Biomechanical comparison of median sternotomy closures

被引:75
|
作者
Losanoff, JE
Collier, AD
Wagner-Mann, CC
Richman, BW
Huff, H
Hsieh, FH
Diaz-Arias, A
Jones, JW
机构
[1] Univ Missouri, Sch Med, Dept Surg, Hlth Sci Ctr M580, Columbia, MO 65212 USA
[2] Univ Missouri, Dept Pathol, Columbia, MO 65201 USA
[3] Univ Missouri, Dept Biol Engn, Columbia, MO 65201 USA
[4] Univ Missouri, Dept Psychiat, Columbia, MO 65201 USA
来源
ANNALS OF THORACIC SURGERY | 2004年 / 77卷 / 01期
关键词
D O I
10.1016/S0003-4975(03)01468-1
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Poor healing of median sternotomy can significantly increase morbidity, mortality, and hospital costs. Effective union requires reliable sternal fixation. Although wire has proven the most reliable and widely used sternotomy closure material, no experimental studies have compared a large variety of wiring techniques in a human model. We developed an easily reproducible experimental model using cadaveric human sterna and compared several wiring methods to assess closure strength and stability. Methods. Fifty-three fresh adult human cadaveric sternal plates with adjacent ribs were fixed with specially designed spiked stainless steel clamps and attached to a texture analyzer. Single peristernal and transsternal, alternating single peristernal and transsternal, figure-eight peristernal, figure-eight pericostal, and Robicsek closures using no. 5 stainless steel wires were tested. We evaluated bone density, stiffness, and displacement using perpendicular, repetitive variable force loads of 800 Newtons cycling at a rate of 0.5 mm/s. Results. There were no significant differences in age, sex, or bone density in outcome measures of the sternal groups. No clamp failures or clamp damage to the specimens occurred. The single peristernal and alternating peristernal and transsternal closures proved superior in strength and stability (p<0.001). The figure-eight peristernal, then the single transsternal, then the Robicsek were next stablest groups in decreasing order. The figure-eight pericostal closure had the highest failure rate (p<0.001). Conclusions. This novel model of sternotomy closure testing was reliable, inexpensive, and easily reproducible. The mechanical stability of peristernal and alternating peristernal and transsternal wires was significantly greater than that of the other tested methods. Pericostal figure-eight closures were not sufficiently stable to be considered a reliable method of primary sternotomy repair.
引用
收藏
页码:203 / 209
页数:7
相关论文
共 50 条
  • [1] Comparison of straight median sternotomy and interlocking sternotomy with respect to biomechanical stability
    Kucukdurmaz, Fatih
    Agir, Ismail
    Bezer, Murat
    [J]. WORLD JOURNAL OF ORTHOPEDICS, 2013, 4 (03): : 134 - 138
  • [2] Fatigue testing median sternotomy closures
    Casha, AR
    Gauci, M
    Yang, L
    Saleh, M
    Kay, PH
    Cooper, GJ
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2001, 19 (03) : 249 - 253
  • [3] Biomechanical porcine model of median sternotomy closure
    Losanoff, JE
    Foerst, JR
    Huff, H
    Richman, BW
    Collier, AD
    Hsieh, FH
    Lee, S
    Jones, JW
    [J]. JOURNAL OF SURGICAL RESEARCH, 2002, 107 (01) : 108 - 112
  • [4] A biomechanical study of median sternotomy closure techniques
    Casha, AR
    Yang, L
    Kay, PH
    Saleh, M
    Cooper, GJ
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1999, 15 (03) : 365 - 369
  • [5] A biological model for biomechanical testing of median sternotomy closure
    Losanoff, JE
    Richman, BW
    Jones, JW
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2001, 20 (02) : 432 - 432
  • [6] Biomechanical comparison of canine median sternotomy closure using suture tape and orthopedic wire cerclage
    Rivenburg, Rachel E. E.
    Maxwell, Elizabeth A. A.
    Bertran, Judith
    Souza, Carlos H. De Mello
    Smith, Benjamin L. L.
    [J]. VETERINARY SURGERY, 2023, 52 (07) : 1057 - 1063
  • [7] A biomechanical comparison of three sternotomy closure techniques
    Cohen, DJ
    Griffin, LV
    [J]. ANNALS OF THORACIC SURGERY, 2002, 73 (02): : 563 - 568
  • [8] MEDIAN STERNOTOMY
    BRAIMBRI.MV
    [J]. LANCET, 1965, 1 (7385): : 585 - &
  • [9] MEDIAN STERNOTOMY
    DALTON, ML
    CONNALLY, SR
    [J]. SURGERY GYNECOLOGY & OBSTETRICS, 1993, 176 (06): : 615 - 624
  • [10] A biomechanical study of median sternotomy closure techniques - Appendix A. Conference discussion
    Loisance, D
    Casha
    Haverich, A
    Van Swieten, H
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1999, 15 (03) : 369 - 369