Investigation of scatter from out of the field of view and multiple scatter in PET using Monte Carlo simulations

被引:1
|
作者
Ye Ting [1 ,2 ]
Chai Pei [1 ]
Gao Juan [1 ,2 ]
Yun Ming-Kai [1 ]
Liu Shuang-Quan [1 ,2 ]
Shan Bao-Ci [1 ]
Wei Long [1 ]
机构
[1] Chinese Acad Sci, Key Lab Nucl Anal Tech, Inst High Energy Phys, Beijing 100049, Peoples R China
[2] Chinese Acad Sci, Grad Univ, Beijing 100049, Peoples R China
基金
国家高技术研究发展计划(863计划); 中国国家自然科学基金;
关键词
PET; Monte Carlo simulation; GATE; scattering; GATE; SCANNER; VALIDATION; PLATFORM; EMISSION; SINGLE; SPECT; MODEL;
D O I
10.1088/1674-1137/35/12/017
中图分类号
O57 [原子核物理学、高能物理学];
学科分类号
070202 ;
摘要
In fully three-dimensional (3D) positron emission tomography (PET) imaging, the scatter fraction (SF) is about 40%-60%, which may degrade the imaging quality severely. Scatter correction is important for high quality image reconstruction. Model-based scatter correction has been proved to be accurate and available in clinical PET. However, it does not correct the scatter from out of the field of view (OFOV) and multiple scatters. In this study, we demonstrate the radial and axial distribution of scatters from OFOV when the source is located in different radial positions. In order to apply the above conclusions to different PET systems, we characterize the scatters from OFOV as a function of the ratio of the scanner diameter to the length of the axial field of view (AFOV) by modeling several typical whole-body and micro PET systems. The proportions of true events (S(0-0)), single scatter of one photon (S(1-0)), single scatter of both photons (S(1-1)), double scatter of one photon (S(2-0)) and multiple scatter (S(m)) are also calculated and compared. Here the 3D-PET Monte Carlo simulations are performed with the Geant4 Application for Tomography Emission (GATE). In summary, the scatters from OFOV tend to be recorded on the lines of response (LOR) far away from the source. They have a much more serious impact on whole-body PET than micro PET depending on the ratio of scanner diameter to the length of AFOV. In whole-body PET, twice scatters including single scatter of both photons (S(1-1)) and double scatter of one photon (S(2-0)) add up to about 12% so that twice scatter correction must be taken into account to acquire a high quality reconstruction image.
引用
收藏
页码:1166 / 1171
页数:6
相关论文
共 50 条
  • [1] Investigation of scatter from out of the field of view and multiple scatter in PET using Monte Carlo simulations
    叶婷
    柴培
    高娟
    贠明凯
    刘双全
    单保慈
    魏龙
    [J]. Chinese Physics C, 2011, (12) : 1166 - 1171
  • [2] Use of Monte Carlo simulations for the scatter events analysis of PET scanners
    Gao Fei
    Yamada, Ryoko
    Watanabe, Mitsuo
    Liu Hua-Feng
    [J]. ACTA PHYSICA SINICA, 2009, 58 (05) : 3584 - 3591
  • [3] A Monte Carlo simulation study of scatter fraction and the impact of patient BMI on scatter in long axial field-of-view PET scanners
    Ghabrial, Amir
    Franklin, Daniel R.
    Zaidi, Habib
    [J]. ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2021, 31 (03): : 305 - 315
  • [4] Scatter analysis of the ClearPET™ neuro using Monte Carlo simulations
    Fulterer, Anna M.
    Schneider, Stephan
    Gundlich, Brigitte
    Musmann, Patrick
    Weber, Simone
    Buzug, Thorsten M.
    [J]. ADVANCES IN MEDICAL ENGINEERING, 2007, 114 : 109 - +
  • [5] MONTE-CARLO PET SIMULATIONS - EFFECT OF PHOTON POLARIZATION ON SCATTER ESTIMATES
    BICE, AN
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (05): : 1185 - 1188
  • [6] MULTIPLE SCATTER CORRECTIONS BY MONTE-CARLO
    PARKER, JB
    TOWLE, JH
    SAMS, D
    JONES, PG
    [J]. NUCLEAR INSTRUMENTS & METHODS, 1961, 14 (01): : 1 - 12
  • [7] A SETUP FOR KV CBCT SCATTER PREDICTION USING MONTE CARLO SIMULATIONS
    Poels, K.
    Van den Heuvel, F.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S47 - S48
  • [8] A MONTE-CARLO STUDY OF MULTIPLE SCATTER EFFECTS IN COMPTON SCATTER DENSITOMETRY
    SPELLER, RD
    HORROCKS, JA
    [J]. MEDICAL PHYSICS, 1988, 15 (05) : 707 - 712
  • [9] Characterization of the scatter component in large axial field-of-view PET scanners: a Monte Carlo simulation study
    Ghabrial, Amir
    Franklin, Daniel
    Zaidi, Habib
    [J]. 2018 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE PROCEEDINGS (NSS/MIC), 2018,
  • [10] MULTIPLE SCATTER CORRECTIONS USING THE MONTE-CARLO PROGRAM MAGGIE
    PARKER, JB
    TOWLE, JH
    SAMS, D
    GILBOY, WB
    PURNELL, AD
    STEVENS, HJ
    [J]. NUCLEAR INSTRUMENTS & METHODS, 1964, 30 (01): : 77 - 87