A new power sensor calibration system incorporating a vector network analyzer (VNA) has been developed. The vector-based system is able to account for the various impedance mismatches at the calibration test port, thus mathematically eliminating the mismatch uncertainty component of the power calibration. With scalar-based systems, this component is a major contributor to the overall calibration uncertainty. The new vector-based system reduces the overall calibration uncertainty significantly. This article discusses the new system in detail and presents measurement data and associated uncertainties.
引用
收藏
页码:144 / +
页数:4
相关论文
共 50 条
  • [1] Vector corrected power sensor calibration system
    Fenton, JR
    51ST ARFTG CONFERENCE DIGEST, 1998, : 94 - 98
  • [2] Acoustic Vector-Corrected Impedance Meter
    Scott, Jonathan
    Pennington, Kyle E.
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2014, 63 (12) : 2726 - 2732
  • [3] An Oscilloscope Correction Method for Vector-Corrected RF Measurements
    Gustafsson, Sebastian
    Thorsell, Mattias
    Stenarson, Jorgen
    Fager, Christian
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2015, 64 (09) : 2541 - 2547
  • [4] Full Two-Port Vector-Corrected Network Analyzer in the Acoustic Domain
    Macdonell, Marcus
    Scott, Jonathan
    JOURNAL OF THE AUDIO ENGINEERING SOCIETY, 2022, 70 (03): : 185 - 198
  • [5] Waveguide Joint Design and Validation for use in Acoustic Vector-corrected Network Analysers
    MacDonell, Marcus
    Basnet, Keshav
    Scott, Jonathan
    2019 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC), 2019, : 874 - 878
  • [6] Vector corrections for an automated power sensor calibration system
    Burns, JG
    50TH ARFTG CONFERENCE DIGEST, 1997, : 84 - 90
  • [7] VECTOR ANALYZER SPEEDS POWER-SENSOR CALIBRATION RESULTS
    BURNS, JG
    MICROWAVES & RF, 1987, 26 (08) : 91 - &
  • [8] Vector-corrected astigmatism after suture removal in penetrating keratoplasty - A comparison between nonmechanical and mechanical trephination
    Langenbucher, A
    Seitz, B
    Kus, MM
    Vilchis, E
    Naumann, GCH
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1996, 37 (03) : 2513 - 2513
  • [9] Vector-corrected Nonlinear Multi-port IQ-mixer Characterization using Modulated Signals
    Gustafsson, Sebastian
    Thorsell, Mattias
    Buisman, Koen
    Fager, Christian
    2017 IEEE MTT-S INTERNATIONAL MICROWAVE SYMPOSIUM (IMS), 2017, : 1429 - 1432
  • [10] INVESTIGATION AND CALIBRATION OF A FORCE VECTOR SENSOR WITH A CALIBRATION ARTEFACT
    Lietz, Sara
    Tegtmeier, Falk
    Roeske, Dirk
    Kumme, Rolf
    Schwind, Daniel
    XIX IMEKO WORLD CONGRESS: FUNDAMENTAL AND APPLIED METROLOGY, PROCEEDINGS, 2009, : 299 - 302