Baumbach, Andreas
    Maisano, Francesco
    Prendergast, Bernard
    Tamburino, Corrado
    Windecker, Stephan
    Piazza, Nicolo
    JACC-CARDIOVASCULAR INTERVENTIONS, 2020, 13 (15) : 1733 - 1748
  • [4] Effect of a novel transcatheter edge-to-edge repair device on the three-dimensional geometry of mitral valve in degenerative mitral regurgitation
    Ge, Zhenyi
    Pan, Wenzhi
    Zhou, Daxin
    Li, Wei
    Wei, Lai
    Liu, Xianbao
    Pu, Zhaoxia
    Shu, Xianhong
    Pan, Cuizhen
    Ge, Junbo
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2021, 97 (01) : 177 - 185
  • [5] Robotic Mitral Valve Repair After Failed Transcatheter Edge-to-Edge Repair
    Rowe, Georgina
    Gill, George
    Trento, Alfredo
    Emerson, Dominic
    Roach, Amy
    Ramzy, Danny
    Chikwe, Joanna
    ANNALS OF THORACIC SURGERY, 2022, 113 (03): : E223 - E225
  • [6] Predictors for long-term survival after transcatheter edge-to-edge mitral valve repair
    Orban, M.
    Lesevic, H.
    Orban, M. W.
    Braun, D.
    Deseive, S.
    Sonne, C.
    Mehilli, J.
    Schunkert, H.
    Massberg, S.
    Hagl, C.
    Kastrati, A.
    Boekstegers, P.
    Nabauer, M.
    Ott, I.
    Hausleiter, J.
    EUROPEAN HEART JOURNAL, 2016, 37 : 405 - 406
  • [7] Mitral valve surgery after failed transcatheter edge-to-edge repair
    Mazur, Piotr
    Arghami, Arman
    Zheng, Clark
    Alkhouli, Mohamad
    Schaff, Hartzell V.
    Dearani, Joseph
    Daly, Richard C.
    Greason, Kevin
    Crestanello, Juan A.
    JTCVS TECHNIQUES, 2022, 14 : 79 - 88
  • [8] Predictors for long-term survival after transcatheter edge-to-edge mitral valve repair
    Orban, Mathias
    Orban, Martin
    Lesevic, Hasema
    Braun, Daniel
    Deseive, Simon
    Sonne, Carolin
    Hutterer, Lisa
    Grebmer, Christian
    Khandoga, Alexander
    Pache, Juergen
    Mehilli, Julinda
    Schunkert, Heribert
    Kastrati, Adnan
    Hagl, Christian
    Bauer, Axel
    Massberg, Steffen
    Boekstegers, Peter
    Nabauer, Michael
    Ott, Ilka
    Hausleiter, Joerg
    JOURNAL OF INTERVENTIONAL CARDIOLOGY, 2017, 30 (03) : 226 - 233
  • [9] DISPARITIES IN TRANSCATHETER EDGE-TO-EDGE MITRAL VALVE REPAIR
    Ebner, Bertrand F.
    Maning, Jennifer
    Grant, Jelani
    Vincent, Louis
    Olarte, Neal
    Olorunfemi, Odunayo
    Colombo, Rosario
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2021, 77 (18) : 1084 - 1084
  • [10] Long-Term Impact of Small Mitral Valve Orifice Area after Transcatheter Edge-to-Edge Mitral Valve Repair on Clinical Outcome: A Three-Dimensional Echocardiography Study
    Kagawa, Shunsuke
    Hasegawa, Hiroko
    Kuwajima, Ken
    Yamane, Takafumi
    Ogawa, Mana
    Patel, Dhairya
    Salseth, Tracy
    Sekhon, Navjot
    Skaf, Sabah
    Chakravarty, Tarun
    Makar, Moody
    Makkar, Raj R.
    Shiota, Takahiro
    JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2024, 37 (03) : 328 - 337